A microCT-based platform to quantify drug targeting

Abstract

Background

Heterotopic ossification (HO) is a frequent and debilitating complication of traumatic musculoskeletal injuries and orthopedic procedures. Prophylactic dosing of botulinum toxin type A (BTxA) holds potential as a novel treatment option if accurately distributed throughout soft-tissue volumes where protection is clinically desired. We developed a high-resolution, microcomputed tomography (microCT)-based imaging strategy to assess drug distribution and validated this platform by quantifying distribution achieved via a prototype delivery system versus a single-bolus injection.

Methods

We injected an iodine-containing contrast agent (iodixanol 320 mg I/mL) into dissected rabbit musculature followed by microCT imaging and analysis. To contrast the performance of distributed versus bolus injections, a three-dimensional (3D) 64-cm3-printed soft-tissue holder was developed. A centered 2-cm3 volume of interest (VOI) was targeted with a single-bolus injection or an equal volume distributed injection delivered via a 3D-printed prototype. VOI drug coverage was quantified as a percentage of the VOI volume that was < 1.0 mm from the injected fluid.

Results

The microCT-based approach enabled high-resolution quantification of injection distribution within soft tissue. The distributed dosing prototype provided significantly greater tissue coverage of the targeted VOI (72 ± 3%, mean ± standard deviation) when compared to an equal volume bolus dose (43 ± 5%, p = 0.031) while also enhancing the precision of injection targeting.

Conclusions

A microCT-based imaging technique precisely quantifies drug distribution within a soft-tissue VOI, providing a path to overcome a barrier for clinical translation of prophylactic inhibition of HO by BTxA.

Relevance statement

This platform will facilitate rapid optimization of injection parameters for clinical devices used to effectively and safely inhibit the formation of heterotopic ossification.

Key points

• MicroCT provides high-resolution quantification of soft-tissue drug distribution.

• Distributed dosing is required to maximize soft-tissue drug coverage.

• Imaging platform will enable rapid screening of 3D-printed drug distribution prototypes.

Graphical Abstract

Background

Heterotopic ossification (HO) is defined as bone formation that occurs in soft tissue outside the skeleton and is a frequent complication of surgical procedures required to address musculoskeletal traumas [1,2,3]. As the location of HO and trauma is commonly coincident, HO frequently forms near joints and causes pain, impaired movement, and, in the most severe cases, joint encapsulation leading to total loss of function [4, 5]. Once HO is symptomatic, the only treatment option is surgical excision of the ectopic bone, which is costly, technically demanding, time-consuming, and has high recurrence rates reported in the range of 6 to 36% predicated on anatomical location of the initial surgical procedure as well as surgical approach and timing [6,7,8,9,10,11].

The current standard of care for HO prevention is either high-dose-focused radiation therapy or a 1- to 6-week postoperative course of a nonsteroidal anti-inflammatory drugs (NSAID) such as indomethacin [12, 13]. The use of radiation therapy is limited by the lack of widespread availability, cost, and potential long-term complications [14,15,16]. The reported benefits of NSAID therapy vary considerably across studies and indications (e.g., less effective for acetabular fracture fixation versus total hip replacement [17,18,19,20]). Recent studies have also raised concerns regarding the risk-to-reward ratio of NSAIDs as an HO treatment [21] due to elevated gastrointestinal bleeding, increased cardiovascular events, and impaired fracture healing [22,23,24,25]. We therefore believe that a prophylactic HO intervention via local treatment of tissue with demonstrably improved efficacy and safety versus NSAIDs will improve patient care and surgical outcomes.

Prophylactic treatment of soft tissue with botulinum toxin type A (BTxA) holds potential to achieve this goal. This strategy is derived from observations that trauma-induced disruption of neuromuscular signaling causes a neuroinflammatory cascade that, in turn, drives differentiation of pluripotent cells into the bone-forming osteoblasts that lead to HO [26,27,28,29]. In context with our recent studies of how neuromuscular dysfunction mediates bone cell activity [30,31,32], we speculated that BTxA-induced inhibition of neuromuscular function would inhibit HO. Using a murine model of bone morphogenetic protein-induced HO, we found that a single dose of BTxA reduced HO volume by 50% [33]. A series of subsequent studies then demonstrated that the effectiveness of this intervention was not due to the paralytic effect of BTxA but was spatially associated with the site where BTxA was injected [34]. Thus, translation of this intervention will require a device that enables targeted delivery of distributed small doses of BTxA within the muscle volume where protection from HO-related pain and dysfunction is clinically desirable.

Unfortunately, this translational goal is not currently achievable as previous imaging strategies to visualize distribution of injected fluid within soft tissue (e.g., magnetic resonance imaging) have a native resolution nearly an order of magnitude lower than that is needed for this purpose [35]. In this study, we address this technical barrier by developing a microcomputed tomography (microCT)-based platform to quantify three-dimensional (3D) drug distribution within muscle. To validate our approach, we assessed whether a prototype 3D-printed distribution device enhanced drug distribution within a target volume of interest versus a bolus dose.

Methods

Study design overview

All studies were performed ex vivo with either harvested (leporine) or procured (porcine) tissues. Quantification of injection distribution was achieved through a combination of contrast agent injection and microCT scanning. A digital microinjection system was used to maximize reproducibility of injection volumes and control injection conditions, while a custom 3D-printed injection template was used to guide and deliver the contrast agent to tissue volumes of interest (VOIs). To first determine if the imaging strategy applied could evaluate injection distribution in our soft-tissue model, we quantified precise imaging thresholds required for microCT visualization of the injected contrast agent. We next applied this imaging strategy to quantify contrast agent in the VOI following either a distributed dosing protocol versus a single-bolus dose and compared these two approaches for their respective ability to provide drug coverage in the VOI.

Injection and microCT scanning

All injections were performed using a UMP3 UltraMicroPump coupled with a Micro4 controller (World Precision Instruments, Sarasota, USA) which aliquoted precise injection volumes at a rate of 10 µL/s. A 500-µL syringe (Trajan Scientific and Medical, Victoria, Australia) coupled with a 2″ 25-G beveled needle (Trajan) was used in all studies. Iodixanol, an iodinated x-ray contrast agent generally used for angiography, was chosen for microCT injection visualization. We used a 33% iodixanol 320 mg I/mL (Visipaque®, General Electric Healthcare, Chicago, USA) dilution (2:1 injectable saline to Visipaque 320; viscosity 5.5 cP) as a model injectable solution. All specimens were imaged using a vivaCT 40 scanner (Scanco Medical, Wanger-Brüttisellen, Switzerland) (voxel resolution 38 µm, 55 kVp, 145 µA, 200 ms). To provide an initial proof of concept for this technique, a portion of rabbit abductor muscle (5.6 cm3) was dissected from a freshly sacrificed rabbit, and a 60-µL bolus was injected. The specimen was imaged within 20 min and the distribution of the injection quantified by standard thresholding algorithms [36] (Fig. 1).

Fig. 1
figure 1

MicroCT rendering of a single injection. A 5.6-cm3 abductor muscle volume (light red) with a single 60-µL injection (arrow) oblique to the primary muscle fiber orientation of the sample (angling from upper right to lower left; blue line). The distributed volume was a three-dimensional spindle aligned with the muscle fiber orientation (dark red)

Specimen holder and prototype injection template

Injection distribution was quantified in porcine shoulder specimens. To minimize tissue handling postinjection, we developed a 3D-printed 4 × 4 × 4 cm3 inner dimension muscle specimen holder (64 cm3). The specimen holder was printed to enable capping with injection templates of various configurations and included a mounting attachment for microCT imaging (Fig. 2). For this study, we modeled a target VOI as a cubic box with a volume of 2 cm3 centered within the overall tissue volume (i.e., the VOI was 3.1% of total tissue volume). Precision injections were achieved using a custom template (forming the lid of the specimen holder) that provided 5 spaced locations in a quincunx pattern and a series of 3-printed spacers for each injection depth for a total of 15 injection sites. All 5 injections were performed consecutively at a single depth before moving to the next depth (moving superior to inferior within the specimen). The bolus injection was made using the centered port and middle depth.

Fig. 2
figure 2

Three-dimensional printed injection specimen holder. The injection specimen holder (gray, volume 64 cm3) enabled quantification of drug distribution strategies within a targeted volume of interest that was 3.1% of the total tissue volume (green, volume 2 cm3). Drug patterning was achieved by altering the injection port(s) orientation on the lid in combination with spacers that precisely controlled injection depth(s). The distributed dosing injection port pattern is visible on the lid

Tissue specimens were randomly assigned to one of three experimental groups (n = 8 per group): group 1 – a single-bolus injection targeted to the center of the VOI (bolus, total injection volume 450 µL); group 2 consisting of 15 injections, each 30-μL distributed throughout the VOI (Distributed); and group 3 naive controls (Uninjected). Immediately following completion of assigned injections, the entire test specimen holder was imaged with tissue specimen in place.

Quantification of fluid distribution in muscle

All high-resolution microCT images were acquired using a Scanco vivaCT 40. Specifically, a 38-µm voxel resolution scan was obtained of the center of the injection specimen holder to quantify both the VOI and a sufficient volume of surrounding tissue (cylindrical scan volume; diameter 3.7 cm, length 2.0 cm). To remove image noise and enhance the edge detection of injection volumes, all raw images were passed through a Gaussian filter (sigma 1.2, support 2.0), followed by identification of the targeted VOI in each scan.

To determine the imaging threshold required to visualize the contrast agent, we performed a parametric examination of the Uninjected group. From this analysis, a threshold of 258 hydroxyapatite (HA)/cm3 was identified as the minimum required to produce zero false-positive voxels in the Uninjected group (Fig. 3). This threshold was then applied to all samples to identify injection distribution within the targeted VOI (Fig. 4). Per standard image thresholding techniques, a visual inspection of the correspondence between two-dimensional binarized images and the original grayscale images was performed to confirm the validity of the threshold value chosen (Fig. 4). Given our previous experimental data suggesting that soft tissue within 1 mm of BTxA injection would be protected from HO formation [34], we next performed a 3D dilation on the segmented volumes (MATLAB, Fig. 4). This dilated volume was used to quantify our primary outcome measure, the effective drug coverage (i.e., the percentage of the targeted VOI that was within 1 mm of any voxel determined to contain injected fluid). As a secondary measure, we quantified off-target injected drug (i.e., dosing that is likely ineffective for this targeted application). We quantified injected fluid outside of concentric volumes larger than the VOI in 1 cm3 increments from 4 to 7 cm3 as a percentage of total injected fluid volume.

Fig. 3
figure 3

MicroCT threshold determination. Following image smoothing with a Gaussian filter, the volumes of interest of Uninjected specimens were segmented across a range of threshold levels to determine the lowest threshold capable of separating soft-tissue and injection of the iodine contrast agent. The global threshold of 258 hydroxyapatite (HA)/cm3 (red arrow) was chosen as it produced no false positives (i.e., soft tissue incorrectly identified as injected with contrast agent) and applied throughout the study

Fig. 4
figure 4

Image processing algorithm from image acquisition to quantification. Representative raw two-dimensional microCT images from Uninjected, bolus, and distributed dosing groups. The two-dimensional images are from the center of the volume of interest (VOI, white box). Standard microCT filtering and thresholding were applied to quantify drug distribution within each VOI (segmented). Finally, the drug coverage was quantified by three-dimensional dilation of the segmented and filtered image (dilated, < 1 mm)

Statistics

Two outcome measures were quantified: (1) effective drug coverage within the targeted VOI and (2) off-target fluid volume. One-way ANOVA were used to determine differences in effective drug coverage between Bolus, Distributed, and Uninjected groups. As homogeneity of variance was not observed between the groups, a nonparametric Kruskal-Wallis test with a post hoc Dunn test was performed. Similarly, differences in off-target fluid volumes in Bolus and Distributed dosing groups were determined by pairwise comparison using Mann-Whitney U-tests. The p-values lower than 0.05 were considered significant.

Results

Injection coverage within the targeted VOI was significantly increased in the Distributed group compared with the Bolus group (Fig. 5). Specifically, a bolus injection targeted at the center of the VOI produced a median injection drug coverage of 44% of the VOI (mean 43%, Fig. 5, bolus). Distributing the same injection volume via the prototype delivery template significantly increased the median injection coverage (median 75%, mean 72%, p = 0.031, Fig. 5, Distributed). Finally, bolus dose off-target fluid volume exceeded that of distributed dosing outside of a 7 cm3 region centered on the VOI (+86%, p = 0.041, Fig. 6).

Fig. 5
figure 5

Bolus versus distributed drug coverage. Drug coverage (red) within the volume of interest (VOI, blue box) was significantly enhanced by distributed dosing (a). To visualize the drug distribution within the VOI, the dilation is constrained by the boundaries of the VOI. Box and whisker plot (b) illustrating the drug coverage within the target VOI. Distributed dosing increased median coverage by 70% when compared to bolus dosing (*p = 0.031)

Fig. 6
figure 6

Off-target fluid volume. Box and whisker plots illustrating the percentage of injection outside of increasing concentric volumes centered about the 2-cm3 volume of interest (from 4 to 7 cm3). Bolus injections (orange) demonstrated elevated variability versus distributed dosing (red). As volume was increased, median escaped fluid volume was reduced by distributed dosing, reaching significance outside of 7 cm3 (*p = 0.041)

Discussion

The inability to visualize and quantify soft-tissue injection volumes at high resolution impairs the ability to rigorously explore how injection parameters (e.g., needle spacing, injection conditions) alter the precision and accuracy of drug distribution within complex muscular anatomy. We developed a microCT-based platform that overcomes this barrier. We assessed the feasibility of this approach by comparing on- and off-target fluid distribution for equal volume bolus versus distributed injections within a tissue volume that was 30 times larger than the targeted VOI. We observed that the 3D-printed prototype drug distribution device enhanced drug coverage within the VOI and reduced off-target injected fluid volume.

Given our interest in developing a platform that would be scalable at various stages of clinical translation (i.e., to large animal model and human cadaver testing), we created a 3D-printed radiolucent injection specimen holder that would enable rapid parametric assessment of injection distribution prototypes. We then used an automated threshold identification protocol to define the diffuse edges of the injection and developed postprocessing algorithms to quantify effective fluid coverage within the VOI and off-target injected fluid outside the VOI. The combination of microCT and an iodinated contrast agent enabled high-resolution visualizations of the injection fields. This was anticipated as similar approaches have been used in varied contexts, albeit with much reduced (3 × to 15 ×) image resolution [37, 38]. For example, visualization of bolus insulin injection in the subcutaneous fat layer of anesthetized pigs successfully quantified injection distribution variation in nonstriated tissue [37]. CT has also been used to produce intramuscular fiducial markers for tumor bed identification [38]. Unlike our study, the authors created a “sticky” injectable that minimally diffused in order to image tumor resection surfaces. In the context of these studies, our data provides further support for the potential of this approach to successfully image fluids of varied viscosity.

Our study has several limitations. First, we precisely quantified drug distribution in ex vivo specimens. Our goal is to use this technique to optimize the design of devices to inhibit HO in live animal and clinical applications. We recognize that the final distribution of injections will vary in vivo due to factors that alter drug profusion/absorption such as muscle disruption (by trauma or surgery) and muscle contraction. However, we believe that the insight gained through ex vivo studies (excised rabbit tissue for preclinical testing and cadaveric samples for clinical applications) will enable development of injection devices whose soft-tissue coverage profile will significantly exceed freehand injection. Ultimately, the proof of this strategy will occur not by imaging injection distribution in live tissue but also by correlating HO inhibition in the soft-tissue volumes we target for protection. Another limitation is that we quantified on-target and off-target injected fluid differently. We defined effective “drug” coverage within the VOI as on-target, as an optimal intervention will most effectively distribute the least required drug to achieve efficacy. We implemented a 1-mm dilation around any voxel identified as fluid based on preclinical preliminary studies confirming that the beneficial effect of the intervention was mitigated 4 mm away from the injected site [34]. While altering this parameter would alter absolute values, it is clear from visual image inspection that our platform readily identifies varied fluid distributions within a small VOI (Fig. 4). Off-target dosing (i.e., dosing that would not effectively inhibit HO in the targeted region) was reduced by the distributed dosing strategy. Although the relative difference between experimental conditions may have been influenced by our confined tissue test volume, we believe that the centered bolus dose provides a conservative comparison for the efficacy of distributed dosing. Finally, we used a single viscosity solution similar to blood (3.5–5.5 cP, [39]), which is higher than the injectable solution being modeled (BTxA in saline). We would anticipate that altering injection viscosity would alter absolute, but not relative results. Given the ability of our platform to image fluid of varied viscosity, future parametric studies will be conducted to confirm this thesis.

Our approach focused on developing a platform and outcome measures capable of detecting differences in injected fluid distribution at sufficient resolution to enable future optimization. We implemented a best-case bolus condition as the bolus injection was centered within the VOI (i.e., 0.63 cm from outer edge of the VOI). In contrast, 80% of the distributed prototype injections (12 of 15) were within 0.13 cm of an outer edge of the VOI. Our goal was not to optimize a distribution strategy, as we recognize that many variables of a drug delivery system must to be explored to achieve this objective (e.g., injection volume, injection speed, number of injections, spacing). However, these data emphasize that a distribution device is essential, as even an unoptimized prototype significantly improved effective drug coverage and reduced off-target injected fluid. Practically, this is a manifestation of a portion of the injected fluid migrating along muscle fiber orientation away from the targeted VOI, a phenomenon that is likely to be exacerbated within a complex surgical field with multiple muscle fiber orientations and surgical disruptions. This injection phenomenon highlights the difficulty inherent in using BTxA as HO prophylaxis, as targeting a variety of endplate zones in muscle will require a diffuse distribution of drug [35, 40] while seeking the minimize effective dose for drug safety considerations. Ultimately, our data suggest that although freehand manual delivery (as approximated by the spatially centered bolus injection) may randomly be accurate, the elevated variability of this strategy is not desirable. Distributed dosing will not only enhanced targeting precision (thus minimizing the required dose of BTxA), but it will also reduce the potential for excessive and/or off-target BTxA to induce undesirable musculoskeletal side effects [41, 42].

In summary, we have developed a microCT-based platform capable of quantifying drug distribution within muscle at a 38-µm voxel resolution. As the injection locations and depths of a 3D-printed template can be readily permutated, this platform will enable rapid optimization of injection parameters (e.g., port locations, dose volume, injection speed, and needle gauge) to most efficiently target a soft-tissue region where HO inhibition is desired while minimizing off-target injected fluid.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

3D:

Three-dimensional

BTxA:

Botulinum toxin type A

HO:

Heterotopic ossification

MicroCT:

Microcomputed tomography

NSAIDs:

Nonsteroidal anti-inflammatory drugs

VOI:

Volume of interest

References

  1. Foruria AM, Augustin S, Morrey BF, Sanchez-Sotelo J (2013) Heterotopic ossification after surgery for fractures and fracture-dislocations involving the proximal aspect of the radius or ulna. J Bone Joint Surg Am 95:e66. https://doi.org/10.2106/JBJS.K.01533

    Article  PubMed  Google Scholar 

  2. Potter BK, Burns TC, Lacap AP, Granville RR, Gajewski DA (2007) Heterotopic ossification following traumatic and combat-related amputations. Prevalence, risk factors, and preliminary results of excision. J Bone Joint Surg Am 89:476–486. https://doi.org/10.2106/JBJS.F.00412

    Article  PubMed  Google Scholar 

  3. Tippets DM, Zaryanov AV, Burke WV et al (2014) Incidence of heterotopic ossification in direct anterior total hip arthroplasty: a retrospective radiographic review. J Arthroplasty 29:1835–1838. https://doi.org/10.1016/j.arth.2014.04.027

    Article  PubMed  Google Scholar 

  4. Kaliyaperumal K, Sathappan SS, Peng LY (2008) Total hip arthroplasty for ankylosed hip secondary to heterotopic ossification. J Arthroplasty 23:470–475. https://doi.org/10.1016/j.arth.2007.03.041

    Article  PubMed  Google Scholar 

  5. Veltman ES, Lindenhovius AL, Kloen P (2014) Improvements in elbow motion after resection of heterotopic bone: a systematic review. Strategies Trauma Limb Reconstr 9:65–71. https://doi.org/10.1007/s11751-014-0192-0

    Article  PubMed  PubMed Central  Google Scholar 

  6. Salazar D, Golz A, Israel H, Marra G (2014) Heterotopic ossification of the elbow treated with surgical resection: risk factors, bony ankylosis, and complications. Clin Orthop Relat Res 472:2269–2275. https://doi.org/10.1007/s11999-014-3591-0

    Article  PubMed  PubMed Central  Google Scholar 

  7. Redmond JM, Keegan MA, Gupta A, Worsham JR, Hammarstedt JE, Domb BG (2017) Outcomes of heterotopic ossification excision following revision hip arthroscopy. J Hip Preserv Surg 4:164–169. https://doi.org/10.1093/jhps/hnx010

    Article  PubMed  PubMed Central  Google Scholar 

  8. Wu XB, Yang MH, Zhu SW et al (2014) Surgical resection of severe heterotopic ossification after open reduction and internal fixation of acetabular fractures: a case series of 18 patients. Injury 45:1604–1610. https://doi.org/10.1016/j.injury.2014.05.018

    Article  PubMed  Google Scholar 

  9. Chalidis B, Stengel D, Giannoudis PV (2007) Early excision and late excision of heterotopic ossification after traumatic brain injury are equivalent: a systematic review of the literature. J Neurotrauma 24:1675–1686. https://doi.org/10.1089/neu.2007.0342

    Article  PubMed  Google Scholar 

  10. Pavey GJ, Polfer EM, Nappo KE, Tintle SM, Forsberg JA, Potter BK (2015) What risk factors predict recurrence of heterotopic ossification after excision in combat-related amputations? Clin Orthop Relat Res 473:2814–2824. https://doi.org/10.1007/s11999-015-4266-1

    Article  PubMed  PubMed Central  Google Scholar 

  11. Chen S, Yu SY, Yan H et al (2015) The time point in surgical excision of heterotopic ossification of post-traumatic stiff elbow: recommendation for early excision followed by early exercise. J Shoulder Elbow Surg 24:1165–1171. https://doi.org/10.1016/j.jse.2015.05.044

    Article  PubMed  Google Scholar 

  12. Sagi HC, Jordan CJ, Barei DP, Serrano-Riera R, Steverson B (2014) Indomethacin prophylaxis for heterotopic ossification after acetabular fracture surgery increases the risk for nonunion of the posterior wall. J Orthop Trauma 28:377–383. https://doi.org/10.1097/BOT.0000000000000049

    Article  PubMed  Google Scholar 

  13. Popovic M, Agarwal A, Zhang L et al (2014) Radiotherapy for the prophylaxis of heterotopic ossification: a systematic review and meta-analysis of published data. Radiother Oncol 113:10–17. https://doi.org/10.1016/j.radonc.2014.08.025

    Article  PubMed  Google Scholar 

  14. Farris MK, Chowdhry VK, Lemke S, Kilpatrick M, Lacombe M (2012) Osteosarcoma following single fraction radiation prophylaxis for heterotopic ossification. Radiat Oncol 7:140. https://doi.org/10.1186/1748-717X-7-140

    Article  PubMed  PubMed Central  Google Scholar 

  15. Hamid N, Ashraf N, Bosse MJ et al (2010) Radiation therapy for heterotopic ossification prophylaxis acutely after elbow trauma: a prospective randomized study. J Bone Joint Surg Am 92:2032–2038. https://doi.org/10.2106/JBJS.I.01435

    Article  PubMed  Google Scholar 

  16. Freije SL, Kushdilian MV, Burney HN, Zang Y, Saito NG (2021) A retrospective analysis of 287 patients undergoing prophylactic radiation therapy for the prevention of heterotopic ossification. Adv Radiat Oncol 6:100625. https://doi.org/10.1016/j.adro.2020.11.010

    Article  PubMed  Google Scholar 

  17. Kan SL, Yang B, Ning GZ et al (2015) Nonsteroidal anti-inflammatory drugs as prophylaxis for heterotopic ossification after total hip arthroplasty: a systematic review and meta-analysis. Medicine (Baltimore) 94:e828. https://doi.org/10.1097/MD.0000000000000828

    Article  PubMed  Google Scholar 

  18. Karunakar MA, Sen A, Bosse MJ, Sims SH, Goulet JA, Kellam JF (2006) Indometacin as prophylaxis for heterotopic ossification after the operative treatment of fractures of the acetabulum. J Bone Joint Surg Br 88:1613–1617. https://doi.org/10.1302/0301-620X.88B12.18151

    Article  CAS  PubMed  Google Scholar 

  19. Matta JM, Siebenrock KA (1997) Does indomethacin reduce heterotopic bone formation after operations for acetabular fractures? A prospective randomised study. J Bone Joint Surg Br 79b:959–963. https://doi.org/10.1302/0301-620x.79b6.6889

    Article  Google Scholar 

  20. Bochat K, Mattin AC, Ricciardo BJ (2021) The efficacy of nonsteroidal anti-inflammatories in the prevention of heterotopic ossification following elbow trauma surgery. JSES Int 5:793–796. https://doi.org/10.1016/j.jseint.2021.04.004

    Article  PubMed  PubMed Central  Google Scholar 

  21. Beckmann JT, Wylie JD, Kapron AL, Hanson JA, Maak TG, Aoki SK (2014) The effect of NSAID prophylaxis and operative variables on heterotopic ossification after hip arthroscopy. Am J Sports Med 42:1359–1364. https://doi.org/10.1177/0363546514526361

    Article  PubMed  Google Scholar 

  22. Banovac K, Williams JM, Patrick LD, Haniff YM (2001) Prevention of heterotopic ossification after spinal cord injury with indomethacin. Spinal Cord 39:370–374. https://doi.org/10.1038/sj.sc.3101166

    Article  CAS  PubMed  Google Scholar 

  23. Bresalier RS, Sandler RS, Quan H et al (2005) Cardiovascular events associated with rofecoxib in a colorectal adenoma chemoprevention trial. N Engl J Med 352:1092–1102. https://doi.org/10.1056/NEJMoa050493

    Article  CAS  PubMed  Google Scholar 

  24. Board TN, Karva A, Board RE, Gambhir AK, Porter ML (2007) The prophylaxis and treatment of heterotopic ossification following lower limb arthroplasty. J Bone Joint Surg Br 89:434–440. https://doi.org/10.1302/0301-620X.89B4.18845

    Article  CAS  PubMed  Google Scholar 

  25. Dahners LE, Mullis BH (2004) Effects of nonsteroidal anti-inflammatory drugs on bone formation and soft-tissue healing. J Am Acad Orthop Surg 12:139–143

    Article  PubMed  Google Scholar 

  26. Kan L, Lounev VY, Pignolo RJ et al (2011) Substance P signaling mediates BMP-dependent heterotopic ossification. J Cell Biochem 112:2759–2772. https://doi.org/10.1002/jcb.23259

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Salisbury E, Rodenberg E, Sonnet C et al (2011) Sensory nerve induced inflammation contributes to heterotopic ossification. J Cell Biochem 112:2748–2758. https://doi.org/10.1002/jcb.23225

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  28. Salisbury E, Sonnet C, Heggeness M, Davis AR, Olmsted-Davis E (2010) Heterotopic ossification has some nerve. Crit Rev Eukaryot Gene Expr 20:313–324

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Davis EL, Davis AR, Gugala Z, Olmsted-Davis EA (2018) Is heterotopic ossification getting nervous?: The role of the peripheral nervous system in heterotopic ossification. Bone 109:22–27. https://doi.org/10.1016/j.bone.2017.07.016

    Article  PubMed  Google Scholar 

  30. Ausk BJ, Huber P, Poliachik SL, Bain SD, Srinivasan S, Gross TS (2012) Cortical bone resorption following muscle paralysis is spatially heterogeneous. Bone 50:14–22. https://doi.org/10.1016/j.bone.2011.08.028

    Article  PubMed  Google Scholar 

  31. Ausk BJ, Huber P, Srinivasan S et al (2013) Metaphyseal and diaphyseal bone loss in the tibia following transient muscle paralysis are spatiotemporally distinct resorption events. Bone 57:413–422. https://doi.org/10.1016/j.bone.2013.09.009

    Article  PubMed  Google Scholar 

  32. Ausk BJ, Worton LE, Smigiel KS et al (2017) Muscle paralysis induces bone marrow inflammation and predisposition to formation of giant osteoclasts. Am J Physiol Cell Physiol 313:C533–C540. https://doi.org/10.1152/ajpcell.00363.2016

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  33. Ausk BJ, Gross TS, Bain SD (2015) Botulinum toxin-induced muscle paralysis inhibits heterotopic bone formation. Clin Orthop Relat Res 473:2825–2830. https://doi.org/10.1007/s11999-015-4271-4

    Article  PubMed  PubMed Central  Google Scholar 

  34. Ausk BJ, Bain SD, Gross TS (2020) Methods for inhibiting heterotopic ossification. UNIV WASHINGTON, US Patent 10,835,689, 17 Nov 2020

  35. Elwischger K, Kasprian G, Weber M et al (2014) Intramuscular distribution of botulinum toxin–visualized by MRI. J Neurol Sci 344:76–79. https://doi.org/10.1016/j.jns.2014.06.028

    Article  CAS  PubMed  Google Scholar 

  36. Bouxsein ML, Boyd SK, Christiansen BA, Guldberg RE, Jepsen KJ, Muller R (2010) Guidelines for assessment of bone microstructure in rodents using micro-computed tomography. J Bone Miner Res 25:1468–1486. https://doi.org/10.1002/jbmr.141

    Article  PubMed  Google Scholar 

  37. Thomsen M, Poulsen M, Bech M et al (2012) Visualization of subcutaneous insulin injections by x-ray computed tomography. Phys Med Biol 57:7191–7203. https://doi.org/10.1088/0031-9155/57/21/7191

    Article  CAS  PubMed  Google Scholar 

  38. Steybe D, Poxleitner P, Voss PJ et al (2021) Evaluation of computed tomography settings in the context of visualization and discrimination of low dose injections of a novel liquid soft tissue fiducial marker in head and neck imaging. BMC Med Imaging 21:157. https://doi.org/10.1186/s12880-021-00689-y

    Article  PubMed  PubMed Central  Google Scholar 

  39. Nader E, Skinner S, Romana M et al (2019) Blood rheology: key parameters, impact on blood flow, role in sickle cell disease and effects of exercise. Front Physiol 10:1329. https://doi.org/10.3389/fphys.2019.01329

    Article  PubMed  PubMed Central  Google Scholar 

  40. Hallett M (2015) Explanation of timing of botulinum neurotoxin effects, onset and duration, and clinical ways of influencing them. Toxicon 107:64–67. https://doi.org/10.1016/j.toxicon.2015.07.013

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  41. Aliprantis AO, Stolina M, Kostenuik PJ et al (2012) Transient muscle paralysis degrades bone via rapid osteoclastogenesis. FASEB J 26:1110–1118. https://doi.org/10.1096/fj.11-196642

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  42. Warner SE, Sanford DA, Becker BA, Bain SD, Srinivasan S, Gross TS (2006) Botox induced muscle paralysis rapidly degrades bone. Bone 38:257–264. https://doi.org/10.1016/j.bone.2005.08.009. (S8756-3282(05)00340-6 [pii])

    Article  CAS  PubMed  Google Scholar 

Download references

Funding

STTR (AR074337, PI: JG, Subcontract PI: SB), Sigvard T. Hansen Jr. Endowed Chair (TG), and Zimmer Fracture Biology Professorship (SB).

Author information

Authors and Affiliations

Authors

Contributions

All authors made substantial contributions to the conception of this work. BA, AT, PH, SB, and TG designed the work. Data acquisition was done by BA, AT, and PH, while data interpretation was performed by BA, PH, SB, and TG. BA and PH created new software used in this work. Finally, BA, PH, SB, and TG drafted the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Brandon J. Ausk.

Ethics declarations

Ethics approval and consent to participate

All animal experiments were performed using protocols and procedures approved by the Institutional Animal Care and Use Committee of the University of Washington.

Consent for publication

Not applicable

Competing interests

BA, JG, TG, and SB are co-founders of the In Situ Therapeutic Solutions Inc. The other authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ausk, B.J., Tucker, A.N., Huber, P. et al. A microCT-based platform to quantify drug targeting. Eur Radiol Exp 7, 38 (2023). https://doi.org/10.1186/s41747-023-00355-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-023-00355-8

Keywords