Skip to main content

Impact of reconstruction parameters on the accuracy of myocardial extracellular volume quantification on a first-generation, photon-counting detector CT

Abstract

Background

The potential role of cardiac computed tomography (CT) has increasingly been demonstrated for the assessment of diffuse myocardial fibrosis through the quantification of extracellular volume (ECV). Photon-counting detector (PCD)-CT technology may deliver more accurate ECV quantification compared to energy-integrating detector CT. We evaluated the impact of reconstruction settings on the accuracy of ECV quantification using PCD-CT, with magnetic resonance imaging (MRI)-based ECV as reference.

Methods

In this post hoc analysis, 27 patients (aged 53.1 ± 17.2 years (mean ± standard deviation); 14 women) underwent same-day cardiac PCD-CT and MRI. Late iodine CT scans were reconstructed with different quantum iterative reconstruction levels (QIR 1−4), slice thicknesses (0.4−8 mm), and virtual monoenergetic imaging levels (VMI, 40−90 keV); ECV was quantified for each reconstruction setting. Repeated measures ANOVA and t-test for pairwise comparisons, Bland–Altman plots, and Lin’s concordance correlation coefficient (CCC) were used.

Results

ECV values did not differ significantly among QIR levels (p = 1.000). A significant difference was observed throughout different slice thicknesses, with 0.4 mm yielding the highest agreement with MRI-based ECV (CCC = 0.944); 45-keV VMI reconstructions showed the lowest mean bias (0.6, 95% confidence interval 0.1–1.4) compared to MRI. Using the most optimal reconstruction settings (QIR4. slice thickness 0.4 mm, VMI 45 keV), a 63% reduction in mean bias and a 6% increase in concordance with MRI-based ECV were achieved compared to standard settings (QIR3, slice thickness 1.5 mm; VMI 65 keV).

Conclusions

The selection of appropriate reconstruction parameters improved the agreement between PCD-CT and MRI-based ECV.

Relevance statement

Tailoring PCD-CT reconstruction parameters optimizes ECV quantification compared to MRI, potentially improving its clinical utility.

Key points

• CT is increasingly promising for myocardial tissue characterization, assessing focal and diffuse fibrosis via late iodine enhancement and ECV quantification, respectively.

• PCD-CT offers superior performance over conventional CT, potentially improving ECV quantification and its agreement with MRI-based ECV.

• Tailoring PCD-CT reconstruction parameters optimizes ECV quantification compared to MRI, potentially improving its clinical utility.

Graphical Abstract

Background

The detection of myocardial scar and fibrosis has diagnostic and prognostic relevance, especially since the introduction of antifibrotic therapies [1, 2]. Magnetic resonance imaging (MRI) is the current noninvasive reference standard for myocardial viability assessment [3], typically based on late gadolinium enhancement, allowing the assessment of focal fibrosis and scars, as well as T1 mapping sequences, which allow the assessment of diffuse, reactive, and interstitial fibrosis through the quantification of extracellular volume (ECV) [4]. Over the last decade, CT has increasingly demonstrated its potential for myocardial tissue characterization, primarily for the evaluation of focal fibrosis through late iodine enhancement (LIE) and later for the assessment of diffuse fibrosis using ECV quantification [5, 6]. CT-based scar detection and ECV quantification are not yet routinely applied, even though their value has already been demonstrated in various clinical scenarios [7,8,9].

The strength of CT-based ECV quantification relies on its high inter- and intra-observer reproducibility and the ability to visualize coronary anatomy during the same session [10]. Dual-energy CT can further improve ECV assessment compared to single‐energy by enabling the creation of iodine maps. Iodine maps automatically and accurately quantify ECV, eliminating the need for true noncontrast acquisitions, thus reducing radiation exposure [11]. Conventional energy-integrating detectors (EID)-CT, however, demonstrate a slight systematic overestimation of ECV compared to MRI [12].

The recently introduced photon-counting detector (PCD)-CT technique offers several advantages over EID-CT such as the availability of spectral data without temporal resolution penalty and improved spatial and contrast resolution [13]. These advancements have the potential to improve ECV-based myocardial tissue characterization, which has been demonstrated to correlate strongly with MRI-ECV when using EID-CT [14, 15]. We hypothesized that an optimal combination of reconstruction parameters improves ECV quantification accuracy and can reduce such overestimation.

Therefore, the aim of this study was to evaluate the impact of CT image reconstruction settings on the accuracy of ECV quantification using a first-generation PCD-CT, with MRI-ECV as reference.

Methods

Study population

This post hoc analysis was performed based on a previously published prospective study [15]. The protocol of the prospective, observational, single-center, Health Insurance Portability and Accountability Act–compliant study was approved by the local institutional review board (Number: Pro00108359). Written informed consent was obtained for all participants. From July 2021 to January 2022, consecutive patients undergoing a clinically indicated cardiac MRI were enrolled and underwent a research cardiac CT on the same day. Inclusion criteria were (i) age ≥ 18 years and (ii) clinical indication for cardiac MRI. Exclusion criteria were (i) refusal to consent and (ii) contraindication to iodine-based contrast media (positive anamnesis of allergy to iodinated contrast media or impaired renal function, defined as a creatinine level > 1.5 mg/dL). Demographic information, medical history, and laboratory values were extracted from electronic medical records.

MRI acquisition protocol

Cardiac MRI was performed on a 1.5-T system (MAGNETOM Avanto; Siemens Healthineers, Erlangen, Germany) using a dedicated electrocardiogram (ECG)-gated acquisition protocol including myocardial native and postcontrast T1 mapping. T1 mapping images of the left ventricle (LV) were acquired in a two-chamber short-axis view at basal, midventricular, and apical positions using a modified Look-Locker (MOLLI) inversion-recovery sequence with a 5(3)3 sampling scheme with the following pulse sequence parameters: repetition time/echo time 2.6/1.1 ms; field of view 300 × 256 mm2; slice thickness 8 mm; image acquisition matrix 192 × 128; reconstruction matrix 192 × 164; in-plane spatial resolution 1.56 × 1.56 mm2; bandwidth 1,085 Hz/pixel; flip angle 35°; and parallel imaging acceleration factor. T1 mapping acquisition was repeated at the equilibrium phase at 10–12 min after the administration of 0.1 mmol/kg gadobutrol (Gadavist; Bayer Healthcare, Berlin, Germany), at the same slice positions as for native T1 mapping, using a MOLLI 4(1)3(1)2 sampling scheme.

MRI postprocessing and ECV analysis

All cardiac MRI examinations were analyzed using dedicated cardiac software (Circle cvi42 v.5.12.2, Circle Cardiovascular Imaging, Calgary, Canada). The LV myocardial wall and the intra-ventricular blood pool were segmented on native and post-contrast T1 maps. The two maps were then coregistered using a contour-based registration method, and ECV was calculated as follows:

$$\mathrm{ECV }= (1 -\mathrm{ Hct}) \times \frac{\left(\frac{1}{{{\text{T}}1}_{\mathrm{myo\ post}}} - \frac{1}{{{\text{T}}1}_{\mathrm{myo\ pre}}}\right)}{\left(\frac{1}{{{\text{T}}1}_{\mathrm{blood\ post }}} - \frac{1}{{{\text{T}}1}_{\mathrm{blood\ pre}}}\right)}$$

where Hct is the hematocrit, T1myo pre and T1myo post are native and postcontrast T1 measurements in the myocardium, respectively, while T1blood pre and T1blood post are native and post-contrast T1 measurements in the LV blood compartment, respectively.

Partial volume averaging due to intraventricular blood and epicardial fat was mitigated by setting an automatic offset of 25% from the LV endocardial and epicardial borders. Since laboratory-based Hct measurements were not available for all patients, a synthetic Hct value was calculated from the native T1 maps following a locally established formula [16]:

$$\mathrm{Synthetic\, Hct}= \left(898\times \left[\frac{1}{{{\text{T}}1}_{{\text{blood}}}}\right]\right)- 0.16$$

where T1blood is the native T1 measured at the LV.

CT acquisition protocol

All CT scans were performed on a first-generation dual-source PCD-CT system (NAEOTOM Alpha; Siemens Healthineers, Forchheim, Germany) equipped with two cadmium telluride PCDs, each with a 144 × 0.4-mm collimation. The gantry rotation time was 0.25 s. Tube voltage was set at 120 kVp per vendor recommendations, while the tube current was automatically adjusted to reach the chosen image quality level (CARE Dose4D, Siemens Healthineers). Coronary CT angiography (CCTA) was performed after the intravenous administration of 100 mL iopromide (Ultravist, 370 mg I/mL; Bayer Healthcare, Berlin, Germany) at a flow rate of 5–6 mL/s along with a 20-mL saline chaser. The image quality level for automated tube voltage selection (CARE kV, Siemens Healthineers) for the CCTA was set to 44 (equivalent to a standard CCTA scan), and an ECG-triggered window was set from 30 to 80% of the RR interval. Scan parameters were aligned to a recently published protocol [14]. LIE scans were acquired using a prospective ECG-triggered protocol with a fixed 280-ms delay from the R-wave, 5 min after contrast administration. CARE kV was set to an image quality level of 55, equivalent to a standard LIE cardiac scan. All scans were performed using a standard Quantum Plus (Siemens Healthineers) acquisition mode, which allows spectral imaging-based iodine map reconstructions.

CT image reconstruction and ECV analysis

CCTA and LIE scans were reconstructed with the use of a proprietary offline image reconstruction platform (ReconCT, version 15.0.58757.0, Siemens Healthineers). The LIE scans were used to generate iodine maps. ECV analysis was performed using a prototype software (CT Cardiac Functional Analysis, version 2.0.9, Siemens Healthineers), as follows: in a preprocessing step, CCTA, LIE-based virtual monoenergetic images (VMI) and iodine maps were aligned by nonrigid registration with the reference coordinate frame of the CCTA to enable accurate segmentation of the myocardium and blood pool. For the ECV quantification, iodine maps associated with the LIE scan contained all the information needed, and ECV was calculated as follows:

$$\mathrm{ECV }= (1-{\text{Hct}}) \times \frac{\mathrm{Iodine\, density\, myocardium}}{\mathrm{Iodine\, density\, bloodpool}}$$

The same synthetic Hct value was used as for the MRI-based ECV quantification. For quantitative analysis, the mid-myocardial wall was considered, using the same 25% offset as in MRI. ECV was calculated for the entire LV volume, and the American Heart Association polar maps with 17 myocardial segments were automatically computed and displayed [17].

To find the optimal combination of reconstruction parameters with the best concordance with MRI, a step-by-step iterative process was chosen in which only one CT reconstruction parameter was modified at a time. ECV was calculated each time one of the reconstruction parameters was changed. The baseline reconstruction setting was the same used in current literature for ECV quantification on a PCD-CT scanner [14, 15]: slice thickness 1.5 mm; increment 1 mm; quantum iterative reconstruction (QIR) 3; kernel Qr40; and LIE VMI reconstructed at 65 keV. In Step 1, the QIR level was modified, obtaining four sets of images ranging from QIR 1 to QIR 4. In Step 2, different slice thicknesses/increments were tested, namely: 0.4/0.3, 1.5/1.0, 3.0/2.5, 5.0/4.0, and 8.0/7.0 mm. In Step 3, VMI levels of the LIE scan were adjusted, including 40, 45, 50, 55, 60, 65, 70, and 90 keV. The parameter with the best tradeoff between the highest concordance and lowest bias compared to MRI was retained in each step for the subsequent step. A visual representation of this step-by-step approach is illustrated in Fig. 1.

Fig. 1
figure 1

Representation of the systematic step-by-step protocol optimization approach. In Step 1, the optimal quantum iterative reconstruction level (14) was identified. In Step 2, the different slice thicknesses were tested from 0.4 to 8 mm. In Step 3, the virtual monoenergetic image reconstruction level of the late iodine enhancement scan was optimized from 40 to 90 keV. The parameter with the best diagnostic performance in each step was retained in the subsequent step

Statistical analysis

Statistical analyses were performed with MedCalc, version 20.110 (MedCalc Software Ltd, Ostend, Belgium). Normality of data was assessed using the Shapiro–Wilk test. Categorical variables were reported as absolute values and percentages, while continuous variables as mean ± standard deviation. At each step, the CT-based ECV quantifications derived from different combinations of reconstruction parameters were compared using the ANOVA test for repeated measures and the t-test for pairwise comparisons. Lin’s concordance correlation coefficient (CCC) was calculated to measure the agreement among the different CT-based ECV values and the MRI-based ECV, which was used as reference standard. CCC values were interpreted as follows: correlation < 0.200 = poor agreement; 0.200−0.399 = fair agreement; 0.400−0.599 = moderate agreement; 0.600−0.799 = good agreement; and 0.800−1.000 = excellent agreement [18]. Bland–Altman plots were used to visualize the agreement between CT- and MRI-based ECV quantification with ± 1.96 standard deviations limits of agreement (LoA) and to identify bias between the two measurements. No data imputation for missing values was used. All tests were two-tailed, and a p-value < 0.05 was considered statistically significant. Bonferroni correction was applied when multiple testing was performed.

Results

Population characteristics

Twenty-seven patients were enrolled in this study and underwent cardiac MRI and CT on the same day. Demographic and clinical characteristics, as well as ECV quantification with MRI and baseline CT, are detailed in Table 1. The mean global MRI-based ECV was 31.1 ± 5.9% and was available only in 21/27 (77.8%) patients since not all subjects had all three myocardial slices available with adequate image quality. The baseline mean global CT-based ECV was 33.2 ± 4.7% and was available in all patients.

Table 1 Baseline patient characteristics

Step 1: QIR optimization

Statistical analysis did not show any significant difference among CT-based ECV values when changing the QIR strength level. Although ECV values were comparable, a slight tendency for better quantification accuracy was observed when increasing the QIR level, as demonstrated by the CCC and the mean bias in the Bland–Altman plot. The concordance between CT- and MRI-based ECV was 0.87 for images reconstructed with a QIR 1, and 0.89 for images reconstructed with a QIR 4; the mean bias was 1.7 (95% confidence interval (CI) 0.8–2.6, LoA 2.3–5.8) with QIR 1, and 1.6 (95% CI 0.7–2.5, LoA -2.2–5.5) using QIR 4. In all the cases, CT demonstrated a systematic overestimation of ECV with respect to MRI.

Step 2: slice thickness optimization

For Step 2, QIR 4, the QIR level showing the highest agreement and lowest mean bias in Step 1, and VMI at 65 keV were kept constant, while slice thickness/increment were adjusted. The combination of reconstruction parameters with slice thickness 0.4 mm and increment 0.3 mm demonstrated a significant difference compared to the other slice thickness combinations, as well as the highest CCC (0.94, 95% CI 0.87–0.97) and lowest mean bias (0.8, 95% CI 0.0–1.5, LoA -2.5–4.1). For thicker slices, we observed a reduced concordance and a higher mean bias compared to the MRI-based ECV. Table 2 shows a detailed pairwise comparison among the different slice thicknesses. Figure 2 shows the concordance scatterplot diagram and the Bland–Altman plot for the 0.4 mm slice thickness reconstruction.

Table 2 Step 2: computed tomography-based myocardial extracellular volume (ECV) quantification using quantum iterative reconstruction level 4 and different slice thicknesses
Fig. 2
figure 2

Correlation of ECV between CT and MRI using 0.4-mm slice thickness. Scatterplot (a) and Bland–Altman plot (b) showing the global ECV correlation between CT and MRI when using QIR 4, slice thickness/increment 0.4/0.3 mm, and VMI 65 keV. Horizontal lines are drawn at the mean difference (blue), and the limits of agreement (red); in the graph are also displayed the line of equality (orange horizontal line), the regression line of differences (pink line), and the 95% confidence interval of mean difference (green bar). CCC Concordance correlation coefficient, CT Computed tomography, ECV Extracellular volume, MRI Magnetic resonance imaging, QIR Quantum iterative reconstruction, VMI Virtual monoenergetic image

Step 3: VMI optimization

The last reconstruction parameter tested was the VMI level of the LIE scans: different VMIs were combined with a slice thickness of 0.4 mm and a QIR level of 4 based on conclusions from Steps 1 and 2. The t-test for paired samples showed a significant difference among the lower keV levels, particularly in the range of 45−60 keV, compared to 70 and 90 kV, as detailed in Table 3. Likewise, we noticed a reduction in the mean bias for the VMIs ranging from 45 to 60 keV, with the lowest bias achieved at 45 keV (mean bias 0.6, 95% CI -0.1–1.4, LoA -2.9–4.2). For 40 keV and any keV higher than 60, the mean bias was higher, indicating an overestimation of CT-based ECV over MRI-based ECV. The CCC showed excellent agreement for all the tested VMIs, ranging from 0.93 for 45 keV to 0.94 for 40 keV.

Table 3 Step 3: computed tomography-based myocardial extracellular volume (ECV) quantification using quantum iterative reconstruction level 4 and 0.4-mm slice thickness combined with different virtual monoenergetic images (VMIs)

Hence, according to the stepwise analysis, the best combination of reconstruction parameters was reached when using QIR 4, slice thickness of 0.4 mm with 0.3 mm increment, and 45-keV VMI level. Figure 3 shows the concordance scatterplot and the Bland–Altman plot for the most optimal combination. This combination, compared to the baseline ECV quantification, showed a reduction in the mean bias by 63% and a 6% increase in the agreement with MRI. Figure 4 demonstrates a case example comparing CT-based ECV quantified with the baseline reconstruction parameters, with the most optimal combination based on our results, and MRI-based ECV.

Fig. 3
figure 3

Correlation of ECV between CT and MRI with the best-found combination. Scatterplot (a) and Bland–Altman plot (b) showing the global ECV correlation between the most optimal combination of CT reconstruction parameters and MRI. The CT parameters were QIR 4, slice thickness 0.4 mm, increment 0.3 mm, and VMI 45 keV. Horizontal lines are drawn at the mean difference (blue), and the limits of agreement (red); in the graph are also displayed the line of equality (orange horizontal line), the regression line of differences (pink line), and the 95% confidence interval of mean difference (green bar). CCC Concordance correlation coefficient, CT Computed tomography, ECV Extracellular volume, MRI Magnetic resonance imaging, QIR Quantum iterative reconstruction, VMI Virtual monoenergetic image

Fig. 4
figure 4

Case of a patient with a history of ischemic cardiomyopathy. CT-based ECV quantification shows anteroseptal 25−50% thickness subendocardial scar and near-transmural apical scar. The CT-based ECV quantification based on the current literature parameters (a, QIR 3, slice thickness 1.5 mm, and VMI 65 keV) overestimates the global ECV compared to the quantification performed with the most optimal combination of parameters found in our study (b, QIR 4, slice thickness 0.4 mm, and VMI 45 keV), with global ECV values of 40.7% and 38.4%, respectively, versus MRI-based ECV of 38.7% (c, midventricular MRI-based ECV map)

Discussion

This post hoc analysis of a prospective study investigated the impact of reconstruction parameters on the accuracy of ECV quantification using PCD-CT and the concordance of CT-based ECV with the reference standard MRI-based ECV. The study revealed three major findings.

First, we demonstrated a good-to-excellent correlation between CT- and MRI-based ECV at every investigated QIR strength level, slice thickness, and keV level, highlighting the robustness of CT-based ECV quantification. Second, we demonstrated that adjusting these parameters improves the quantification accuracy of CT-based ECV compared to MRI. In particular, we were able to reduce the mean bias by 63% and increase the concordance with MRI by 6%, when comparing the optimal combination of reconstruction parameters to the baseline proposed in the literature [14, 15]. Third, our systematic methodology enabled the identification of slice thickness as one of the primary parameters significantly influencing ECV quantification and reliability. We chose this systematic step-by-step approach to observe the specific effects of each reconstruction parameter independently and understand the individual contribution to the overall outcome through incremental learning.

The first parameter we evaluated was the strength level of QIR, the iterative reconstruction algorithm specifically introduced for PCD-CT as the commercially available iterative reconstruction algorithms used for EID-CT are not optimal for PCD-CT imaging [19]. In the analyzed scans, QIR strength did not significantly impact the CT-based ECV quantification. Nonetheless, we could notice a slight trend towards increased concordance and decreased mean bias compared to MRI when using higher QIRs. This observation indicates that, unlike conventional iterative reconstruction algorithms, QIR does not affect low-contrast spatial resolution performance, as typical in LIE scan, and does not influence ECV quantification [20]. While conventional iterative reconstruction techniques affect the spatial resolution of low-contrast scans [21], high QIR levels even improve soft-tissue image quality by reducing noise and enhancing contrast-to-noise ratio (CNR) without compromising image texture or CT attenuation values [20].

The second step of our study demonstrated superior accuracy and concordance of CT-based ECV when a thinner slice, 0.4 mm with 0.3 mm increment, was used. This result can be explained by the significant reduction in partial volume averaging, which can lead to significant errors in ECV quantification. When reducing the slice thickness with EID-CT, it should be considered that image noise significantly increases since it is inversely proportional to the square root of the slice thickness [22, 23]. As detailed above, EID-CTs use iterative reconstruction algorithms based on non-linear reconstruction methods, causing a loss in spatial and contrast resolution [20, 21]. Because of its technical limitation, iterative reconstruction cannot compensate for the increased image noise at thinner slices; therefore, to evaluate ECV with EID-CT, thicker slices are usually acquired or reformatted, even though this implies the presence of partial volume averaging. PCD-CT offers the unique possibility to overcome the limitation of partial volume averaging without reducing image quality due to its intrinsic increased spatial resolution and reduced image noise compared to EID-CT [24, 25]. The highest concordance that thinner slices have with MRI quantification of ECV is bound to the higher contrast resolution of myocardial tissue that is achieved with reduced slice thickness and hence reduced z-axial partial volume effect [26]. The increased spatial resolution and consequent increase in contrast resolution are fundamental in CT compared to MRI, where the presence of partial volume effect (MOLLI sequences conventionally use 8-mm slice thickness [27]) is balanced by the strength of pixel-wise mapping and the ability of MRI to detect small abnormalities and discriminate low-contrast structures [28]. Even with MRI’s intrinsic high contrast resolution, it is important to note that the use of relatively thick slices may blur the boundary between the myocardium and the blood pool, especially if the myocardial walls are thinned or cardiac motion artifacts are present. In our cohort, such issues prevented the calculation of global MRI-based ECV in six patients.

Finally, we observed a better ECV quantification, with reduced overestimation compared to the MRI reference, at the lower keV range (45 to 60 keV), with 45 keV being the most accurate VMI level. The VMI reconstruction of the LIE scan is pivotal to achieve reliable quantification of ECV, as it delivers both sufficient signal-to-noise ratio and CNR to allow the best alignment with CCTA and consequent extrapolation of quantitative data from the LIE scan-based iodine map. Our results confirm prior evidence demonstrating that lower VMI energies lead to higher diagnostic accuracy compared to higher keV [29, 30]. This difference can be explained by the increased CNR at lower keV and the relative stability of the signal-to-noise ratio throughout different keV levels [30]. As Gutjahr et al. demonstrated this is related to the capability of PCD systems to count all x-ray quanta whose energy is above the lowest threshold so that lower-energy x-rays, bearing most of the low-contrast information, contribute as much to the detector signal as higher energy x-rays [31]. Even though the iodine CNR is highest at 40 keV, studies demonstrated a tendency to prefer slightly higher keV reconstructions for diagnosis to avoid the high image noise present at 40 keV [13]. Euler et al., for example, showed that VMIs in the range 45 to 50 keV represent the best trade-off between objective and subjective image quality [32]; also Albrecht et al. demonstrated that the use of 50-keV VMI provides a reasonable combination of iodine attenuation with preserved image sharpness [33]. This evidence can be explanatory of our results, which show a higher mean bias in Bland–Altman plots for 40 keV compared to the 45–60-keV range, indicating a reduced accuracy in ECV quantification. The role of VMI reconstructions and QIR strength level with PCD-CT has been investigated in other settings and demonstrated a significant impact on CT diagnostic performances in both phantom and clinical studies [34]. This evidence, along with our results, emphasizes the importance to investigate the best acquisition and reconstruction parameters further to achieve the highest diagnostic performance in different clinical settings.

Our study has the following limitations: first, the single-center study design with a relatively small sample size, with a consequent small spectrum of ECV-altering cardiomyopathies investigated; second, the lack of a histological reference; third, the use of MRI-derived synthetic Hct; fourth, the choice of a step-by-step approach rather than a grid approach testing all the possible combinations, which could have introduced a bias; fifth, the use of only a definite range of slice thicknesses and increment; sixth, we did not investigate the impact of different kernels on the accuracy of ECV quantification. Finally, we used a 5-min delayed scan for ECV quantification; however, there are no current studies that investigated the optimal timing for the delayed scan for both ECV quantification and focal scar detection with the novel PCD-CT, an issue that should be investigated in further studies. However, given the novelty of PCD-CT technology, even a limited sample size with the mentioned limitations holds value, particularly considering the unique cohort of patients who underwent same-day MRI and CT imaging.

In conclusion, we demonstrated that an appropriate selection of image reconstruction parameters improves the agreement between ECV values quantified by PCD-CT and MRI. These results highlight the importance of tailoring reconstruction parameters to optimize PCD-CT performance for myocardial tissue characterization.

Availability of data and materials

The data that support the findings of this study are available from the corresponding authors upon reasonable request.

Abbreviations

CCC:

Concordance correlation coefficient

CCTA:

Coronary computed tomography angiography

CI :

Confidence interval

CNR:

Contrast-to-noise ratio

CT:

Computed tomography

ECV:

Extracellular volume

EID:

Energy-integrating detector

Hct:

Hematocrit

LIE:

Late iodine enhancement

LoA:

Limits of agreement

LV:

Left ventricle

MOLLI:

Modified Look-Locker inversion-recovery

PCD:

Photon-counting detector

QIR:

Quantum iterative reconstruction

VMI:

Virtual monoenergetic imaging

References

  1. Liu T, Song D, Dong J et al (2017) Current understanding of the pathophysiology of myocardial fibrosis and its quantitative assessment in heart failure. Front Physiol 8:238. https://doi.org/10.3389/fphys.2017.00238

    Article  PubMed  PubMed Central  Google Scholar 

  2. Aoki T, Fukumoto Y, Sugimura K et al (2011) Prognostic impact of myocardial interstitial fibrosis in non-ischemic heart failure. Comparison between preserved and reduced ejection fraction heart failure. Circ J 75:2605–2613. https://doi.org/10.1253/circj.cj-11-0568

    Article  CAS  PubMed  Google Scholar 

  3. Garcia MJ, Kwong RY, Scherrer-Crosbie M et al (2020) State of the Art: imaging for myocardial viability: a scientific statement from the American Heart Association. Circ Cardiovasc Imaging 13:e000053. https://doi.org/10.1161/HCI.0000000000000053

    Article  PubMed  Google Scholar 

  4. Haaf P, Garg P, Messroghli DR, Broadbent DA, Greenwood JP, Plein S (2016) Cardiac T1 Mapping and Extracellular Volume (ECV) in clinical practice: a comprehensive review. J Cardiovasc Magn Reson 18:89. https://doi.org/10.1186/s12968-016-0308-4

    Article  PubMed  PubMed Central  Google Scholar 

  5. Nacif MS, Kawel N, Lee JJ et al (2012) Interstitial myocardial fibrosis assessed as extracellular volume fraction with low-radiation-dose cardiac CT. Radiology 264:876–883. https://doi.org/10.1148/radiol.12112458

    Article  PubMed  PubMed Central  Google Scholar 

  6. Lardo AC, Cordeiro MA, Silva C et al (2006) Contrast-enhanced multidetector computed tomography viability imaging after myocardial infarction: characterization of myocyte death, microvascular obstruction, and chronic scar. Circulation 113:394–404. https://doi.org/10.1161/circulationaha.105.521450

    Article  PubMed  PubMed Central  Google Scholar 

  7. Vignale D, Palmisano A, Gnasso C et al (2023) Extracellular volume fraction (ECV) derived from pre-operative computed tomography predicts prognosis in patients undergoing transcatheter aortic valve implantation (TAVI). Eur Heart J Cardiovasc Imaging. https://doi.org/10.1093/ehjci/jead040

  8. Gerber BL, Belge B, Legros GJ et al (2006) Characterization of acute and chronic myocardial infarcts by multidetector computed tomography. Circulation 113:823–833. https://doi.org/10.1161/CIRCULATIONAHA.104.529511

    Article  PubMed  Google Scholar 

  9. Schuleri KH, Centola M, George RT et al (2009) Characterization of peri-infarct zone heterogeneity by contrast-enhanced multidetector computed tomography. J Am Coll Cardiol 53:1699–1707. https://doi.org/10.1016/j.jacc.2009.01.056

    Article  PubMed  PubMed Central  Google Scholar 

  10. Kurita Y, Kitagawa K, Kurobe Y et al (2016) Estimation of myocardial extracellular volume fraction with cardiac CT in subjects without clinical coronary artery disease: a feasibility study. J Cardiovasc Comput Tomogr 10:237–241. https://doi.org/10.1016/j.jcct.2016.02.001

    Article  PubMed  Google Scholar 

  11. van Assen M, De Cecco CN, Sahbaee P et al (2019) Feasibility of extracellular volume quantification using dual-energy CT. J Cardiovasc Comput Tomogr 13:81–84. https://doi.org/10.1016/j.jcct.2018.10.011

    Article  PubMed  Google Scholar 

  12. Lee H-J, Im DJ, Youn J-C et al (2016) Myocardial Extracellular Volume Fraction with Dual-Energy equilibrium contrast-enhanced cardiac CT in nonischemic cardiomyopathy: a prospective comparison with cardiac MR Imaging. Radiology 280:49–57. https://doi.org/10.1148/radiol.2016151289

    Article  PubMed  Google Scholar 

  13. Flohr T, Schmidt B (2023) Technical basics and clinical benefits of Photon-Counting CT. Invest Radiol. https://doi.org/10.1097/rli.0000000000000980

  14. Mergen V, Sartoretti T, Klotz E et al (2022) Extracellular Volume Quantification with cardiac late enhancement scanning using dual-source Photon-Counting Detector CT. Invest Radiol 57:406–411. https://doi.org/10.1097/rli.0000000000000851

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Aquino GJ, O’Doherty J, Schoepf UJ et al (2023) Myocardial characterization with extracellular volume mapping with a first-generation photon-counting detector CT with MRI reference. Radiology 307:e222030. https://doi.org/10.1148/radiol.222030

    Article  PubMed  Google Scholar 

  16. Robison S, Karur GR, Wald RM, Thavendiranathan P, Crean AM, Hanneman K (2018) Noninvasive hematocrit assessment for cardiovascular magnetic resonance extracellular volume quantification using a point-of-care device and synthetic derivation. J Cardiovasc Magn Reson 20. https://doi.org/10.1186/s12968-018-0443-1

  17. Cerqueira MD, Weissman NJ, Dilsizian V et al (2002) Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation 105:539–542. https://doi.org/10.1161/hc0402.102975

    Article  PubMed  Google Scholar 

  18. Ashby D (1991) Practical statistics for medical research. Douglas G. Altman, Chapman and Hall, London, 1991. Stat Med 10:1635–1636. https://doi.org/10.1002/sim.4780101015

    Article  Google Scholar 

  19. Willemink MJ, Noël PB (2019) The evolution of image reconstruction for CT—from filtered back projection to artificial intelligence. Eur Radiol 29:2185–2195. https://doi.org/10.1007/s00330-018-5810-7

    Article  PubMed  Google Scholar 

  20. Sartoretti T, Landsmann A, Nakhostin D et al (2022) Quantum iterative reconstruction for abdominal photon-counting detector CT improves image quality. Radiology 303:339–348. https://doi.org/10.1148/radiol.211931

    Article  PubMed  Google Scholar 

  21. McCollough CH, Yu L, Kofler JM et al (2015) Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology 276:499–506. https://doi.org/10.1148/radiol.15142047

    Article  PubMed  Google Scholar 

  22. Nakaura T, Iyama Y, Kidoh M et al (2016) Comparison of iterative model, hybrid iterative, and filtered back projection reconstruction techniques in low-dose brain CT: impact of thin-slice imaging. Neuroradiology 58:245–251. https://doi.org/10.1007/s00234-015-1631-4

    Article  PubMed  Google Scholar 

  23. Sato T, Takagi T (2021) Validation of the influence of CT slice thickness on the quantitative accuracy and image quality of single photon emission computed tomography. Asia Ocean J Nucl Med Biol 9:148–157. https://doi.org/10.22038/aojnmb.2021.54670.1376

    Article  PubMed  PubMed Central  Google Scholar 

  24. Rajendran K, Petersilka M, Henning A et al (2022) First Clinical Photon-counting detector CT system: technical evaluation. Radiology 303:130–138. https://doi.org/10.1148/radiol.212579

    Article  PubMed  Google Scholar 

  25. Shao J, Jiang J-S, Wang X-Y et al (2022) Measurement of myocardial extracellular volume using cardiac dual-energy computed tomography in patients with ischaemic cardiomyopathy: a comparison of different methods. Int J Cardiovasc Imaging 38:1591–1600. https://doi.org/10.1007/s10554-022-02532-z

    Article  Google Scholar 

  26. Monnin P, Sfameni N, Gianoli A, Ding S (2017) Optimal slice thickness for object detection with longitudinal partial volume effects in computed tomography. J Appl Clin Med Phys 18:251–259. https://doi.org/10.1002/acm2.12005

    Article  PubMed  Google Scholar 

  27. Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP (2004) Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Magn Reson Med 52:141–146. https://doi.org/10.1002/mrm.20110

    Article  PubMed  Google Scholar 

  28. Kellman P, Hansen MS (2014) T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson 16:2. https://doi.org/10.1186/1532-429x-16-2

    Article  PubMed  PubMed Central  Google Scholar 

  29. Sartoretti T, McDermott M, Mergen V et al (2023) Photon-counting detector coronary CT angiography: impact of virtual monoenergetic imaging and iterative reconstruction on image quality. Br J Radiol 96:20220466. https://doi.org/10.1259/bjr.20220466

    Article  PubMed  PubMed Central  Google Scholar 

  30. Jungblut L, Abel F, Nakhostin D et al (2023) Impact of photon counting detector CT derived virtual monoenergetic images and iodine maps on the diagnosis of pleural empyema. Diagn Interv Imaging 104:84–90. https://doi.org/10.1016/j.diii.2022.09.006

    Article  PubMed  Google Scholar 

  31. Gutjahr R, Halaweish AF, Yu Z et al (2016) Human imaging with photon counting-based computed tomography at clinical dose levels: contrast-to-noise ratio and cadaver studies. Invest Radiol 51:421–429. https://doi.org/10.1097/rli.0000000000000251

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  32. Euler A, Higashigaito K, Mergen V et al (2022) High-pitch photon-counting detector computed tomography angiography of the aorta: intraindividual comparison to energy-integrating detector computed tomography at equal radiation dose. Invest Radiol 57:115–121. https://doi.org/10.1097/rli.0000000000000816

    Article  CAS  PubMed  Google Scholar 

  33. Albrecht MH, Vogl TJ, Martin SS et al (2019) Review of clinical applications for virtual monoenergetic Dual-Energy CT. Radiology 293:260–271. https://doi.org/10.1148/radiol.2019182297

    Article  PubMed  Google Scholar 

  34. Booij R, van der Werf NR, Dijkshoorn ML, van der Lugt A, van Straten M (2022) Assessment of iodine contrast-to-noise ratio in virtual monoenergetic images reconstructed from dual-source energy-integrating CT and photon-counting CT data. Diagnostics (Basel) 12. https://doi.org/10.3390/diagnostics12061467

Download references

Acknowledgements

The authors hereby declare that large language models (LLMs) were not used in the writing of this article.

Funding

Open Access funding enabled and organized by Projekt DEAL. This study was partially funded by an institutional research grant from Siemens Healthcare.

Author information

Authors and Affiliations

Authors

Contributions

CG, DP, AV-S, and TE designed the study; CG wrote the study, and UJS, AV-S, TE, and MV-N reviewed and corrected the paper; GJA, NF, EZ, RJH, JS, and PS provided clinical and technical expertise. CG and DP performed experiments and statistical analyses. All authors read the manuscript and agreed to the submission of this paper.

Corresponding author

Correspondence to Tilman Emrich.

Ethics declarations

Ethics approval and consent to participate

The protocol of this study, Health Insurance Portability and Accountability Act–compliant, was approved by the local institutional review board of the Medical University of South Carolina (MUSC). Written informed consent was for all participants.

Consent for publication

All patients signed a written informed consent form, which includes the following statement: “Results of this research will be used for the purposes described in this study. This information may be published, but patients will not be identified. Information that is obtained concerning this research that can be identified will remain confidential to the extent possible within State and Federal law.” Potentially identifying information including age and gender are not provided for the case example.

Competing interests

CG: none; DP: none; UJS: receives institutional research support and/or personal fees by Bayer, Bracco, Elucid BioImaging, Guerbet, HeartFlow, Inc., Keya Medical, and Siemens; MV-N: none; GJA: none, NF: none; EZ: none; RJH: none; JS: none; PS: none; AV-S: receives institutional research support and/or personal subscriptions from Siemens and Elucid Bioimaging. He is the Deputy Editor of European Radiology Experimental; TE: receives travel support and speaker fee from Siemens Medical Solutions USA Inc, and institutional research support by Siemens Healthineers. He is a member of the Scientific Review Board of European Radiology Experimental (Section Editor, Computed Tomography).

The authors affiliated with European Radiology Experimental were not involved in the selection or review processes of this article.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gnasso, C., Pinos, D., Schoepf, U.J. et al. Impact of reconstruction parameters on the accuracy of myocardial extracellular volume quantification on a first-generation, photon-counting detector CT. Eur Radiol Exp 8, 70 (2024). https://doi.org/10.1186/s41747-024-00469-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-024-00469-7

Keywords