Developing and testing a robotic MRI/CT fusion biopsy technique using a purpose-built interventional phantom

Abstract

Background

Magnetic resonance imaging (MRI) can be used to target tumour components in biopsy procedures, while the ability to precisely correlate histology and MRI signal is crucial for imaging biomarker validation. Robotic MRI/computed tomography (CT) fusion biopsy offers the potential for this without in-gantry biopsy, although requires development.

Methods

Test–retest T1 and T2 relaxation times, attenuation (Hounsfield units, HU), and biopsy core quality were prospectively assessed (January–December 2021) in a range of gelatin, agar, and mixed gelatin/agar solutions of differing concentrations on days 1 and 8 after manufacture. Suitable materials were chosen, and four biopsy phantoms were constructed with twelve spherical 1–3-cm diameter targets visible on MRI, but not on CT. A technical pipeline was developed, and intraoperator and interoperator reliability was tested in four operators performing a total of 96 biopsies. Statistical analysis included T1, T2, and HU repeatability using Bland–Altman analysis, Dice similarity coefficient (DSC), and intraoperator and interoperator reliability.

Results

T1, T2, and HU repeatability had 95% limits-of-agreement of 8.3%, 3.4%, and 17.9%, respectively. The phantom was highly reproducible, with DSC of 0.93 versus 0.92 for scanning the same or two different phantoms, respectively. Hit rate was 100% (96/96 targets), and all operators performed robotic biopsies using a single volumetric acquisition. The fastest procedure time was 32 min for all 12 targets.

Conclusions

A reproducible biopsy phantom was developed, validated, and used to test robotic MRI/CT-fusion biopsy. The technique was highly accurate, reliable, and achievable in clinically acceptable timescales meaning it is suitable for clinical application.

Key points

  • The biopsy phantom had magnetic resonance imaging visible/computed tomography invisible targets and was highly reproducible.

  • Four operators with different levels of experience sampled all targets (96/96) successfully.

  • The fastest operator biopsied all twelve targets within 32 min.

Background

Magnetic resonance imaging (MRI) confers exquisite soft-tissue contrast and the ability to probe multiple biophysical characteristics in a single examination [1] and may direct biopsy sampling of deterministic tumour regions. To understand the underpinnings of MRI signals, imaging biomarker studies [2] and radiogenomic studies [3] also require precisely localised pathological correlation with a view toward ‘digital biopsy’ [4, 5]. Whilst in-gantry biopsy could achieve these aims, biopsies are usually performed under ultrasound or computed tomography (CT) guidance due to lower cost, wider availability, and ease of use [6].

MRI/CT-fusion provides a potential solution, exploiting the advantages of both imaging modalities, although remains largely undeveloped. A CT-guided interventional robot (MAXIO, Perfint, Chennai, IN) has capabilities in multiplanar planning, stereotactic targeting, and new work-in-progress MRI/CT-fusion software, which is still in development outside of United States Food and Drug Administration labelled use and has not yet been used clinically. The IDEAL recommendations for new interventional procedures state that preclinical research should be conducted prior to first-in-human studies [7] and a standardised interventional phantom would be a suitable means of achieving this, assisting with the clinical development and validation of the fusion software. However, commercially available phantoms cannot be biopsied, degrade with use and have components which can be seen on both MRI and CT, making image fusion unnecessary [8].

Here, we design and validate a purpose-built MRI/CT fusion biopsy phantom, containing targets that can be seen on MRI but not on CT, can be biopsied and assessed for diagnostic material. We then use the phantom to establish a biopsy technique pipeline, test procedural reliability, and minimise risk prior to potential clinical translation.

Methods

This prospective study was carried out at the Royal Marsden Hospital between January and December 2021. Ethical review was waived due to non-clinical nature. Data generated or analysed during the study are available from the corresponding author by request.

Phantom development

Material investigation (tray experiments)

High-strength (250 ‘bloom’) bovine gelatin and agar were chosen as potential materials due to non-toxicity, low cost, availability, prior use in interventional phantoms, and the ability to generate different MRI contrast [9, 10]. We investigated a range of concentrations, including mixed solutions, in silicone trays (100 mL per cube) as controlled experiments (Fig. 1). Trays underwent both CT and MRI to determine their attenuation and relaxation times respectively, and biopsy for core quality (Fig. 1).

Fig. 1
figure 1

Concentrations and core adequacy of gelatin, agar, and mixed gelatin/agar samples. Concentrations (shown as white text overlaid on images) denote ratios of gelatin or agar to water, by weight ratio. Mixed gelatin/agar samples contain equal volumes of gelatin and agar solutions at specified concentrations. Biopsy samples taken on day 1 are shown below each tray

All CT acquisitions in this study were carried out using the same protocol. Images were acquired in the axial plane, using a CT scanner with 64 detector rows (Definition Edge, Siemens Healthineers, Erlangen, Germany) and a 3-mm slice thickness with 1-mm slice interval.

Room temperature trays were placed in a bath of doped water (770 mg × L−1 CuSO4, 2,000 mg × L−1NaCl) and magnetic resonance images were acquired using a 1.5-T scanner (Magnetom Aera, Siemens Healthineers, Erlangen, Germany). T1 relaxation times were estimated using inversion-prepared turbo spin-echo (TSE) sequences (inversion times 35−3,000 ms, repetition time 4,500 ms) and T2 relaxation times using a multi-contrast spin-echo sequence (echo times 10−320 ms, repetition time 5,000 ms) as described previously [11]. Trays were imaged twice on day 1 after manufacture (2 h apart) to assess repeatability, and day 8 to assess medium-term stability over a typical period of use.

T1 and T2 relaxation times were estimated using in-house software (Adept, ICR). Curves were fitted pixel-by-pixel and median values of a region-of-interest (750 mm2) calculated for each cube. Hounsfield units (HU) were measured at the same position using CT images, acquired on days 1 and 8. A pair of biopsies was taken from each cube using a 16-G biopsy instrument on both days 1 and 8 (Tru-Core II, Argon Medical, TX, USA), and core quality assessed visually.

Phantom design and construction

A full description of phantom design, construction, and costings are provided in the Additional file 1. In brief, 1:10 gelatin was chosen as the background material, and equal parts 1:10-gelatin and 6:100-agar as the target material since these workable materials provided adequate cores, approximately matched HU (both around 25), and different T2 relaxation times (approximately 500 ms versus 200 ms respectively) to generate MRI contrast. A total of four identical biopsy phantoms were constructed using acrylic boxes (25 × 25 × 15 cm3) filled with background (non-target) dyed gelatin, with four sets of three spherical targets composed of mixed gelatin/agar (diameters 1, 2, and 3 cm) at 10-cm depth (12 targets in total), dyed with different food colourings to distinguish targets from background when assessing biopsy cores. A custom-made stamp was used to create wells as the background gelatin set, on which the targets were fixed at reproducible positions. A square aperture (13.5 × 13.5 cm2) in the lid mandated in-plane biopsy (Y displacement only, targets 1–3); single oblique (XY for targets 4–6 and ZY for 7–9), and double oblique (XZY for targets 10–12) approaches (Fig. 2). Two of the phantoms (A and B) were constructed one week apart for reproducibility assessment and technique development, and two further phantoms (C and D) were constructed for multioperator reliability assessment.

Fig. 2
figure 2

Phantom. Feature demonstration (a–d); construction steps (j–l). a Acrylic box with scribed lines (arrow) for scanner laser alignment. b Black blind to prevent observation during biopsy. c Plate with hemispheres for well impressions. d Plate in position, with flanges (arrow) to keep stamp 5 cm from base. e 1:10 gelatin:water, heated to 50 °C in a water bath. f Stamp in position, making wells in the half-filled gelatin as the gelatin cools. g Stamp removed, leaving wells (arrow) in the set gelatin in which the targets rest. h Two hemispherical domes form each sphere, with a 2-mm hole (arrow) in the apex of one of the domes which accommodates an 18-G needle for injection filling of target material. i Formed target spheres, filled with coloured gelatin/agar mix. Red, blue, and black food colouring for 3-, 2-, and 1-cm targets, respectively. j Spheres resting in wells. k Complete phantom, with gelatin filled to the top and surface stickers on the side of the phantom for orientation (arrow). l Phantom with biopsy needle in situ, with surface markers for confirmation of orientation (arrow)

Phantom validation

Phantoms were imaged on the same MRI scanner as the trays using a three-dimensional T2-weighted TSE acquisition (axial, 104 slices, slice thickness 2.5 mm, echo time 140 ms, repetition time 1,500 ms, 1.4 signal averages, in-plane reconstructed pixel size 1.3 × 1.3 mm2, acquisition time 6 min 14 s). Target visibility was assessed visually while target position reproducibility was assessed qualitatively using image fusion (Horos, horosproject.org) and quantitatively using a Dice similarity coefficient (DSC) (MATLAB 2021a, Mathworks, Natick, MA) for phantoms A and B. All acquisitions were carried out the day after phantom construction.

Robotic biopsy

A commercially available interventional CT robot (MAXIO™, Perfint Healthcare, Chennai, India), licensed for interventional procedures in the chest, abdomen, and pelvis was used to perform fusion biopsies [12]. The integrated planning workstation allows up to six needle trajectories to be planned using multiplanar reformats, and has work-in-progress MRI fusion software (off label use). Each planned trajectory is translated to object space by an electromechanical arm (stereotaxy), which has grippers that hold a needle guide, through which biopsy needles are manually placed. Rigid MRI/CT fusion was performed using the planning workstation by defining between 3 and 5 corresponding points-of-interest using fixed structures on both the MRI and planning CT images, namely the corners of the acrylic box.

Feasibility and pipeline

One of the authors (E.J. 4-year experience in interventional radiology, including 3 months of clinical robotic intervention) developed a technical pipeline (Fig. 3) by taking biopsies from the targets in phantom A, with steps fully described in Additional file 1.

Fig. 3
figure 3

a Coronal/birds-eye on T2-weighted turbo spin-echo images with numbered targets. b Phantom on the CT table next to the robot. c Aligned phantom, fixed with adhesive putty. d MRI/CT-fusion and needle path planning using robot workstation. e Workstation screenshot showing CT (left), fused MR/CT (middle), and MRI (right) reference points. f Robot and inserted co-axial needle. g Radiologist performing a biopsy. h Robot workstation showing planned (left), actual (right), and fused (planned and actual, middle) needle paths. i Biopsy core with target (black) and non-target surround (very pale yellow); core length and target length are being measured. CT Computed tomography, MRI Magnetic resonance imaging

Multioperator reliability assessment

Four operators each attempted biopsy of all targets twice (back-to-back), 5−7 days after manufacture. Two operators biopsied phantom C and two operators biopsied phantom D using a 16-G × 20-cm biopsy needle (Tru-Core II, Argon Medical, TX, USA) which has a maximum core length of 19 mm. Operators had different levels of experience: E.J. (again, following pipeline development), N.F. (19-year experience with biopsy procedures including 3 months of clinical robotic intervention), J.L. (in-training, 2-year experience with biopsy procedures, including two months of clinical robotic intervention), J.W. (MRI physicist with no prior experience of conducting biopsy procedures but 11-year experience in medical imaging). Operators were blind to each other’s biopsies, apart from E.J. who attended all procedures to explain some aspects of software functionality, without physical input.

Freehand biopsy

After the robotic biopsies were carried out, a 16-G co-axial needle was positioned into the phantom in multiple (> 100) non-target regions of phantom D, to introduce non-target gas locules and make the location of targets uncertain. The quickest robotic operator then attempted freehand biopsy on day 7 after manufacture under CT guidance using low-dose sequential acquisitions and cognitive fusion, where measurements were translated from MRI to CT to estimate target position. As many targets as possible were biopsied in a 1-h time limit, starting with the 3-cm targets.

Biopsy core analysis

Cores were measured immediately after each biopsy (to minimise desiccation) by an independent operator (J.B.), in consensus with an operator not performing the biopsy procedure, to minimise bias. Each core was straightened (if necessary) on a piece of white card, and core length (CL)–the total length of the biopsy core (target + background material) and target length (TL)–(target material length only) marked alongside the core using an ultrafine (0.1 mm) black tip pen (adapted from [13]), and measured using a ruler to the nearest 0.5 mm.

Statistical analysis

Statistical analysis was carried out using GraphPad Prism (version 9, San Diego, CA) and MATLAB (2021a, Mathworks, Natick, MA). A p value of < 0.05 was used to indicate statistical significance.

Data were checked for normality using the Shapiro–Wilk test, and column statistics calculated. Repeatability of T1, T2, and HU in trays and phantoms was assessed using Bland–Altman analysis [14] and time taken to perform the first versus second phantom biopsies compared using paired t tests.

A flow diagram of study methods is shown in Fig. 4.

Fig. 4
figure 4

Flow diagram of study methods. CL Core length, DSC Dice similarity coefficient, HU Hounsfield units, TL Target length

Results

Tray experiments

T1, T2, and HU demonstrated strong dependence on gelatin and agar concentration and good repeatability with lower with upper 95% limits of agreement of -2.8 and 5.5%, -1.7% and 1.7%, and -3.8% and 14.1% respectively (Fig. 5).

Fig. 5
figure 5

a Coronal T2-weighted (echo time 50 ms) magnetic resonance images of pure gelatin (left) and agar (right), trays 1 and 2, showing that gelatin qualitatively returns high signal at all concentrations, whereas agar returns lower signal as concentration increases (the bottom two wells were filled with doped water). Attenuation values (HU) (b) T1 (c), and T2 (d) at different gelatin/agar concentrations. These graphs confirm relatively small differences in HU as the concentration of both substances increases and show that the T2 relaxation time is long in all gelatin samples and short in agar samples. Bland–Altman plots comparing differences between the two test–retest measurements for HU (e) T1 (f), and T2 (g). Solid lines represent the mean difference, and dashed lines the 95% limits of agreement

There was a tendency for greater core integrity with higher substance concentration at the expense of higher cost and difficulty dissolving substances. Biopsy core quality did not differ between days 1 and 8, confirming material stability.

Phantom validation

There was excellent contrast between targets and background material on MRI, without discernible CT contrast (Fig. 6). Target size and position highly reproducible between phantoms A and B, with a DSC of 0.92 (0.93 for scanning the same phantom).

Fig. 6
figure 6

Biopsy phantom. a Axial/side view T2-weighted magnetic resonance image, with dashed yellow line showing location of coronal reformats. b Coronal/birds’ eye view reformat of T2-weighted images. c Fused magnetic resonance images of two separate biopsy phantoms showing the same position of targets. d Computed tomography appearances using a narrow soft tissue (brain) window to confirm targets are invisible

Robotic biopsy

Technique feasibility was initially demonstrated by successfully taking 12 biopsies in 89 min with 100% hit rate (12/12), including scanning after each needle insertion, which showed from 0.0 to 1.0 mm of tip deviation.

At multioperator assessment, visible target material > 1 mm was achieved in all attempted biopsies (hit rate 100%, 96/96). Median CL was 15.6 mm (interquartile range [IQR] 14.6−17.0) across all targets. Median TL was 15.0 mm (IQR 14.0–17.3) for 3-cm targets, 15.0 mm (IQR 10.0–6.0) for 2-cm targets, and 7.8 mm (IQR 6.6–9.0) for 1-cm targets. The maximum CL, defined by sample notch size, was 19 mm. The second set of biopsies was significantly quicker than the first, (61 min 0 s ± 27 min 42 s [mean ± standard deviation] versus 49 min 30 s ± 20 min 30 s, p = 0.049), with the longest at 101 min (MRI physicist) and the quickest time at 32 min (attending radiologist) for all 12 biopsies, from start to finish (scanning, image fusion, and biopsy). Each set of biopsies was performed using a single volumetric acquisition, mean dose-length product of 145 mGy*cm (range 140–150), with a single needle positioning for each biopsy (i.e., no readjustments).

Using freehand guidance using cognitive image fusion, 3 targets (1, 5, and 12) were successfully biopsied, and 1 was missed entirely (target 9), with total dose-length product of 675 mGy*cm in a 1-h time slot, which is 8 times longer per target, and 18 times higher radiation dose per target than the fastest robotic operator.

Discussion

In this study, we sought to develop and rigorously validate a technique for MRI-guided biopsy in the CT scanner using image fusion and stereotactic targeting to better decide whether it is suitable for translation to clinical studies.

We first developed and validated a unique purpose-built, low-cost, non-toxic, MRI/CT fusion biopsy phantom, which had 12 dyed targets of varying difficulty levels that could be seen on MRI but not on CT, can be biopsied, and core adequacy assessed for presence of dyed target material. The similar DSC for two separate phantoms compared with two MRI scans of the same phantom (0.92 and 0.93) means that our well-controlled construction process yielded highly reproducible phantoms, which is crucial to maintain consistent difficulty and compare true operator performance.

Intraoperator and interoperator reliability can be readily assessed using phantoms, since they can be biopsied more than once, which allows risk to be minimised through training and development or abandonment of unsuccessful techniques. After showing initial feasibility, we tested reliability in a range of operators with different experience levels, including an MRI physicist without prior experience of performing biopsies. All operators biopsied targets with 100% (96/96) hit-rate including highly complex 20°–40° double oblique angulations in small targets, using a single volumetric acquisition and a clinically acceptable duration of around 30 min in the fastest operator, demonstrating that our technique is reliable and therefore suitable for clinical translation. The ability to select and execute complex oblique needle trajectories using multiplanar reformats is a key advantage of our robotic technique, which provides more available needle path options, and allows vital structures to be avoided.

The diagnostic yield of conventional CT-guided biopsy for lesions less than 1 cm in diameter is limited [15,16,17,18] without the added complexity of out-of-plane paths and image fusion. Indeed, our 0.0−1.0-mm tip deviation is highly accurate and in line with the results reported by Scharll et al. [19], who quoted needle placement accuracy of 1.5 ± 0.87 mm (mean ± standard deviation) in a structurally simple gelatin phantom with embedded aluminium targets, without MRI/CT fusion. Our studies are more accurate than another group who used the same robot without image fusion and a commercially available hydrogel polymer phantom, and reported a tip deviation of 6.5 ± 2.5 mm [12], which could be due to differences in phantom materials (hydrogel polymer gelatin), analytical methods, and a greater internal complexity of their phantom, with resemblance to anatomical structures.

Limitations of the phantom stem from its differences with patients including lack of movement, object simplicity making fusion straightforward, poor anatomical resemblance, and degradation of the phantom with use. Whilst needle path tracks were sometimes faintly visible on CT, they were not used to guide biopsy in our experiments. Their minimal impact is evidenced by the fact cores were successfully obtained in all targets on first attempt using the robotic technique.

Further work regarding phantom development could include reproduction at other centres, use for in-gantry biopsy and three-dimensional printing patient-specific phantoms [20]. Clinical translation should include the addition of a vacuum immobilisation mattress to reduce external motion [21]. We also advise taking a stepwise approach, where relatively fixed tumours are first investigated (e.g., MRI-guided bone biopsy [22, 23], followed by more mobile tumours using internal motion correction techniques as has been applied in robotic lung biopsy without image fusion using breath hold monitoring [24]. For example, whereas rigid registration (as used in our study) might be more suitable for non-deformable structures with fixed internal points (e.g., bone), deformable structures may benefit from more complex deformable registration algorithms [25].

If successful, MRI/CT fusion in biopsy procedures could allow the most deterministic intratumoural components to be targeted for more accurate classification and better management decisions. Tumour targeting in ablation procedures, combined with apnoea or high-frequency jet ventilation [26] would be improved for index tumours that are occult on CT, but not on MRI [27]. Consequently, there is an unmet clinical need for the development and validation of fusion software in combination with stereotactic interventional devices.

In conclusion, a reproducible biopsy phantom was successfully developed, validated, and used to rigorously test a robotic technique for MRI/CT-fusion biopsy. The technique was shown as highly accurate, reliable, and suitable for clinical translation.

Availability of data and materials

Available from the corresponding author on request.

Abbreviations

CL:

Core length

CT:

Computed tomography

DSC:

Dice similarity coefficient

HU:

Hounsfield units

IQR:

Interquartile range

MRI:

Magnetic resonance imaging

TL:

Target length

TSE:

Turbo spin-echo

References

  1. Mitchell DG, Burk DL Jr, Vinitski S, Rifkin MD (1987) The biophysical basis of tissue contrast in extracranial MR imaging. AJR Am J Roentgenol 149:831–837

    Article  CAS  PubMed  Google Scholar 

  2. O’Connor JPB, Aboagye EO, Adams JE et al (2017) Imaging biomarker roadmap for cancer studies. Nat Rev Clin Oncol 14:169–186. https://doi.org/10.1038/nrclinonc.2016.162

    Article  CAS  PubMed  Google Scholar 

  3. Lam A, Bui K, Hernandez Rangel E et al (2018) Radiogenomics and IR. J Vasc Interv Radiol 29:706–713. https://doi.org/10.1016/j.jvir.2017.11.021

    Article  PubMed  Google Scholar 

  4. Banna GL, Olivier T, Rundo F et al (2019) The promise of digital biopsy for the prediction of tumor molecular features and clinical outcomes associated with immunotherapy. Front Med 6:6–11. https://doi.org/10.3389/fmed.2019.00172

    Article  Google Scholar 

  5. Arthur A, Johnston EW, Winfield JM et al (2022) Virtual biopsy in soft tissue sarcoma. How Close Are We? Front Oncol 12:892620. https://doi.org/10.3389/fonc.2022.892620

    Article  PubMed  PubMed Central  Google Scholar 

  6. Veltri A, Bargellini I, Giorgi L et al (2017) CIRSE guidelines on percutaneous needle biopsy (PNB). Cardiovasc Intervent Radiol 40:1501–1513. https://doi.org/10.1007/s00270-017-1658-5

    Article  PubMed  Google Scholar 

  7. Hirst A, Philippou Y, Blazeby J et al (2019) No surgical innovation without evaluation: evolution and further development of the IDEAL Framework and Recommendations. Ann Surg 269:211–220. https://doi.org/10.1097/SLA.0000000000002794

    Article  PubMed  Google Scholar 

  8. CIRS inc IMAGE-GUIDED ABDOMINAL BIOPSY PHANTOM. https://www.cirsinc.com/products/ultrasound/zerdine-hydrogel/image-guided-abdominal-biopsy-phantom/. Accessed 20 Jan 2022

  9. Stoffner R, Augschöll C, Widmann G et al (2009) Accuracy and feasibility of frameless stereotactic and robot-assisted ct-based puncture in interventional radiology: a comparative phantom study. Rofo 181:851–858. https://doi.org/10.1055/s-0028-1109380

    Article  CAS  PubMed  Google Scholar 

  10. Zettinig O, Shah A, Hennersperger C et al (2015) Multimodal image-guided prostate fusion biopsy based on automatic deformable registration. Int J Comput Assist Radiol Surg 10:1997–2007. https://doi.org/10.1007/s11548-015-1233-y

    Article  PubMed  Google Scholar 

  11. Stupic KF, Ainslie M, Boss MA et al (2021) A standard system phantom for magnetic resonance imaging. Magn Reson Med 86:1194–1211. https://doi.org/10.1002/mrm.28779

    Article  PubMed  PubMed Central  Google Scholar 

  12. Koethe Y, Xu S, Velusamy G et al (2014) Accuracy and efficacy of percutaneous biopsy and ablation using robotic assistance under computed tomography guidance: A phantom study. Eur Radiol 24:723–730. https://doi.org/10.1007/s00330-013-3056-y

    Article  PubMed  Google Scholar 

  13. Carmona Echeverria LM, Haider A, Freeman A et al (2020) A critical evaluation of visual proportion of Gleason 4 and maximum cancer core length quantified by histopathologists. Sci Rep 10:1–9. https://doi.org/10.1038/s41598-020-73524-z

    Article  CAS  Google Scholar 

  14. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310. https://doi.org/10.1016/S0140-6736(86)90837-8

    Article  CAS  PubMed  Google Scholar 

  15. Stattaus J, Kuehl H, Ladd S et al (2007) CT-guided biopsy of small liver lesions: Visibility, artifacts, and corresponding diagnostic accuracy. Cardiovasc Intervent Radiol 30:928–935. https://doi.org/10.1007/s00270-007-9023-8

    Article  PubMed  Google Scholar 

  16. Chang YY, Chen CK, Yeh YC, Wu MH (2018) Diagnostic feasibility and safety of CT-guided core biopsy for lung nodules less than or equal to 8 mm: A single-institution experience. Eur Radiol 28:796–806. https://doi.org/10.1007/s00330-017-5027-1

    Article  PubMed  Google Scholar 

  17. De HM, Weng HH, Hsu SL et al (2019) Accuracy and complications of CT-guided pulmonary core biopsy in small nodules: A single-center experience. Cancer Imaging 19:1–10. https://doi.org/10.1186/s40644-019-0240-6

    Article  Google Scholar 

  18. Tian P, Wang Y, Li L et al (2017) CT-guided transthoracic core needle biopsy for small pulmonary lesions: Diagnostic performance and adequacy for molecular testing. J Thorac Dis 9:333–343. https://doi.org/10.21037/jtd.2017.02.16

    Article  PubMed  PubMed Central  Google Scholar 

  19. Scharll Y, Mitteregger A, Laimer G et al (2022) Comparison of a robotic and patient-mounted device for CT-guided needle placement : a phantom study. J Clin Med 11:3746. https://doi.org/10.3390/jcm11133746

    Article  PubMed  PubMed Central  Google Scholar 

  20. Filippou V, Tsoumpas C (2018) Recent advances on the development of phantoms using 3D printing for imaging with CT, MRI, PET, SPECT, and ultrasound. Med Phys 45:e740–e760. https://doi.org/10.1002/mp.13058

    Article  Google Scholar 

  21. Johnston EW, Basso J, Winfield J et al (2022) Starting CT guided robotic interventional oncology at a UK centre. Br J Radiol 95:20220217. https://doi.org/10.1259/bjr.20220217

    Article  PubMed  Google Scholar 

  22. Donners R, Figueiredo I, Tunariu N et al (2022) Multiparametric bone MRI can improve CT-guided bone biopsy target selection in cancer patients and increase diagnostic yield and feasibility of next-generation tumour sequencing. Eur Radiol 32:4647–4656. https://doi.org/10.1007/s00330-022-08536-6

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Kumar R, Singh SK, Mittal BR et al (2022) Safety and diagnostic yield of 68Ga prostate-specific membrane antigen PET/CT-guided robotic-assisted transgluteal prostatic biopsy. Radiology 303:392–398. https://doi.org/10.1148/RADIOL.204066

    Article  PubMed  Google Scholar 

  24. Anzidei M, Argirò R, Porfiri A et al (2015) Preliminary clinical experience with a dedicated interventional robotic system for CT-guided biopsies of lung lesions: a comparison with the conventional manual technique. Eur Radiol 25:1310–1316. https://doi.org/10.1007/s00330-014-3508-z

    Article  PubMed  Google Scholar 

  25. Oguro S, Tuncali K, Elhawary H, Morrison PR, Hata N, Silverman SG (2011) Image registration of pre-procedural MRI and intra-procedural CT images to aid CT-guided percutaneous cryoablation of renal tumors. Int J Comput Assist Radiol Surg 6:111–117. https://doi.org/10.1007/s11548-010-0485-9

    Article  PubMed  Google Scholar 

  26. Beyer LP, Wiggermann P (2017) Planning and guidance: New tools to enhance the human skills in interventional oncology. Diagn Interv Imaging 98:583–588. https://doi.org/10.1016/j.diii.2017.07.004

    Article  CAS  PubMed  Google Scholar 

  27. Schullian P, Johnston E, Laimer G et al (2020) Thermal ablation of CT “invisible” liver tumors using MRI fusion: a case control study. Int J Hyperthermia 37:564–572. https://doi.org/10.1080/02656736.2020.1766705

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Funding was obtained from the Royal of College of Radiologists, Pump Priming Grant. This study represents independent research supported by the National Institute for Health and Care Research (NIHR) Biomedical Research Centre and the Clinical Research Facilities at The Royal Marsden NHS Trust and the Institute of Target Research, London. The views expressed here are those of the author(s) and not necessarily those of the National Health Service, the National Institute for Health and Care Research, or the Department of Health. The robotic platform was supplied by Perfint Healthcare Pvt. Ltd. (Chennai, India) under a Materials Transfer Agreement between the Royal Marsden Hospital and Perfint Healthcare. The authors had full control of the data and made the decision to publish.

Funding

Funding was obtained from the Royal of College of Radiologists, Pump Priming Grant.

Author information

Authors and Affiliations

Authors

Contributions

EJ: concept, experimental study, data collection, data analysis, wrote paper. NF: experimental study, contributed to paper. CC: concept, experimental study, contributed to paper. JB: experimental study, data collection, data analysis. TT: concept, experimental study, contributed to paper. JL: experimental study, contributed to paper. CM: experimental study, contributed to paper. D-M K: experimental study, contributed to paper. JW: concept, experimental study, data collection, data analysis, contributed to paper. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Edward W. Johnston.

Ethics declarations

Ethics approval and consent to participate

Ethical review was waived due to the non-clinical nature of the study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Supplemental Table 1.

Phantom costings. Supplemental Figure 1. Targets accessible via in plane approach (red arrow), single oblique (green arrow) and double oblique approach (dotted blue arrow) as mandated by the box lid, and its opening. Note the ‘F’ markings on the side of the box and lid, denoting the ‘feet’ for orientation purposes (black arrowheads). Supplemental Figure 2. Design features of the phantom. A: schematic birds eye view of the phantom stamp. B: CAD model of the under surface of the stamp. White asterisks demonstrate defects which reduce surface area and thus interfacial tension between the stamp and gelatin, facilitating removal. C. CAD model of hemispheres (1, 2 and 3cm) + locating pegs. Supplemental Figure 3. The stamp hanging down onto gelatin as it sets, to create well imprints in the set gelatin.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Johnston, E.W., Fotiadis, N., Cummings, C. et al. Developing and testing a robotic MRI/CT fusion biopsy technique using a purpose-built interventional phantom. Eur Radiol Exp 6, 55 (2022). https://doi.org/10.1186/s41747-022-00308-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-022-00308-7

Keywords