Impact of fixation, coil, and number of excitations on diffusion tensor imaging of rat brains at 7.0 T

Abstract

Background

We sought to compare diffusion tensor imaging (DTI) parameters in vivo and ex vivo in the brain and to explore the effects of radiofrequency coil and number of excitations on ex vivo DTI parameters.

Methods

Six Sprague–Dawley rat brains were used to obtain in vivo and ex vivo DTI maps with different coils and number of excitations. DTI parameters of white matter and grey matter including diffusivities, fractional anisotropy, and other dimensionless ratios (λ2/λ1, λ3/λ1, and λ2/λ3) were obtained from reconstruction maps. Comparisons of ex vivo signal-to-noise ratio with different coils and number of excitations were conducted.

Results

Diffusivities decreased significantly after fixation in all the selected white matter and grey matter regions of interest (all at p < 0.001). The diffusivities in white matter integrity decreased more than in grey matter integrity after fixation (all at p < 0.001). The ratio of λ2/λ3 in the major brain structures changed after fixation (most at p < 0.05). There were differences in major ex vivo brain structures in DTI parameters and signal-to-noise ratio between surface coil and volume coil, and between one and four excitations (most at p < 0.05).

Conclusion

The impact of fixation, coil, and number of excitations on DTI parameters should be taken into consideration in clinical and experimental studies at 7.0 T.

Key points

  • Diffusivities of rat brain at 7.0 T were lower ex vivo than in vivo.

  • The dimensionless parameters were different under different scan conditions.

  • DTI studies should consider the effects of fixation, coil, and number of excitations.

Background

Diffusion tensor imaging (DTI) has been recognised to be able to characterize the central nervous system tissues, including the white matter (WM) and grey matter (GM) [1,2,3]. The main and essential DTI parameters are eigenvalues, sorted into the longest diffusivity (λ1), intermediate diffusivity (λ2), and lowest diffusivity (λ3) in the ellipsoid model [4]. On the basis of the eigenvalues, the conventionally used indices include axial diffusivity (AD, i.e. λ1), radial diffusivity (RD), mean diffusivity (MD), and fractional anisotropy (FA). Usually, DTI parameters change after axonal damage, demyelination, cerebral ischaemia, traumatic brain injury, and other nervous system diseases [5,6,7,8,9].

However, in vivo DTI has a disadvantage in that it is vulnerable to movement and it is time-consuming to acquire images with high resolution. Therefore, ex vivo DTI is widely used to acquire high-quality images with high resolution in clinical and basic studies with long-time scanning [10,11,12]. Moreover, high b value, thin slice, and a greater number of excitations are possible with ex vivo DTI. Previous studies have compared in vivo and ex vivo characteristics of DTI, finding that the diffusion decreased after fixation [12,13,14]. There were different results of comparison of FA in vivo and ex vivo in previous studies [14,15,16]. However, the scan conditions were not rigorously controlled in most previous investigations. To date, few studies have systematically compared the live and fixed DTI features under the same imaging conditions. In addition, the effects of radiofrequency coil and number of excitations on ex vivo DTI parameters remain to be determined. Therefore, we aimed at comparing in vivo and ex vivo DTI parameters under the same scan conditions and evaluating the effects of coil and number of excitations on ex vivo DTI.

Methods

In vivo and ex vivo preparations

All the procedures and experiments were approved by local Experimental Animal Ethics Committee. Six male Sprague–Dawley rats (body weight 380 ± 27 g, mean ± standard deviation) were used for the in vivo and ex vivo DTI studies. For in vivo preparation, rats were anaesthetised with isoflurane (3% for induction, 2–2.5% for maintenance, depending on respiration). Respiration and body temperature were continuously monitored during the scanning. For ex vivo preparation, brains were fixed by transcardial perfusion with 4% paraformaldehyde after euthanasia. The fixed brains were placed in plastic tubes with 4% paraformaldehyde and stored at 4 °C for more than 2 days. Before the scanning, the excised brains were immersed in the oil named Fomblin (Solvay, Brussels, Belgium) and sealed with plastic film to reduce susceptibility artefacts.

DTI acquisition

Magnetic resonance imaging was performed on a 7.0 T animal Scanner (Biospec 70/30, Bruker, Ettlingen, Germany). The brain receiving surface coil with a transmit radiofrequency coil was used for both in vivo and ex vivo brain imaging, whereas a volume coil with a 23-mm inner diameter and 44-mm outer diameter for both transmission and reception was only available for fixed brains, due to the size. After a shim, DTI images were acquired with the echo planar imaging sequence with 30 diffusion gradient directions (b = 1000 s/mm2), diffusion gradient duration/separation 4/20 ms, gradient speed 3353.45 T/m/s, repetition time/echo time 6250/ 32.2 ms, field of view 35 mm × 35 mm, matrix 128 × 128, and slice thickness 1 mm. Five additional DTI images with b = 0 s/mm2 were also obtained. One excitation was used in vivo, and both one and four excitations were used ex vivo. The scanning duration was about 15 min for one excitation and 60 min for four excitations.

Data analysis

Maps of anisotropic parameters including λ1, λ2, λ3, MD, and FA were reconstructed by ParaVision version 5.0 (Bruker, Ettlingen, Germany) using diffusion-weighted images. A reconstruction filter was used for the correction of the phase mismatch between even and odd echoes in ParaVision version 5.0. RD was calculated, where the formula is:

RD = (λ2 + λ3)/2

MD is the mean of λ1, λ2, and λ3. We also used λ1, λ2, and λ3 to calculate the dimensionless ratios of λ2/λ1 (R1), λ3/λ1 (R2), and λ2/λ3 (R3). The fractional reduction of different diffusivities after fixation was computed by using the equation:

(Din-Dex)/Din

(Din) is in vivo diffusivity and (Dex) is ex vivo diffusivity. The regions of interest (ROIs) comprised WM ROIs of the corpus callosum, external capsule, anterior commissure, and optic tract and GM ROIs of the cortex, hippocampus, and striatum [13, 17, 18]. Two to three axial slices were selected to bilaterally delineate each ROI for the mean value. Additionally, the internal capsule and cerebral peduncle were measured together as a WM ROI because the posterior internal capsule is near the cerebral peduncle while the anterior internal capsule was difficult to outline under the thickness of 1 mm. WM integrity was defined as all the WM ROIs as a whole, and GM integrity was defined as all the WM ROIs as a whole. Signal-to-noise ratio (SNR) was defined as the mean signal intensity acquired from reconstructed intensity maps in each ROI against the standard deviation of the background noise (4 mm2 box) divided by 0.66, in consideration of Rayleigh statistics [19].

Statistical analysis

Data were expressed as mean ± standard deviation (n = 6). SPSS 22 (IBM Corp., Armonk NY, USA) was used for statistical analysis. The paired t test was used for comparisons between in vivo and ex vivo parameters, between surface coil and volume coil ex vivo parameters, and between one and four excitations ex vivo. The data on WM integrity and GM integrity were compared using the independent-samples t test. A p value < 0.05 (two-tailed) was considered statistically significant.

Results

Comparisons of DTI parameters between WM integrity and GM integrity

FA and R3 of WM integrity were significantly higher than those of GM integrity both in vivo and ex vivo under different scanning conditions (all at p < 0.001; Fig. 1). Nevertheless, RD, R1, and R2 of WM integrity were significantly lower than those of GM integrity both in vivo and ex vivo under different scanning conditions (all at p < 0.001; Fig. 1). AD in WM integrity was similar to that in GM integrity.

Fig. 1
figure 1

Comparisons between white matter (WM) integrity and grey matter (GM) integrity in fractional anisotropy (FA), radial diffusivity (RD), λ2/λ1 ratio (R1), λ3/λ1 ratio (R2), and λ2/λ3 ratio (R3). There was higher FA and R3 in WM integrity than in GM integrity on both in vivo (IV) and ex vivo scans, whereas there was lower RD, R1, and R2 in WM integrity than GM integrity (all at p < 0.001). SC-1, surface coil and one excitation; SC-4, surface coil and four excitations; VC-1, volume coil and one excitation; VC-4, volume coil and four excitations

Comparisons of DTI parameters in vivo and ex vivo

The comparisons of DTI parameters in vivo and ex vivo were performed with the surface coil and one excitation. The diffusivities of ex vivo ROIs were significantly lower than those of in vivo ROIs (all at p < 0.001; Table 1). The reduction of MD, AD, and RD after fixation is summarised in Table 2. Reduction of all the ROIs ranged from 0.54 to 0.70 in MD, from 0.51 to 0.70 in AD, and from 0.54 to 0.72 in RD after fixation (Table 2). The MD, AD, and RD in WM integrity decreased more than those in GM integrity after fixation (WM integrity versus GM integrity, 0.66 ± 0.05 versus 0.51 ± 0.16 in MD; 0.66 ± 0.06 versus 0.54 ± 0.06 in AD; 0.65 ± 0.06 versus 0.56 ± 0.05 in RD; all at p < 0.001). After fixation, dimensionless DTI parameters including FA, R1, and R2 in the major brain structures did not change significantly, whereas the R3 in the major brain structures changed significantly (Table 3).

Table 1 Diffusivities of in vivo and ex vivo regions of interest using the surface coil and one excitation
Table 2 Reduction of diffusivities in selected regions of interest after fixation, using the surface coil and one excitation
Table 3 Dimensionless DTI parameters of in vivo and ex vivo regions of interest using the surface coil and one excitation

Comparisons of ex vivo DTI parameters under different scan conditions

Figure 2 and Table 4 show the comparisons of ex vivo diffusivities after using different coils and number of excitations. There were significant differences in the MD of major brain structures using the surface coil compared to the volume coil for four excitations. There were significant differences in the MD of major brain structures with one compared to four excitations for both coils. Also there were significant differences between the AD of major brain structures obtained using the surface coil and those obtained using the volume coil, and differences between brain structures imaged with one compared to four excitations. The RD of major brain ROIs with one excitation was significantly different from that with four excitations.

Fig. 2
figure 2

Comparisons of ex vivo diffusivities using the surface coil and the volume coil, and using one and four excitations. Most diffusivities in major brain structures differed when using different coils and numbers of excitations. AC, anterior commissure; AD, axial diffusivity; CC, corpus callosum; CT, cortex; EC, external capsule; HP, hippocampus; ICCP, internal capsule and cerebral peduncle; MD, mean diffusivity; ns, not significant; OT, optic tract; RD, radial diffusivity; ST, striatum; SC-1, surface coil and one excitation; SC-4, surface coil and four excitations; VC-1, volume coil and one excitation; VC-4, volume coil and four excitations. *p < 0.05; **p < 0.01; ***p < 0.001

Table 4 Comparisons of DTI parameters and SNR using different coils and different numbers of excitations

Figure 3 and Table 4 show the comparisons of ex vivo dimensionless parameters after using different coils and numbers of excitations. The FA of major brain ROIs using the surface coil was significantly different from that using the volume coil for one excitation. For the surface coil, there were significant differences in the FA of major brain structures using one compared to four excitations. There were significant differences in R1 using one compared to four excitations using the surface coil. There were significant differences in R2 of major brain structures using one compared to four excitations using the surface coil. The R2 of major brain ROIs using the surface coil was significantly different from that using the volume coil for one excitation. The R3 of major brain ROIs using the surface coil was significantly different from that using the volume coil for one excitation. There were significant differences in R3 of brain structures using one compared to four excitations and the surface coil.

Fig. 3
figure 3

Comparisons of ex vivo dimensionless diffusion tensor imaging (DTI) parameters using the surface coil and the volume coil, and using one and four excitations. Most dimensionless DTI parameters in major brain structures differed using different coils and numbers of excitations. AC, anterior commissure; AD, axial diffusivity; CC, corpus callosum; CT, cortex; EC, external capsule; HP, hippocampus; ICCP, internal capsule and cerebral peduncle; MD, mean diffusivity; ns, not significant; OT, optic tract; RD, radial diffusivity; ST, striatum; SC-1, surface coil and one excitation; SC-4, surface coil and four excitations; VC-1, volume coil and one excitation; VC-4, volume coil and four excitations. *p < 0.05; **p < 0.01; ***p < 0.001

There were significant differences in the SNR of brain structures using the surface coil compared to the volume coil, and with one compared to four excitations (Fig. 4 and Table 4). With the same coil, the SNRs of all the selected ROIs were significantly greater with four excitations than with one excitation. The SNR of brain structures was significantly greater using the volume coil than using the surface coil when the same number of excitations was selected.

Fig. 4
figure 4

Comparisons of signal-to-noise (SNR) using the surface coil and the volume coil, and using one and four excitations. There were significant differences in the SNR of brain structures using the surface coil compared to the volume coil, and using one compared to four excitations. AC, anterior commissure; AD, axial diffusivity; CC, corpus callosum; CT, cortex; EC, external capsule; HP, hippocampus; ICCP, internal capsule and cerebral peduncle; MD, mean diffusivity; ns, not significant; OT, optic tract; RD, radial diffusivity; ST, striatum; SC-1, surface coil and one excitation; SC-4, surface coil and four excitations; VC-1, volume coil and one excitation; VC-4, volume coil and four excitations. *p < 0.05; **p < 0.01; ***p < 0.001

Discussion

In terms of the degree of WM and GM, our study showed that the FA of WM integrity was greater than that of GM integrity, in agreement with previous research [16, 20]. Interestingly, the results for the RD, R1, and R2 were opposite to those for the FA, and the AD in WM integrity and GM integrity was similar, suggesting that the RD constitutes the main difference in diffusion between WM integrity and GM integrity.

The diffusivities of brain WM ROIs and GM ROIs decreased after fixation, which is consistent with previous studies [13, 14, 16]. The principle of fixation is related to the formation of cross-links between proteins or between proteins and nucleic acids, which involves hydroxymethylene bridges and coordinate bonds for calcium ions, altering the three-dimensional structure of proteins [21,22,23,24]. Changes in the molecular structure may affect the molecular diffusivities. Tissue shrinkage after fixation may also contribute to diffusivity alterations [25,26,27]. Fixation processes and fixatives may also affect the change in tissue property [20, 28]. Additionally, previous work has found that diffusion is associated with temperature [29,30,31,32]. A 2.4% alteration in water diffusion per degree has been determined [31]. The temperature of the magnetic resonance room is 20 °C, whereas the rat body temperature is 37 °C. Furthermore, magnetic resonance tissue characteristics including T1 and T2 relaxations have been proved to tend to decrease after fixation [33,34,35,36].

The present study demonstrated that WM diffusivities decreased more than GM diffusivities, which is in accordance with the results of the previous investigations [13, 14]. However, what causes this diversified reduction in the diffusivity of WM and GM remains to be investigated. McGrath et al. [37] have confirmed the different effects of fixation on different tissues. Therefore, the different components of WM and GM may also cause differences in the reduction of diffusivity [13].

For dimensionless DTI parameters, we observed a significant difference in vivo compared to ex vivo only in R3, which may be related to the non-proportional reduction of λ2 and λ3. A better understanding of the details of diffusion change in the microstructure in different directions may be promoted by diffusion kurtosis imaging or multi compartment models [38,39,40,41]. In terms of FA, the current study found that anisotropies of the major brain ROIs were similar in the live and in the fixed brain, in agreement with previous investigations [13,14,15].

There are several limitations in this study. First, there were no in vivo brain data obtained using the volume coil and four excitations, due to the compromised volume coil size and the scanning duration. Second, a slice thickness of 1 mm is probably too thick for ex vivo scanning. Given the weak tolerance of the scanning duration in experimental models, we chose 1 mm to compare in vivo and ex vivo DTI data. Third, it is difficult to observe the histological appearance of the brain both in vivo and ex vivo.

In summary, we observed significant effects of the fixation, coil, and number of excitations on major DTI parameters and the SNR of magnetic resonance studies of WM and GM in a rat model imaged at 7.0 T. As a consequence, we suggest that the fixation, coil, and number of excitations should be taken into account in clinical and experimental DTI brain studies.

Abbreviations

AD:

Axial diffusivity

DTI:

Diffusion tensor imaging

FA:

Fractional anisotropy

GM:

Grey matter

MD:

Mean diffusivity

R1:

λ2/λ1 ratio

R2:

λ3/λ1 ratio

R3:

λ2/λ3 ratio

RD:

Radial diffusivity

ROI:

Region of interest

SNR:

Signal-to-noise ratio

WM:

White matter

References

  1. Kumar M, Duda JT, Yoon SY et al (2016) Diffusion tensor imaging for assessing brain gray and white matter abnormalities in a feline model of alpha-mannosidosis. J Neuropathol Exp Neurol 75:35–43

    Article  PubMed  CAS  Google Scholar 

  2. Badea A, Kane L, Anderson RJ et al (2016) The fornix provides multiple biomarkers to characterize circuit disruption in a mouse model of Alzheimer's disease. Neuroimage 142:498–511

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  3. Zhang J, Aggarwal M, Mori S (2012) Structural insights into the rodent CNS via diffusion tensor imaging. Trends Neurosci 35:412–421

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  4. Basser PJ, Mattiello J, LeBihan D (1994) MR diffusion tensor spectroscopy and imaging. Biophys J 66:259–267

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  5. Sun SW, Liang HF, Trinkaus K, Cross AH, Armstrong RC, Song SK (2006) Noninvasive detection of cuprizone induced axonal damage and demyelination in the mouse corpus callosum. Magn Reson Med 55:302–308

    Article  PubMed  Google Scholar 

  6. Budde MD, Kim JH, Liang HF et al (2007) Toward accurate diagnosis of white matter pathology using diffusion tensor imaging. Magn Reson Med 57:688–695

    Article  PubMed  Google Scholar 

  7. Tuor UI, Morgunov M, Sule M et al (2014) Cellular correlates of longitudinal diffusion tensor imaging of axonal degeneration following hypoxic-ischemic cerebral infarction in neonatal rats. Neuroimage Clin 6:32–42

    Article  PubMed  PubMed Central  Google Scholar 

  8. Morales-Guadarrama A, Mejía-González I, Hernández-Godínez B, Ibáñez-Contreras A, Reyes-Pantoja SA, Olayo R (2013) DTI tractography and MRI in chronic cerebral ischemia in rhesus monkey (Macaca mulatta) in vivo. Arq Neuropsiquiatr 71:134

    Article  PubMed  Google Scholar 

  9. Yuh EL, Cooper SR, Mukherjee P et al (2014) Diffusion tensor imaging for outcome prediction in mild traumatic brain injury: a TRACK-TBI study. J Neurotrauma 31:1457–1477

    Article  PubMed  PubMed Central  Google Scholar 

  10. Laitinen T, Sierra A, Pitkänen A, Grohn O (2010) Diffusion tensor MRI of axonal plasticity in the rat hippocampus. Neuroimage 51:521–530

    Article  PubMed  Google Scholar 

  11. Ruest T, Holmes WM, Barrie JA et al (2011) High-resolution diffusion tensor imaging of fixed brain in a mouse model of Pelizaeus-Merzbacher disease: comparison with quantitative measures of white matter pathology. NMR Biomed 24:1369–1379

    Article  PubMed  Google Scholar 

  12. Chuang N, Mori S, Yamamoto A et al (2011) An MRI-based atlas and database of the developing mouse brain. Neuroimage 54:80–89

    Article  PubMed  Google Scholar 

  13. Sun SW, Neil JJ, Song SK (2003) Relative indices of water diffusion anisotropy are equivalent in live and formalin-fixed mouse brains. Magn Reson Med 50:743–748

    Article  PubMed  Google Scholar 

  14. Wu D, Xu J, McMahon MT et al (2013) In vivo high-resolution diffusion tensor imaging of the mouse brain. Neuroimage 83:18–26

    Article  PubMed  Google Scholar 

  15. Guilfoyle DN, Helpern JA, Lim KO (2003) Diffusion tensor imaging in fixed brain tissue at 7.0 T. NMR Biomed 16:77–81

    Article  PubMed  Google Scholar 

  16. Rane S, Duong TQ (2011) Comparison of in vivo and ex vivo diffusion tensor imaging in rhesus macaques at short and long diffusion times. Open Neuroimage J 5:172–178

    Article  Google Scholar 

  17. Hui ES, Cheung MM, Chan KC, Wu EX (2010) B-value dependence of DTI quantitation and sensitivity in detecting neural tissue changes. Neuroimage 49:2366–2374

    Article  PubMed  Google Scholar 

  18. van Meer MP, Otte WM, van der Marel K et al (2012) Extent of bilateral neuronal network reorganization and functional recovery in relation to stroke severity. J Neurosci 32:4495–4507

    Article  PubMed  CAS  Google Scholar 

  19. Edelstein WA, Bottomley PA, Pfeifer LM (1984) A signal-to-noise calibration procedure for NMR imaging systems. Med Phys 11:180–185

    Article  PubMed  CAS  Google Scholar 

  20. Madi S, Hasan KM, Narayana PA (2005) Diffusion tensor imaging of in vivo and excised rat spinal cord at 7 T with an icosahedral encoding scheme. Magn Reson Med 53:118–125

    Article  PubMed  Google Scholar 

  21. Fox CH, Johnson FB, Whiting J, Roller PP (1985) Formaldehyde fixation. J Histochem Cytochem 33:845–853

    Article  PubMed  CAS  Google Scholar 

  22. Helander KG (1994) Kinetic studies of formaldehyde binding in tissue. Biotech Histochem 69:177–179

    Article  PubMed  CAS  Google Scholar 

  23. Morgan JM, Navabi H, Schmid KW, Jasani B (1994) Possible role of tissue-bound calcium ions in citrate-mediated high-temperature antigen retrieval. J Pathol 174:301–307

    Article  PubMed  CAS  Google Scholar 

  24. Werner M, Chott A, Fabiano A, Battifora H (2000) Effect of formalin tissue fixation and processing on immunohistochemistry. Am J Surg Pathol 24:1016–1019

    Article  PubMed  CAS  Google Scholar 

  25. Hillman H, Deutsch K (1978) Area changes in slices of rat brain during preparation for histology or electron microscopy. J Microsc 114:77–84

    Article  PubMed  CAS  Google Scholar 

  26. Quester R, Schröder R (1997) The shrinkage of the human brain stem during formalin fixation and embedding in paraffin. J Neurosci Methods 75:81–89

    Article  PubMed  CAS  Google Scholar 

  27. Schulz G, Crooijmans HJ, Germann M, Scheffler K, Muller-Gerbl M, Müller B (2011) Three-dimensional strain fields in human brain resulting from formalin fixation. J Neurosci Methods 2:17–27

    Article  Google Scholar 

  28. Wehrl HF, Bezrukov I, Wiehr S et al (2015) Assessment of murine brain tissue shrinkage caused by different histological fixatives using magnetic resonance and computed tomography imaging. Histol Histopathol 30:601–613

    PubMed  CAS  Google Scholar 

  29. Morvan D, Leroy-Willig A (1995) Simultaneous measurements of diffusion and transverse relaxation in exercising skeletal muscle. Magn Reson Imaging 13:943–948

    Article  PubMed  CAS  Google Scholar 

  30. Thelwall PE, Shepherd TM, Stanisz GJ, Blackband SJ (2006) Effects of temperature and aldehyde fixation on tissue water diffusion properties, studied in an erythrocyte ghost tissue model. Magn Reson Med 56:282–289

    Article  PubMed  Google Scholar 

  31. Le Bihan D (1995) Diffusion and perfusion magnetic resonance imaging-applications to functional MRI. Raven Press, New York

    Google Scholar 

  32. Hasegawa Y, Latour LL, Sotak CH, Dardzinski BJ, Fisher M (1994) Temperature dependent change of apparent diffusion coefficient of water in normal and ischemic brain of rats. J Cereb Blood Flow Metab 14:383–390

    Article  PubMed  CAS  Google Scholar 

  33. Tovi M, Ericsson A (1992) Measurements of T1 and T2 over time in formalin-fixed human whole-brain specimens. Acta Radiol 33:400–404

    Article  PubMed  CAS  Google Scholar 

  34. Yong-Hing CJ, Obenaus A, Stryker R, Tong K, Sarty GE (2005) Magnetic resonance imaging and mathematical modeling of progressive formalin fixation of the human brain. Magn Reson Med 54:324–332

    Article  PubMed  Google Scholar 

  35. Purea A, Webb AG (2006) Reversible and irreversible effects of chemical fixation on the NMR properties of single cells. Magn Reson Med 56:927–931

    Article  PubMed  Google Scholar 

  36. Raman MR, Shu Y, Lesnick TG, Jack CR, Kantarci K (2016) Regional T relaxation time constants in ex vivo human brain: longitudinal effects of formalin exposure. Magn Reson Med 77:774–778

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  37. McGrath DM, Foltz WD, Al-Mayah A, Niu CJ, Brock KK (2012) Quasi-static magnetic resonance elastography at 7 T to measure the effect of pathology before and after fixation on tissue biomechanical properties. Magn Reson Med 68:152–165

    Article  PubMed  Google Scholar 

  38. Rosenkrantz AB, Padhani AR, Chenevert TL et al (2015) Body diffusion kurtosis imaging: basic principles, applications, and considerations for clinical practice. J Magn Reson Imaging 42:1190–1202

    Article  PubMed  Google Scholar 

  39. Chung S, Fieremans E, Kucukboyaci NE et al (2018) Working memory and brain tissue microstructure: white matter tract integrity based on multi-Shell diffusion MRI. Sci Rep 8:3175

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  40. Ferizi U, Schneider T, Panagiotaki E et al (2014) A ranking of diffusion MRI compartment models with in vivo human brain data. Magn Reson Med 72:1785–1792

    Article  PubMed  Google Scholar 

  41. Zhu X, Gur Y, Wang W, Fletcher PT (2013) Model selection and estimation of multi-compartment models in diffusion MRI with a Rician noise model. Inf Process Med Imaging 23:644–655

    Article  PubMed  Google Scholar 

Download references

Availability of data and materials

All data generated or analysed during this study are included in this published article.

Funding

No specific funding was received for this article.

Acknowledgements

The authors would like to acknowledge Lei Wang for the assistance in the animal experiment.

Author information

Authors and Affiliations

Authors

Contributions

CW: design of the study, realisation of the experiment, data analysis and interpretation, and manuscript preparation. LS: realisation of the experiment, data analysis, and statistical analysis. RZ: realisation of the experiment, manuscript revision. FG: design and coordination of the study, data analysis and interpretation, and manuscript revision. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Fabao Gao.

Ethics declarations

Ethics approval

The study was approved by local Experimental Animal Ethics Committee on 25 February 2016. The approval number is 2016005A.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wang, C., Song, L., Zhang, R. et al. Impact of fixation, coil, and number of excitations on diffusion tensor imaging of rat brains at 7.0 T. Eur Radiol Exp 2, 25 (2018). https://doi.org/10.1186/s41747-018-0057-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-018-0057-2

Keywords