Skip to main content

Virtual noncontrast images reveal gouty tophi in contrast-enhanced dual-energy CT: a phantom study

Abstract

Background

Dual-energy computed tomography (DECT) is useful for detecting gouty tophi. While iodinated contrast media (ICM) might enhance the detection of monosodium urate crystals (MSU), higher iodine concentrations hamper their detection. Calculating virtual noncontrast (VNC) images might improve the detection of enhancing tophi. The aim of this study was to evaluate MSU detection with VNC images from DECT acquisitions in phantoms, compared against the results with standard DECT reconstructions.

Methods

A grid-like and a biophantom with 25 suspensions containing different concentrations of ICM (0 to 2%) and MSU (0 to 50%) were scanned with sequential single-source DECT using an ascending order of tube current time product at 80 kVp (16.5–220 mAs) and 135 kVp (2.75–19.25 mAs). VNC images were equivalently reconstructed at 80 and 135 kVp. Two-material decomposition analysis for MSU detection was applied for the VNC and conventional CT images. MSU detection and attenuation values were compared in both modalities.

Results

For 0, 0.25, 0.5, 1, and 2% ICM, the average detection indices (DIs) for all MSU concentrations (35–50%) with VNC postprocessing were respectively 25.2, 36.6, 30.9, 38.9, and 45.8% for the grid phantom scans and 11.7, 9.4, 5.5, 24.0, and 25.0% for the porcine phantom scans. In the conventional CT image group, the average DIs were respectively 35.4, 54.3, 45.4, 1.0, and 0.0% for the grid phantom and 19.4, 17.9, 3.0, 0.0, and 0.0% for the porcine phantom scans.

Conclusions

VNC effectively reduces the suppression of information caused by high concentrations of ICM, thereby improving the detection of MSU.

Relevance statement

Contrast-enhanced DECT alone may suffice for diagnosing gout without a native acquisition.

Key points

• Highly concentrated contrast media hinders monosodium urate crystal detection in CT imaging

• Virtual noncontrast imaging redetects monosodium urate crystals in high-iodinated contrast media concentrations.

• Contrast-enhanced DECT alone may suffice for diagnosing gout without a native acquisition.

Graphical Abstract

Background

Gout is a disease characterized by a granulomatous inflammatory response to the presence of monosodium urate (MSU) crystals in the joints, bones, and soft tissues with the formation of the so-called tophi [1]. The tophus contains cellular, humoral, and crystal components causing the clinical symptoms of gout, which typically presents itself as an acute monoarticular flare in the lower extremities. However, atypical and chronic forms of gout, such as involvement of the axial skeleton or polyarticular manifestations, can mimic other diseases like rheumatoid or psoriatic arthritis and may lead to wrong therapy decisions [2, 3].

Dual-energy computed tomography (DECT) has proven its efficacy in the detection of MSU using the different attenuation properties of tissues in different voltages [4]. Moreover, contrast-enhanced multimodal DECT has gained relevance in recent years, allowing the evaluation of periarticular soft tissue inflammation in addition to erosions, bone marrow edema, and other lesions [5, 6]. A single contrast-enhanced DECT scan in patients with an unclear arthritis etiology can thus provide objective evidence for active inflammation or rule out differential diagnoses. A recent phantom study of gouty arthritis has shown that DECT images acquired after administration of iodinated contrast medium (ICM) can improve the detection of tophi with a lower density that would remain undetected in unenhanced CT images [7]. At the same time, however, high ICM almost completely hampered tophus detection. Therefore, tophi with a strong contrast uptake may be obscured.

Virtual noncontrast (VNC) imaging is a postprocessing method for DECT scans. VNC images are generated by subtracting iodine maps from the contrast-enhanced images. This results in a set of native images that can be compared to an acquisition without administering iodine contrast to the patient. If successfully applied, this subtraction technique could potentially enable the identification of monosodium urate (MSU) despite high ICM concentrations.

The aim of our study was to investigate MSU detection in VNC images generated from contrast-enhanced DECT datasets.

Methods

Due to the phantom nature of the study, approval by the institutional review board was not necessary.

Phantom setup and scan protocol

To simulate different concentrations of MSU in vivo, we used two established phantom models with 25 suspensions containing 3 components in various concentrations: MSU (linear formula, C5H3N4O3Na; molecular weight, 190.09; Zeff, 7.7; Sigma Aldrich, St. Louis, MO, USA), ultrasound gel, and ICM containing 370 mg iodine/mL (Ultravist 370, iopromide; linear formula, C18H24I3N3O8; molecular weight, 791.1; Zeff, 41.4; Bayer Vital GmbH, Leverkusen, Germany). Five sets of suspensions containing MSU concentrations of 0%, 35%, 40%, 45%, and 50% mixed with varying ICM concentrations of 0%, 0.25%, 0.5%, 1%, and 2% (0.925, 1.85, 3.7, and 7.4 mg/mL of iodine, respectively) were prepared. The five MSU concentrations were selected based on previous work [7,8,9]. The suspensions were filled in 25 individual syringes. MSU masses were measured and documented. For the first phantom, the syringes were placed in a grid-like structure immersed in water. For the second phantom, a biophantom, five freshly prepared porcine forelegs with intact cutaneous, subcutaneous, muscular, and bone structures were bought from the slaughterhouse to mimic gouty tophi in realistic surroundings. In each foreleg, five syringes containing the same ICM concentration, but different MSU concentrations, were placed in the tissue surrounding the joint (Fig. 1).

Fig. 1
figure 1

Porcine legs with respectively five inserted syringes containing the same ICM concertation

All phantoms were scanned in a 320-row scanner (Aquilion ONE Vision Edition; Canon Medical Systems, Otawara, Japan) with the “dual-energy volume scan” mode with a total number of 10 acquisitions per phantom using a single-source technique with 2 sequential scans at 80 kVp and 135 kVp with 16-cm z-axis coverage without table movement and a rotation time of 0.275 s. The 10 acquisitions had an ascending order of tube currents ranging from 60 to 800 mA at 80 kVp and 10 to 140 mA at 135 kVp. All scans were reconstructed using adaptive iterative dose reduction at a strong iteration level (AIDR3Dstrong) with a medium soft tissue kernel without beam hardening compensation (FC13), a 0.5-mm slice thickness, and a 0.25-mm reconstruction interval. The tube current–time product, CT dose index, and dose-length product were documented, and the estimated effective dose was calculated using a conversion coefficient for upper extremities of 0.0008 \(\frac{{\text{mSv}}}{{\text{mGy}}\times {\text{cm}}}\) [10, 11].

Image postprocessing

Three-material decomposition was used to calculate iodine maps from the original images, resulting in two VNC images equivalent to 80 and 135 kVp. All 60 phantom acquisitions were reconstructed using AIDR3Dstrong image reconstructions (10 grid phantom scans and 50 porcine biophantom scans). The VNC images were generated using a proprietary clinical software package (version 6; Canon Medical Systems). Standard software settings by the vendor were used for calculation: water, 0/0 HU; MSU, 100/107 HU for low and high kVp with a gradient of 0.55 for iodine.

The calculated sets of two VNC images at 80 and 135 keV were processed in the same way as conventional DECT scans using a two-material decomposition analysis algorithm (Dual-Energy Composition Analysis, version 6.0, Canon Medical Systems) for the detection of uric acid using clinical standard settings established in previous studies [7,8,9,10,11,12].

The 60 phantom acquisitions without any prior subtraction were processed as conventional CT images with the same two-material decomposition analysis algorithm, thereby creating our reference group.

Quantitative analysis

The MSU volumes detected using two-material decomposition analysis and MSU masses estimated using an established method were recorded for both VNC images and conventional CT images [12]. False-positive detections within the phantom but outside of the syringes were classified as artifacts and quantified by their volume. Due to the variability in absolute MSU content within the syringes, we established a detection index (DI) to enhance the comparability of results.

$$Detection\;index\left[\mathit\%\right]\mathit=\left(\frac{detected\mathit\;MSU\mathit\;mass\mathit\;\left[\textit{mg}\right]}{real\mathit\;MSU\mathit\;mass\mathit\;\left[\textit{mg}\right]}\right)\times100$$

The resulting DIs from the VNC images were compared with the results of the datasets of the conventional CT images, described in detail in a previous study [7].

Attenuation values of the syringes were measured in the grid-like and pig foreleg phantom. Consistency was maintained by employing the same layer, location, and size for the region of interest in both VNC and conventional CT images.

Statistical analysis

Statistical significance was measured by using a paired t-test for intergroup comparability after VNC usage. Statistical analysis and visualization of data were carried out using Microsoft Excel (2016, Microsoft Corporation, Redmond, WA, USA) and Jupyter Notebook (2022, Anaconda Inc., Austin, TX, USA) using the pandas, matplotlib, scipy, and seaborn libraries.

Results

Phantom setup and scan protocol

Phantom setup and imaging were executed successfully. All acquisition and dose parameters are summarized in Table 1.

Table 1 Acquisition parameters

Post CT image processing

All datasets were successfully postprocessed. False-positive MSU artifacts inside the bone compartment were excluded from the analysis. For images of the postprocessing steps, please see Fig. 2.

Fig. 2
figure 2

Upper row (grid phantom): The syringes are ordered by their concentration of iodinated contrast medium (ICM) in ascending order and monosodium urate (MSU) in descending order. a MSU map without virtual-non-contrast (VNC), highlighting detected MSU coded in red. b Iodine map (ICM coded in gray). c MSU map with VNC postprocessing. Lower row (porcine phantom): The syringes are placed around the elbow joint in ascending clockwise order. d MSU map of a phantom without ICM showing MSU detection. e Phantom with 2% ICM completely hampering the MSU detection. f VNC postprocessing enables MSU detection in the same phantom with 2% ICM. False-positive MSU detection in bone structure is observed

Quantitative analysis

With VNC postprocessing, the average DIs for all MSU concentrations (35–50%) and tube currents were 25.2, 36.6, 30.9, 38.9, and 45.8% respectively for 0, 0.25, 0.5, 1, and 2% ICM for the grid phantom and 11.7, 9.4, 5.5, 24.0, and 25.0% for the porcine biophantom. In conventional CT images, the average DIs were respectively 35.4, 54.3, 45.4, 1.0, and 0.0% for the grid phantom and 19.4, 17.9, 3.0, 0.0, and 0.0% for the porcine biophantom. The results are presented in detail in Fig. 3.

Fig. 3
figure 3

Heatmaps of detection indices. Results are shown for the grid phantom (a, b) and the porcine phantom (c, d) with the use of virtual noncontrast (VNC) and conventional CT images. The x-axis in each of the heatmaps represents the concentration of iodinated contrast medium (ICM), while the y-axis represents the uric acid (MSU) concentration. The heatmap is ordered similarly to our grid phantom array. Each of the squares contains the value of the mean detection index (DI) of every syringe and is color-coded according to the color bar on the right side. Both in the grid-like and in the biophantom, VNC imaging enables MSU detection even in the presence of higher ICM concentrations (1% and 2%), at which no MSU is detected without VNC imaging

The DIs measured in the grid phantom images were significantly higher than those of the porcine biophantom across all ICM concentrations investigated. While DIs varied significantly with the concentrations of MSU and ICM present in the syringes and the tube currents used, no trend was discernible.

In syringes with higher ICM concentrations (1% and 2%), the use of VNC images significantly improved MSU detection to a rate similar to that achieved for conventional CT images without ICM. The VNC images showed lower DIs in syringes without ICM or with lower ICM concentrations (0.25 and 0.5%) in comparison with conventional CT images (p = 0.015 and p = 0.008, respectively). For comparisons, see Fig. 4.

Fig. 4
figure 4

Detection index for all iodinated contrast medium (ICM) concentrations. Boxplot diagram with different phantom setups (x-axis) and their respective DI (y-axis): contrast-enhanced images without the use of virtual noncontrast (VNC) imaging have higher DI in low ICM concentrations (0.25% and 0.5%), while DI are reduced to a value of zero in high ICM concentrations (1% and 2%). VNC images show lower DI in noncontrast and low IC concentrations (0.25% and 0.5%), while reaching significantly higher results in higher ICM concentrations (1% and 2%); p-values are shown above the compared groups

Syringes not containing MSU never showed false-positive MSU detection. There was also no false-positive MSU detection in the surrounding connective and soft tissue components in the porcine legs.

In conventional CT images, attenuation values in syringes with 50% MSU in the grid phantom showed averages of 263.2, 279.2, 289.7, 364.6, and 446.9 HU in 80 kVp and 257.5, 267.9, 270.7, 324.5, and 369.7 HU in 135 kVp for 0.0, 0.25, 0.5, 1.0, and 2% ICM, respectively. With VNC imaging, syringes with 50% MSU in the grid phantom showed average attenuations of 218.9, 221.7, 214.0, 238.9, and 238.6 HU in 80 kVp equivalent images and 233.2, 236.3, 229.0, 255.4, and 255.2 HU in 135 kVp equivalent images for 0.0, 0.25, 0.5, 1.0, and 2% ICM, respectively.

Attenuation measurements showed lower values in VNC images and revealed an increasing difference between VNC and conventional CT images as the concentration of ICM increased for both phantoms and all tube voltages investigated. In all ICM concentrations, VNC successfully altered the attenuation values to match a dual-energy gradient optimized for MSU detection. For an explanatory example, see Fig. 5.

Fig. 5
figure 5

Attenuation values for 50% monosodium urate (MSU) concentration. Exemplary visual representation of attenuation values in 50% MSU syringes in the raster phantom under varying tube voltages (80 and 135 kVp). Attenuation values in enhanced images were increasingly higher at 80 kV than at 135 kV, converging toward the dual-energy gradient of iodine (≈ 0.55) as the ICM concentration increased. In virtual noncontrast (VNC) imaging, the dual-energy gradient (≈ 1.05) becomes optimized for MSU detection

Discussion

This phantom study shows that VNC imaging enables the detection of MSU crystals even in the presence of high ICM concentrations, thereby resolving the hampering effect of iodine on MSU detection described before [7].

We assessed MSU volumes in phantoms using a two-material decomposition analysis. Subsequently, we estimated MSU masses through an established method that considers local attenuation levels. Given the acknowledged limited sensitivity of DECT for detecting non-tophaceous gout, as reported in other studies [12,13,14], owing to its reliance on local MSU concentrations and subthreshold CT attenuation, we employed the DI to objectively substantiate our findings.

The DIs found in our study for syringes with higher ICM concentrations in VNC images showed comparable magnitudes to those of scans with low or no ICM without VNC postprocessing. However, syringes with low or no ICM showed lower DIs in VNC imaging than the same syringes in images without VNC postprocessing.

Two-material decomposition differentiates between tissues by taking advantage of their different attenuation properties at different voltages. MSU is an organic purine derivative with a low effective atomic number of Zeff 7.7, similar to that of soft tissues. In DECT imaging, MSU is distinguished from soft tissues by setting a certain threshold value. This threshold can be modified to improve either the sensitivity or specificity of detection [7].

Adding an iodinated contrast medium, which has a high Zeff of 41.4, can therefore have two distinct effects on a voxel: (i) it raises the density of the voxel and (ii) it increases the Zeff of the tissue within the voxel, thus moving the dual-energy index toward that of iodine.

These mechanisms explain the positive results of VNC postprocessing on detection for syringes with higher ICM concentrations as subtraction of iodine reduces the Zeff and thus enables proper characterization. On the other hand, poorer MSU sensitivity with VNC in samples with a low iodine concentration is easily explained by the previously described beneficial effect of ICM on tophus detection: ICM raises the attenuation values above the detection threshold without grossly affecting the dual-energy index, thus enhancing the characterization of MSU. Removing iodine negates this effect. Finally, the lower detection in VNC images of MSU samples without ICM needs attention. We used standard clinical settings for postprocessing, which include soft tissue and fat as basis materials to calculate the VNC images. However, MSU and other components typically found in the joints are not considered in this formula. This is also evident through the impact of VNC imaging on calcium detection. Our application of VNC imaging leads to a partial elimination of the ability to differentiate between lesions containing calcium and those containing MSU (Fig. 2f), as discussed by other authors [15]. These factors might result in an unwanted influence of the VNC postprocessing that might be eliminated or optimized in further research. Therefore, we recommend MSU analysis with both VNC and conventional CT images, when suspecting gouty tophi in a contrast-enhanced DECT.

Interestingly, a very similar issue exists in the imaging of urinary calculi, which consist of uric acid. Unenhanced CT is the reference standard for the detection of stones in the urinary tract because they can be obscured by ICM in the renal parenchyma or collecting system. Several phantoms and in vivo studies have shown that VNC postprocessing resolves the hampering effect of ICM, consistent with our findings [16,17,18,19].

In recent years, DECT has evolved into a powerful multiparametric tool in arthritis imaging, enabling the detection of erosions, bone marrow edema, and MSU depositions [6, 20, 21]. With the application of ICM and the use of iodine maps, it even allows the detection of increased perfusion in synovitis and other musculoskeletal inflammatory lesions [22, 23].

Our data show that, depending on the enhancement level, MSU detection in contrast-enhanced DECT can vary dramatically, without an actual change in a tissue MSU content. In clinical practice, the local ICM concentration is influenced by the perfusion and inflammatory activity in a tophus. This poses a possible restriction in the quantification of MSU volume in serial examinations over the course of the disease as well as considering the given amount of iodine during the contrast administration. These dependencies might preclude objective monitoring of the response to urate-lowering therapy in comparison with the sole use of unenhanced DECT, as discussed by other authors [24].

Our study has several limitations. Firstly, despite our meticulous preparation of the syringes, we cannot entirely rule out the possibility of inhomogeneous distribution of the syringe contents. Secondly, it is important to note that our results may not be readily applicable to other DECT systems, since different vendors offer varying hardware and software configurations, which could yield different outcomes.

Additionally, our use of VNC postprocessing was not optimized for periarticular imaging and was primarily designed to differentiate three materials which might affect the depiction of other structures such as calcium-containing tissue. Furthermore, it is essential to acknowledge that our study was conducted as a phantom study with clinical parameters while lacking a patient cohort to investigate clinical relevance. This limitation leaves open the question of whether contrast uptake of tophi in actual patients would reach the high levels of ICM analyzed in our study. More research with patient data is needed to establish clinical significance.

In summary, this study demonstrates that VNC imaging successfully reduces the hampering effects caused by high ICM concentrations in the detection of MSU crystals by DECT. While the full clinical implications of these findings require further investigation, it appears feasible to opportunistically screen for gout by analyzing contrast-enhanced DECT scans with and without VNC postprocessing, especially when highly vascularized lesions are present. Nevertheless, it is advisable to perform an unenhanced scan when gouty arthritis is suspected to ensure comparability of follow-up examinations.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

CT:

Computed tomography

DECT:

Dual-energy CT

DI:

Detection index

ICM:

Iodinated contrast medium

MSU:

Monosodium urate

VNC:

Virtual noncontrast

References

  1. Chhana A, Dalbeth N (2015) The gouty tophus: a review. Curr Rheumatol Rep 17:19. https://doi.org/10.1007/s11926-014-0492-x

    Article  CAS  PubMed  Google Scholar 

  2. López-Reyes A, Hernández-Díaz C, Hofmann F et al (2012) Gout mimicking psoriatic arthritis flare. J Clin Rheumatol 18:220. https://doi.org/10.1097/RHU.0b013e318259aa2e

    Article  PubMed  Google Scholar 

  3. Nunes EA, Rosseti AG, Ribeiro DS et al (2014) Gout initially mimicking rheumatoid arthritis and later cervical spine involvement. Case Rep Rheumatol 2014:357826. https://doi.org/10.1155/2014/357826

    Article  PubMed  PubMed Central  Google Scholar 

  4. Carotti M, Salaffi F, Filippucci E (2020) Clinical utility of dual energy computed tomography in gout: current concepts and applications. Acta Biomed 91:116–124. https://doi.org/10.23750/abm.v91i8-S.9942

    Article  PubMed  PubMed Central  Google Scholar 

  5. Ulas ST, Ziegeler K, Richter ST et al (2022) Contrast-enhanced CT techniques and MRI perform equally well in arthritis imaging of the hand: a prospective diagnostic accuracy study. Eur Radiol 32:6376–6383. https://doi.org/10.1007/s00330-022-08744-0

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Diekhoff T, Engelhard N, Fuchs M et al (2019) Single-source dual-energy computed tomography for the assessment of bone marrow oedema in vertebral compression fractures: a prospective diagnostic accuracy study. Eur Radiol 29:31–39. https://doi.org/10.1007/s00330-018-5568-y

    Article  PubMed  Google Scholar 

  7. Kotlyarov M, Mews J, Ulas ST et al (2023) Influence of contrast medium on tophus detection using dual-energy CT: phantom study and clinical illustration. Eur Radiol Exp 7:43. https://doi.org/10.1186/s41747-023-00348-7

    Article  PubMed  PubMed Central  Google Scholar 

  8. Diekhoff T, Kiefer T, Stroux A et al (2015) Detection and characterization of crystal suspensions using single-source dual-energy computed tomography: a phantom model of crystal arthropathies. Invest Radiol 50:255–260. https://doi.org/10.1097/RLI.0000000000000099

    Article  PubMed  Google Scholar 

  9. Diekhoff T, Kotlyarov M, Mews J et al (2018) Iterative reconstruction may improve diagnosis of gout: an ex vivo (bio)phantom dual-energy computed tomography study. Invest Radiol 53:6–12. https://doi.org/10.1097/RLI.0000000000000402

    Article  PubMed  Google Scholar 

  10. Saltybaeva N, Jafari ME, Hupfer M et al (2014) Estimates of effective dose for CT scans of the lower extremities. Radiology 273:153–159. https://doi.org/10.1148/radiol.14132903

    Article  PubMed  Google Scholar 

  11. European Commission (2008) CT safety & efficacy: a broad perspective. 2004 CT quality criteria. Appendix A - MSCT dosimetry, guidelines on radiation dose to the patient. Available via: https://www.biophysicssite.com/Documents/MSCT2004b/Appendix_A_MSCT_Dosimetry.pdf Accessed 01 Dec 2023

  12. Kotlyarov M, Hermann KGA, Mews J et al (2020) Development and validation of a quantitative method for estimation of the urate burden in patients with gouty arthritis using dual-energy computed tomography. Eur Radiol 30:404–412. https://doi.org/10.1007/s00330-019-06350-1

    Article  CAS  PubMed  Google Scholar 

  13. Ahn SJ, Zhang D, Levine BD et al (2021) Limitations of dual-energy CT in the detection of monosodium urate deposition in dense liquid tophi and calcified tophi. Skeletal Radiol 50:1667–1675. https://doi.org/10.1007/s00256-021-03715-w

    Article  PubMed  Google Scholar 

  14. Gamala M, Jacobs JWG, Van Laar JM (2019) The diagnostic performance of dual energy CT for diagnosing gout: a systematic literature review and meta-analysis. Rheumatology (Oxford) 58:2117–2121. https://doi.org/10.1093/rheumatology/kez180

    Article  PubMed  Google Scholar 

  15. Christiansen SN, Müller FC, Østergaard M et al (2020) Dual-energy CT in gout patients: do all colour-coded lesions actually represent monosodium urate crystals? Arthritis Res Ther 22:212. https://doi.org/10.1186/s13075-020-02283-z

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Scheffel H, Stolzmann P, Frauenfelder T et al (2007) Dual-energy contrast-enhanced computed tomography for the detection of urinary stone disease. Invest Radiol 42:823–829. https://doi.org/10.1097/RLI.0b013e3181379bac

    Article  PubMed  Google Scholar 

  17. Takahashi N, Hartman RP, Vrtiska TJ et al (2008) Dual-energy CT iodine-subtraction virtual unenhanced technique to detect urinary stones in an iodine-filled collecting system: a phantom study. AJR Am J Roentgenol 190:1169–1173. https://doi.org/10.2214/AJR.07.3154

    Article  PubMed  PubMed Central  Google Scholar 

  18. Takahashi N, Vrtiska TJ, Kawashima A et al (2010) Detectability of urinary stones on virtual nonenhanced images generated at pyelographic-phase dual-energy CT. Radiology 256:184–190. https://doi.org/10.1148/radiol.10091411

    Article  PubMed  Google Scholar 

  19. Wang J, Qu M, Duan X et al (2012) Characterisation of urinary stones in the presence of iodinated contrast medium using dual-energy CT: a phantom study. Eur Radiol 22:2589–2596. https://doi.org/10.1007/s00330-012-2532-0

    Article  PubMed  PubMed Central  Google Scholar 

  20. Nicolaou S, Liang T, Murphy DT et al (2012) Dual-energy CT: a promising new technique for assessment of the musculoskeletal system. AJR Am J Roentgenol 199:S78-86. https://doi.org/10.2214/AJR.12.9117

    Article  PubMed  Google Scholar 

  21. Ziegeler K, Hermann S, Hermann KGA et al (2020) Dual-energy CT in the differentiation of crystal depositions of the wrist: does it have added value? Skeletal Radiol 49:707–713. https://doi.org/10.1007/s00256-019-03343-5

    Article  PubMed  Google Scholar 

  22. Fukuda T, Umezawa Y, Asahina A et al (2017) Dual energy CT iodine map for delineating inflammation of inflammatory arthritis. Eur Radiol 27:5034–5040. https://doi.org/10.1007/s00330-017-4931-8

    Article  PubMed  Google Scholar 

  23. Fukuda T, Umezawa Y, Tojo S et al (2017) Initial experience of using dual-energy CT with an iodine overlay image for hand psoriatic arthritis: comparison study with contrast-enhanced MR imaging. Radiology 284:134–142. https://doi.org/10.1148/radiol.2016161671

    Article  PubMed  Google Scholar 

  24. Dalbeth N, Choi HK (2013) Dual-energy computed tomography for gout diagnosis and management. Curr Rheumatol Rep 15:301. https://doi.org/10.1007/s11926-012-0301-3

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

No LLM was used for ideation. No images were produced by generative artificial intelligence.

The authors thank Ms. Bettina Herwig for the language editing.

Funding

Open Access funding enabled and organized by Projekt DEAL. The Department of Radiology received a research grant from Canon Medical Systems. We acknowledge the financial support from the Open Access Publication Fund of Charité — Universitätsmedizin Berlin and the German Research Foundation (DFG).

Author information

Authors and Affiliations

Authors

Contributions

KK had lead responsibility for data curation, formal analysis, investigation, validation, visualization, and writing (original draft). TD had lead responsibility for funding acquisition, was equally responsible for conceptualization, and had a supporting role in formal analysis, investigation, methodology, and writing (review and editing). JM had lead responsibility for software and was equally responsible for investigation, resources, and writing (review and editing). He had a supporting role in conceptualization, methodology, and supervision. SS was equally responsible for writing (review and editing). MK had a leading role in data acquisition, project administration, supervision, and writing (review and editing). He was equally responsible for the conceptualization and had a supporting role in formal analysis, investigation, and methodology.

Corresponding author

Correspondence to Karim Khayata.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

JM is an employee of Canon Medical Systems Europe BV. However, the corresponding nonemployee author (KK) had full control of the data and information submitted for publication. The company did not influence or authorize the results or the publication of our research. The other authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Khayata, K., Diekhoff, T., Mews, J. et al. Virtual noncontrast images reveal gouty tophi in contrast-enhanced dual-energy CT: a phantom study. Eur Radiol Exp 8, 69 (2024). https://doi.org/10.1186/s41747-024-00466-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-024-00466-w

Keywords