Reproducibility and repeatability of quantitative T2 and T2* mapping of osteosarcomas in a mouse model

Abstract

Background

New immunotherapies activate tumor-associated macrophages (TAMs) in the osteosarcoma microenvironment. Iron oxide nanoparticles (IONPs) are phagocytosed by TAMs and, therefore, enable TAM detection on T2*- and T2-weighted magnetic resonance images. We assessed the repeatability and reproducibility of T2*- and T2-mapping of osteosarcomas in a mouse model.

Methods

Fifteen BALB/c mice bearing-murine osteosarcomas underwent magnetic resonance imaging (MRI) on 3-T and 7-T scanners before and after intravenous IONP infusion, using T2*-weighted multi-gradient-echo, T2-weighted fast spin-echo, and T2-weighted multi-echo sequences. Each sequence was repeated twice. Tumor T2 and T2* relaxation times were measured twice by two independent investigators. Repeatability and reproducibility of measurements were assessed.

Results

We found excellent agreement between duplicate acquisitions for both T2* and T2 measurements at either magnetic field strength, by the same individual (repeatability), and between individuals (reproducibility). The repeatability concordance correlation coefficient (CCC) for T2* values were 0.99 (coefficients of variation (CoV) 4.43%) for reader 1 and 0.98 (CoV 5.82%) for reader 2. The reproducibility of T2* values between the two readers was 0.99 (CoV 3.32%) for the first acquisitions and 0.99 (CoV 6.30%) for the second acquisitions. Regarding T2 values, the repeatability of CCC was similar for both readers, 0.98 (CoV 3.64% for reader 1 and 4.45% for reader 2). The CCC of the reproducibility of T2 was 0.99 (CoV 3.1%) for the first acquisition and 0.98 (CoV 4.38%) for the second acquisition.

Conclusions

Our results demonstrated high repeatability and reproducibility of quantitative T2* and T2 mapping for monitoring the presence of TAMs in osteosarcomas.

Relevance statement

T2* and T2 measurements of osteosarcomas on IONP-enhanced MRI could allow identifying patients who may benefit from TAM-modulating immunotherapies and for monitoring treatment response. The technique described here could be also applied across a wide range of other solid tumors.

Key points

• Optimal integration of TAM-modulating immunotherapies with conventional chemotherapy remains poorly elucidated.

• We found high repeatability of T2* and T2 measurements of osteosarcomas in a mouse model, both with and without IONPs contrast, at 3-T and 7-T MRI field strengths.

• T2 and T2* mapping may be used to determine response to macrophage-modulating cancer immunotherapies.

Graphical Abstract

Background

Osteosarcoma is the most common bone tumor in children and adolescents. The 5-year survival rate of metastatic osteosarcoma is 30% [1]. Therefore, new treatment options are being investigated. New immunotherapies that activate tumor-associated macrophages (TAMs) in the osteosarcoma microenvironment have demonstrated promising results [2]. Since the tumor does not change in size in response to immunotherapy, at least not in the immediate post-treatment phase, an imaging technology that can visualize TAM activation would greatly help to identify responders to these new therapies. Intravenously injected iron oxide nanoparticles (IONPs) are phagocytosed by TAMs and shorten T2* and T2 relaxation times of osteosarcomas on magnetic resonance imaging (MRI) [3,4,5]. A hypointense contrast enhancement, as quantified by decreasing tumor T2* and T2 relaxation times, correlated with the quantity of macrophages within the tumor on histology [5, 6]. IONP-enhanced MRI can therefore help to detect TAMs in osteosarcomas, using T2* and T2 mapping techniques. However, the reproducibility and repeatability of these tumor T2* and T2 measurements have not been investigated in osteosarcomas.

The repeatability and reproducibility of T2* and T2 measurements for measurements of the iron content in the liver and heart muscle have been studied [7,8,9]. Several research groups demonstrated excellent to good agreement in intra-observer and inter-observer reproducibility in the T2* measurement of liver tissue [10,11,12]. Similarly, the T2* measurement of myocardial tissue has shown consistent reproducibility in different studies [8, 11, 13]. However, the accumulation of endogenous iron in the liver and heart muscle in the setting of hemosiderosis is typically homogenous, whereas the accumulation of IONPs in solid tumors is typically heterogeneous.

Relatively few publications report the reproducibility of T2* or T2 measurement in tumors [14, 15]. The reproducibility of R2* (R2* = 1/T2*) measurements of prostate adenocarcinoma was reported to be 64.6% [14] and the reproducibility of T2* measurements of liver metastasis in patients with colorectal cancer was reported to be 55.4% for 84th percentiles [15]. While we recognize that tumor tissue is more heterogeneous than that of most normal organs, we hypothesized that standardized approaches should yield significant agreements between repeated acquisitions and repeated measurements of tumor T2* and T2 values.

The purpose of our study was to assess the repeatability and reproducibility of T2* and T2 mapping as imaging biomarkers of TAM activation in a mouse osteosarcoma model.

Methods

Cell line culture

The murine osteosarcoma K7M2 (CRL-2836™) cell line was obtained from the American Type Culture Collection and cultured in Dulbecco’s Modified Eagle’s Medium supplemented with 10% fetal bovine serum, 100 mg/mL streptomycin, 100 U/mL penicillin and 2 mM L-glutamine. All cell culture reagents were provided by Gibco, Invitrogen, Carlsbad, CA USA. Cells were maintained at 37 °C in a humidified incubator infused with 5% CO2.

Osteosarcoma mouse model

All experimental procedures involving mice were approved by the Stanford University Administrative Panel on Laboratory Animal Care (Protocol 24,965). Fifteen female 6- to 8-week-old BALB/c mice (000651, Jackson Laboratory, Bar Harbor, ME, USA) were included in the experiment (Fig. 1). The mice were anesthetized via inhalation of 1.5−2.0% isoflurane in oxygen, and then 1 × 105 K7M2 tumor cells suspended in phosphate-buffered saline were implanted into the right proximal tibial metaphysis. Tumors with intraosseous and extraosseous soft tissue components, resembling the growth pattern of human tumors, typically grew to a size of 1 cm within 3 weeks after inoculation.

Fig. 1
figure 1

Experimental design: K7M2 murine osteosarcoma cells were implanted into the proximal right tibia of BALB/c mice. After 3 weeks, tumor-bearing mice underwent magnetic resonance imaging (MRI) on 7-T and 3-T scanners. Multiecho T2- and T2*-weighted sequences were acquired twice (acquisition 1 [Acq 1] and acquisition 2 [Acq 2]). Immediately after pre-contrast scanning (pre-Cx), iron oxide nanoparticles (IONPs) were injected through the tail vein at a dose of 30 Fe mg/kg. Twenty-four hours later, animals underwent post-contrast scanning (post-Cx) on both 7-T and 3-T scanners. The yellow arrows indicate tumors on T2-weighted images. A region of interest was used to delineate the tumors on these images

MRI scans

When the tumor had reached a size of 1 cm, tumor-bearing mice underwent MRI on 7-T and 3-T scanners (both Bruker Corporation, Billerica, MA, USA) with the parameters shown in Table 1. T2-weighted fast spin-echo images were acquired for anatomic reference and region of interest definition. Multislice and multiecho gradient-echo and spin-echo sequences were acquired for T2*- and T2-map calculations. Each mapping sequence was acquired twice. Technical parameters of both sequences at 3 T and 7 T are shown in Table 1. Immediately after MRI scanning, each mouse received a single intravenous injection of IONPs (Ferumoxytol; AMAG Pharmaceuticals, Inc., Waltham, MA, USA) at a dose of 30 Fe mg/kg (Fig. 1). Twenty-four hours after the IONP injection, animals underwent MRI on both 3-T and 7-T scanners as described above.

Table 1 Acquisition parameters of multislice sequences

T2* and T2 mapping

T2* and T2-maps were calculated based on the multi-gradient-echo and multi-slice multi-echo acquisitions, respectively. A mono-exponential decay model with a constant offset was fit to the signal in the input images, S(t), and the time of each echo, t, according to:

$$S\left(t\right)=A+S_o^\ast\text{exp}\left(-t/T\right)$$

where the parameters calculated by the model are A = absolute bias, So = initial signal intensity, and T = either T2* or T2. Those parameters are evaluated and mapped in Paravision 360 (3-T) and 7.1 (7-T) (Bruker Corp, Billerica MA, USA).

Parametric data analysis

MRI anatomic and parametric data were exported for analysis to Osirix version 8.0 (Pixmeo SARL, Bernex, Switzerland). Two trained researchers (R.R. and F.P.) independently measured T2* and T2 relaxation times of all tumors on pre-contrast and post-contrast acquisitions at both field strengths. These eight measurements were performed twice by each researcher. The standardized analysis procedure in Osirix was as follows:

  1. 1)

    Outline the entire tumor by manually drawing regions of interest circumscribing the tumor on each slice of the T2-weighted anatomic images;

  2. 2)

    Add a mask to exclude pixels below 7% (7 T) and 30% (3 T) of the signal of normal muscle (i.e., the mean signal value in the calf muscle of the contralateral leg) on the minimum TE images; and

  3. 3)

    Record the mean T2* and T2 relaxation times of the tumor entire tumor volume.

Statistical analysis

Descriptive data of tumor T2* and T2 relaxation times are shown as mean data ± standard deviation. Shapiro–Wilk and Shapiro-Francia tests were conducted to prove normality distribution of the data. A linear regression model was used to correlate T2* or T2 with field strength. Statistical significance was considered for p-values < 0.05.

The agreement between T2* and T2 measurements on first and second acquisitions by each reader (intraobserver) refers to the repeatability, and agreement between the two readers (inter-observer) refers to the reproducibility. Repeatability and reproducibility were examined using concordance correlation coefficient (CCC) and Bland–Altman analysis and the mean difference is presented with 95% limits of agreement (LOA) [16, 17]. In addition, we used coefficients of variation (CoV) to evaluate the distribution of the data relative to their average [18]. The CoV (%) is defined as the standard deviation of the means divided by the average of means. The CoV was assessed as poor (CoV > 20%), moderate (10% < CoV ≤ 20%), good (5% < CoV ≤ 10%), or excellent (CoV ≤ 5%) [19].

CoV is a measure of precision and provides an indication of how spread out the data are. However, CoV does not address bias, which refers to systematic error in measurements. CCC assesses the agreement between two sets of measurements, considering not only precision but also whether they have a bias. A CCC of 1 indicates perfect agreement, 0 indicates no agreement beyond chance, and -1 indicates perfect disagreement. Therefore, CoV is valuable for assessing precision, whereas CCC provides a more comprehensive assessment of accuracy by considering both precision and bias [16].

Results

T2*- and T2-weighted MRI

Pre-contrast MRI of the 15 female BALB/c mice confirmed osteosarcoma tumor growth. These images showed the development of a primary tumor in the proximal tibia, with an intraosseous tumor component, cortical disruption, and extraosseous soft tissue component, similar to the growth pattern in human patients (Figs. 2 and 3). Our analysis demonstrated normal distribution of T2* relaxation times for both first and second acquisitions as well as reader 1 and reader 2. In addition, T2* relaxation times showed normal distribution in terms of acquisitions and readers.

Fig. 2
figure 2

Reproducibility of T2* measurements. a Representative repeated T2*-weighted acquisitions (Acq) of osteosarcomas at 7 T before iron oxide nanoparticle (IONP) injection. b Quantitative tumor T2* relaxation times at 7 T before IONP injection. c Representative repeated T2*-weighted Acq of osteosarcomas at 3 T before IONP injection. d Quantitative tumor T2* relaxation times at 3 T before IONP injection. e Representative repeated T2*-weighted Acq of osteosarcomas at 7 T after IONP injection. f Quantitative tumor T2* relaxation times at 7 T before IONP injection. g Representative repeated T2*-weighted Acq of osteosarcomas at 3 T after IONP injection. h Quantitative tumor T2* relaxation times at 3 T after IONP injection. All quantitative data represent the mean data of 15 tumors in each group and standard deviations

Fig. 3
figure 3

Reproducibility of T2 measurements. a Representative repeated T2-weighted acquisitions (Acq) of osteosarcomas at 7 T before iron oxide nanoparticle (IONP) injection. b Quantitative tumor T2 relaxation times at 7 T before IONP injection. c Representative repeated T2-weighted Acq of osteosarcomas at 3 T before IONPs injection. d Quantitative tumor T2 relaxation times at 3 T before IONP injection. e Representative repeated T2-weighted Acq of osteosarcomas at 7 T after IONP injection. f Quantitative tumor T2 relaxation times at 7 T after IONP injection. g Representative repeated T2-weighted acquisitions of osteosarcomas at 3 T IONP injection. h Quantitative tumor T2 relaxation times at 3 T after IONP injection. All quantitative data represent the mean data of 15 tumors in each group and standard deviations

Pre-contrast T2* relaxation times of osteosarcomas were significantly shorter at 7 T (first acquisition, 14.2 ± 2.35 ms; second acquisition, 14.28 ± 2.08 ms; Fig. 2a, b) than at 3 T (first acquisition, 26.75 ± 5.56 ms; second acquisition, 26.71 ± 6.09 ms; p < 0.001; Fig. 2c, d). At 24 h after IONP infusion, all tumors demonstrated significant shortening of T2* relaxation times at 7 T (first acquisition, 10.21 ± 3 ms; second acquisition, 10.64 ± 2.97 ms; Fig. 2e, f) and at 3 T (first acquisition,16.63 ± 5.67 ms; second acquisition, 16.70 ± 5.73 ms, p < 0.001, Fig. 2g, h and Supplemental Fig. S1a).

Similarly, pre-contrast T2 relaxation times of osteosarcomas were significantly shorter at 7 T (first acquisition, 41.24 ± 6.58 ms; second acquisition, 41.60 ± 5.68 ms, Fig. 3a, b) than at 3 T (first acquisition 65.06 ± 2.64 ms; second acquisition, 65.11 ± 3.01 ms; p < 0.001; Fig. 3c, d). At 24 h after IONP infusion, all tumors demonstrated significant shortening of T2 relaxation times at 7 T (first acquisition, 34.44 ± 6.26 ms; second acquisition, 34.42 ± 6.13 ms; Fig. 3e, f) and at 3 T (first acquisition, 47.46 ± 7.89 ms; second acquisition, 48.18 ± 7.70 ms; p < 0.001, Fig. 3g, h and Supplemental Fig. S1b).

Repeatability of tumor T2* and T2 relaxation times

The intra-observer analyses of repeated T2* measurements demonstrated a strong agreement between the first and second T2* sequences for both readers using CCC. The intraobserver CCC for T2* values were 0.99 (95% CI 0.98−0.99, Fig. 4a) for reader 1 and 0.98 (95% CI (0.97−0.99, Fig. 4b) for reader 2. Concerning T2 measurements, an excellent intraobserver agreement was noted for readers 1 and 2. As depicted in Fig. 4c, d, intraobserver CCC was 0.98 (95% CI 0.98−0.99) and 0.98 (95% CI 0.97−0.98) for readers 1 and 2, respectively.

Fig. 4
figure 4

Repeatability of tumor T2* and T2 measurements. Bland–Altman plots of the difference between the first and second T2* measurements of the same tumor by reader 1 (a), and reader 2 (b). Bland–Altman plots of the difference between the first and second T2 measurements of the same tumor by reader 1 (c), and reader 2 (d). The green line indicates the mean difference, and the red lines indicate the 95% limits of agreement (LoA)

Reproducibility of tumor T2* and T2 relaxation times

We found an inter-observer variability of less than 5% for repeated measurements of T2* and T2 values. The inter-observer agreement of repeated T2* measurements by two readers was 0.99 (95% CI 0.99−0.99) for the first acquisitions (Fig. 5a) and 0.99 (95% CI 0.98−0.99) for the second acquisitions (Fig. 5b). The Bland–Altman plot also showed a strong inter-observer agreement of T2 values for the first (0.99, 95% CI 0.99−0.99), Fig. 5c) and second acquisitions (0.98, 95% CI 0.97−0.99, Fig. 5d).

Fig. 5
figure 5

Reproducibility of tumor T2* and T2 measurements. Bland–Altman plots of the difference between the first T2* measurements of the same tumor between two readers (a), and Bland–Altman plots of the difference between the second T2* measurements of the same tumor between two readers (b). Bland–Altman plots of the difference between the first T2 measurements of the same tumor between two readers (c), and Bland–Altman plots of the difference between the second T2 measurements of the same tumor between two readers (d). The green line indicates the mean difference, and the red lines indicate the 95% limits of agreement (LoA)

Coefficients of variation

As summarized in Table 2, the CoV for the first acquisition of T2* values was 3.32% indicating low variability. The CoV for the second acquisition obtained from T2*-mapping was 6.30%. Excellent agreement of 4.43% was seen in the T2* measurements by reader 1, and CoV was 5.82% for reader 2 for T2* values. The T2 estimates obtained from the first and second acquisitions also showed excellent agreement (3.1% and 4.38%, respectively). Both readers had an excellent agreement in T2-map measurements (Reader 1, 3.64%; Reader 2, 4.45%).

Table 2 Coefficients of variation (CoV) of T2* and T2 relaxation time measurement in repeated acquisitions by two readers

Discussion

Our research has established the repeatability and reproducibility of T2* and T2 quantification in an animal model of osteosarcoma, before and after the administration of IONP contrast, and at 3-T and 7-T MRI field strengths. Furthermore, our results demonstrate minimal intra- and inter-observer variability in T2* and T2 measurements in our osteosarcoma mouse model.

Quantitative imaging has progressively gained importance in both preclinical research and clinical imaging applications. Our data showed in accordance with the literature that post-contrast T2* and T2 values are shorter than their pre-contrast counterparts at both field strengths [10]. As anticipated, T2* and T2 values at 3 T were higher than at 7 T.

Our findings regarding the low variability of repeated T2* and T2 measurements at both field strengths align with previous studies across various diseases and organs [20,21,22,23,24]. For instance, Li et al. demonstrated the excellent reproducibility of T2 measurements at 3 T in cervical cancer [20], while Vietti Violi et al. exhibited strong intra- and inter-observer agreement in measuring T2 at 3T in different anatomical locations of the pancreas [21]. Ge et al. noted a shorter T2 signal decay time in patients with malignant lymph nodes, underlining the diagnostic potential of T2, with high intra- and inter-observer agreement [22]. Similarly, analysis of the lumbar spine in healthy volunteers indicated excellent repeatability in T2 measurements [23]. All of these previous studies focused on the repeatability and reproducibility of T2 and T2* measurements on unenhanced MRI images, whereas our study investigated the repeatability and reproducibility of T2 and T2* measurements after IONP administration.

We found an inter-observer variability of less than 5% for T2* and T2 measurements of osteosarcomas, which is consistent with the reported inter-observer variability in other tissues and other tumors.

For T2* measurements of iron overload in the liver, Positano et al. reported an intra- and inter-observer variability of 3.7% and 5.6%, respectively [10], Kirk reported an inter-observer variability of 4.4% [11] and Meloni et al. reported an intra- and inter-observer variability of 4% and 6.9%, respectively [12]. For T2* measurements of iron overload in the heart, Anderson et al. reported an inter-study variability of 5% [8], Westwood et al. reported a variability of 3.5% and 2.4% at two different centers for T2* [13] and Kirk et al. reported an inter-study and inter-observer variability of 5.9% and 5.4%, respectively [11]. Panek et al. found that T2* analysis is a sensitive approach for the clinical detection of oxygenation levels in head and neck squamous cell carcinoma at 3T [25].

In conclusion, we found significant repeatability and reproducibility of T2* and T2 measurements of osteosarcomas in a mouse model, both with and without IONP contrast, at 3T and 7T MRI field strengths. Furthermore, we have demonstrated strong intra- and inter-observer consistency of these measurements.

Availability of data and materials

The datasets generated for this study are available from the first author upon request.

Abbreviations

CCC:

Concordance correlation coefficient

CI:

Confidence interval

CoV:

Coefficients of variation

IONPs:

Iron oxide nanoparticles

MRI:

Magnetic resonance imaging

TAMs:

Tumor-associated macrophages

References

  1. Strauss SJ, Frezza AM, Abecassis N et al (2021) Bone sarcomas: ESMO–EURACAN–GENTURIS–ERN PaedCan Clinical Practice Guideline for diagnosis, treatment and follow-up. Ann Oncol 32:1520–1536. https://doi.org/10.1016/j.annonc.2021.08.1995

    Article  CAS  PubMed  Google Scholar 

  2. Mohanty S, Aghighi M, Yerneni K, Theruvath JL, Daldrup-Link HE (2019) Improving the efficacy of osteosarcoma therapy: combining drugs that turn cancer cell “don”t eat me’ signals off and “eat me” signals on. Mol Oncol 13:2049–2061. https://doi.org/10.1002/1878-0261.12556

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Aghighi M, Theruvath AJ, Pareek A et al (2018) Magnetic resonance imaging of tumor-associated macrophages: clinical translation. Clin Cancer Res 24:4110–4118. https://doi.org/10.1158/1078-0432.CCR-18-0673

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  4. Tong S, Zhu H, Bao G (2019) Magnetic iron oxide nanoparticles for disease detection and therapy. Mater Today (Kidlington) 31:86–99. https://doi.org/10.1016/j.mattod.2019.06.003

    Article  CAS  PubMed  Google Scholar 

  5. Roudi R, Pisani L, Pisani F, Kiru L, Daldrup-Link HE (2023) Novel clinically translatable iron oxide nanoparticle for monitoring anti-CD47 cancer immunotherapy. Invest Radiol 59:391–403. https://doi.org/10.1097/RLI.0000000000001030

    Article  CAS  PubMed  Google Scholar 

  6. Mohanty S, Yerneni K, Theruvath JL et al (2019) Nanoparticle enhanced MRI can monitor macrophage response to CD47 mAb immunotherapy in osteosarcoma. Cell Death Dis 10:36. https://doi.org/10.1038/s41419-018-1285-3

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. Hernando D, Levin YS, Sirlin CB, Reeder SB (2014) Quantification of liver iron with MRI: state of the art and remaining challenges. J Magn Reson Imaging 40:1003–1021. https://doi.org/10.1002/jmri.24584

    Article  PubMed  PubMed Central  Google Scholar 

  8. Anderson LJ, Holden S, Davis B et al (2001) Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J 22:2171–2179. https://doi.org/10.1053/euhj.2001.2822

    Article  CAS  PubMed  Google Scholar 

  9. Cancado R, Watman NP, Lobo C et al (2018) Assessment of liver and cardiac iron overload using MRI in patients with chronic anemias in Latin American countries: results from ASIMILA study. Hematology 23:676–682. https://doi.org/10.1080/10245332.2018.1461292

    Article  CAS  PubMed  Google Scholar 

  10. Positano V, Salani B, Pepe A et al (2009) Improved T2* assessment in liver iron overload by magnetic resonance imaging. Magn Reson Imaging 27:188–197. https://doi.org/10.1016/j.mri.2008.06.004

    Article  PubMed  Google Scholar 

  11. Kirk P, He T, Anderson LJ et al (2010) International reproducibility of single breathhold T2* MR for cardiac and liver iron assessment among five thalassemia centers. J Magn Reson Imaging 32:315–319. https://doi.org/10.1002/jmri.22245

    Article  PubMed  PubMed Central  Google Scholar 

  12. Meloni A, Luciani A, Positano V et al (2011) Single region of interest versus multislice T2* MRI approach for the quantification of hepatic iron overload. J Magn Reson Imaging 33:348–355. https://doi.org/10.1002/jmri.22417

    Article  PubMed  Google Scholar 

  13. Westwood MA, Firmin DN, Gildo M et al (2005) Intercentre reproducibility of magnetic resonance T2* measurements of myocardial iron in thalassaemia. Int J Cardiovasc Imaging 21:531–538. https://doi.org/10.1007/s10554-005-0651-2

    Article  PubMed  Google Scholar 

  14. Alonzi R, Taylor NJ, Stirling JJ et al (2010) Reproducibility and correlation between quantitative and semiquantitative dynamic and intrinsic susceptibility-weighted MRI parameters in the benign and malignant human prostate. J Magn Reson Imaging 32:155–164. https://doi.org/10.1002/jmri.22215

    Article  PubMed  Google Scholar 

  15. Ter Voert EGW, Heijmen L, de Wilt JHW et al (2013) Reproducibility and biological basis of in vivo T(2)* magnetic resonance imaging of liver metastasis of colorectal cancer. Magn Reson Med 70:1145–1152. https://doi.org/10.1002/mrm.24543

    Article  PubMed  Google Scholar 

  16. Lin LI (1989) A concordance correlation coefficient to evaluate reproducibility. Biometrics 45:255–268

    Article  CAS  PubMed  Google Scholar 

  17. Bartlett JW, Frost C (2008) Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables. Ultrasound Obstet Gynecol 31:466–475. https://doi.org/10.1002/uog.5256

    Article  CAS  PubMed  Google Scholar 

  18. Aerts S, Haesbroeck G, Ruwet C (2015) Multivariate coefficients of variation: comparison and influence functions. J Multivar Anal 142:183–198. https://doi.org/10.1016/j.jmva.2015.08.006

    Article  Google Scholar 

  19. Koo TK, Li MY (2016) A guideline of selecting and reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med 15:155–163. https://doi.org/10.1016/j.jcm.2016.02.012

    Article  PubMed  PubMed Central  Google Scholar 

  20. Li S, Liu J, Zhang F et al (2020) Novel T2 mapping for evaluating cervical cancer features by providing quantitative T2 maps and synthetic morphologic images: a preliminary study. J Magn Reson Imaging 52:1859–1869. https://doi.org/10.1002/jmri.27297

    Article  PubMed  Google Scholar 

  21. Vietti Violi N, Hilbert T, Bastiaansen JAM et al (2019) Patient respiratory-triggered quantitative T2 mapping in the pancreas. J Magn Reson Imaging 50:410–416. https://doi.org/10.1002/jmri.26612

    Article  PubMed  PubMed Central  Google Scholar 

  22. Ge Y-X, Hu S-D, Wang Z et al (2021) Feasibility and reproducibility of T2 mapping and DWI for identifying malignant lymph nodes in rectal cancer. Eur Radiol 31:3347–3354. https://doi.org/10.1007/s00330-020-07359-7

    Article  PubMed  Google Scholar 

  23. Raudner M, Schreiner M, Hilbert T et al (2020) Accelerated T2 mapping of the lumbar intervertebral disc: highly undersampled k-space data for robust T2 relaxation time measurement in clinically feasible acquisition Times. Invest Radiol 55:695–701. https://doi.org/10.1097/RLI.0000000000000690

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Cheng K, Duan Q, Hu J et al (2023) Evaluation of postcontrast images of intracranial tumors at 7T and 3T MRI: An intra-individual comparison study. CNS Neurosci Ther 29:559–565. https://doi.org/10.1111/cns.14036

    Article  PubMed  Google Scholar 

  25. Panek R, Welsh L, Dunlop A et al (2016) Repeatability and sensitivity of T2* measurements in patients with head and neck squamous cell carcinoma at 3T. J Magn Reson Imaging 44:72–80. https://doi.org/10.1002/jmri.25134

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Funding

This work was supported by a grant from the National Cancer Institute (grant number U24CA264298 [principal investigator HEDL]). The small animal imaging and biomedical statistics research resources used for this project were supported by the Stanford Cancer Institute (P30CA124435 grant [principal investigator Steven Artandi.] from the National Cancer Institute).

Author information

Authors and Affiliations

Authors

Contributions

All authors contributed to the study’s conception and design. Material preparation, data collection, and analysis were performed by RR, LJP, FP, and TL. The first draft of the manuscript was written by RR and substantially edited by HEDL. All authors read, commented, and approved the final manuscript. Funding was obtained by HEDL and all the studies were supervised by HEDL.

Corresponding authors

Correspondence to Raheleh Roudi or Heike E. Daldrup-Link.

Ethics declarations

Ethics approval and consent to participate

All experimental procedures involving mice were approved by the Stanford University Administrative Panel on Laboratory Animal Care (Protocol 24965).

Competing interests

The authors have no relevant financial or nonfinancial interests to disclose.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1:

Supplementary Fig. S1. Tumor T2* and T2 relaxation time measurements before and after iron oxide nanoparticle (IONP) injection: (a) Tumor T2* relaxation time at 3 T and 7 T before (pre) and after (post) IONP injection. (b) T2 relaxation time (ms) at 3 T and 7 T before (pre) and after (post) IONPs. All quantitative data represent the mean data of 15 tumors in each group and standard deviations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Roudi, R., Pisani, L.J., Pisani, F. et al. Reproducibility and repeatability of quantitative T2 and T2* mapping of osteosarcomas in a mouse model. Eur Radiol Exp 8, 74 (2024). https://doi.org/10.1186/s41747-024-00467-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-024-00467-9

Keywords