Strategies for calculating contrast media dose for chest CT

Abstract

Background

Total body weight (TBW) is a frequently used contrast media (CM) strategy for dose calculation in enhanced CT, yet it is suboptimal as it lacks consideration of patient characteristics, such as body fat percentage (BFP) and muscle mass. Alternative CM dosage strategies are suggested by the literature. Our objectives were to analyze the CM dose impact when adjusting to body composition using methods of obtaining lean body mass (LBM) and body surface area (BSA) along with its correlation with demographic factors in contrast enhanced chest CT examinations.

Methods

Eighty-nine adult patients referred for CM thoracic CT were retrospectively included, categorized as either normal, muscular, or overweight. Patient body composition data was used to calculate the CM dose according to LBM or BSA. LBM was calculated with the James method, Boer method, and bioelectric impedance (BIA). BSA was calculated using the Mostellar formula. We then correlated the corresponding CM doses with demographic factors.

Results

BIA demonstrated the highest and lowest calculated CM dose in muscular and overweight groups respectively, compared to other strategies. For the normal group, the lowest calculated CM dose was achieved using TBW. The calculated CM dose was more closely correlated with BFP using the BIA method.

Conclusions

The BIA method is more adaptive to variations in patient body habitus especially in muscular and overweight patients and is most closely correlated to patient demographics. This study could support utilizing the BIA method for calculating LBM for a body-tailored CM dose protocol for enhanced chest CT examinations.

Relevance statement

The BIA-based method is adaptive to variations in body habitus especially in muscular and overweight patients and is closely correlated to patient demographics for contrast-enhanced chest CT.

Key points

• Calculations based on BIA showed the largest variation in CM dose.

• Lean body weight using BIA demonstrated the strongest correlation to patient demographics.

• Lean body weight BIA protocol may be considered for CM dosing in chest CT.

Graphical Abstract

Background

Image quality and lesion detection in computed tomography (CT) scans of the chest may be improved with the use of contrast media (CM). Traditionally, enhanced CT protocols have used fixed CM doses [1]; however, personalized CM approaches have been shown to be nearly as effective as personalized approaches to patient care [2, 3]. While iodine concentration, injection rate, scan delay, blood pressure and cardiac function can affect contrast enhancement, so too can body composition, which impacts distribution of CM in extracellular spaces due to differences in blood volume and flow [4, 5]. As a result, new strategies have emerged aiming to customize CM dose to both total body weight (TBW) [1, 6] and body composition [7]. Linear weight-based approaches, however, do not accurately estimate ideal enhancement as they fail to take into consideration patient characteristics, such as body fat percentage (BFP) and muscle mass [6, 8].

Literature documents numerous methods to customize CM doses, however none are recognized as standard or preferred [1, 8]. Lean body mass (LBM), defined as the difference between TBW and body fat weight, is suggested as one appropriate CM administration strategy [9]. CM dose based on LBM decreases CM doses while consistently improving image quality, as it corrects for high body fat percentage (BFP) by correlating with extracellular fluid volumes [10]. LBM can be calculated using the James method, the Boer method, and bioelectric impedance analysis (BIA) [6, 9, 11]. For obese patients, the Boer formula is more consistent and reliable [7] as the James method underestimates CM doses in this population and is therefore not appropriate in patients with body mass index (BMI) above approximately 40 kg/m2 [6, 9, 11]. BIA is considered the standard for evaluating body composition and BFP, measuring conduction of an electrical current through intracellular and extracellular water [6, 12]; however, it relies on constant patient hydration and may be inaccurate if a patient is NPO (Nothing by Mouth) prior to scanning [12]. Iodine dose may also be adjusted by measuring body surface area (BSA) [1], and when compared to LBM methods, it may be highly applicable clinically to adjust CM dose across a range of body sizes and weights [1]. Research shows that BSA calculated using the Mosteller formula is a strong indicator for metabolic mass [1]. However, controversy exists regarding BSA calculations as using actual weight-based calculations compared to ideal TBW may overestimate CM dose by 15–30% [13].

With a global shortage of iodinated CM in recent years [14], it is necessary to modify thinking with respect to how CM can most effectively be utilized [15,16,17]. Consequently, there is an increased focus on the reduction and personalization of CM doses to ensure adequate image quality. Current research on patient-tailored CM dosage focuses mostly on abdominal protocols [1, 4, 9, 18] without examining the impact of these methods in evaluation of the chest. In this setting, we aimed to investigate the impact of body composition data on various CM dose calculations.

Methods

Population

The present study uses patient data from a larger prospective study evaluating the CM dose used in contrast enhanced thoracic CT examinations. The larger study was approved by the Norwegian Regional Medical Research Ethics Committee (No. 2016/674) and Data Protection Officer at the institution. Informed consent was waived from patients referred for various chest CT examinations. The patients were randomized to receive CM according to either fixed or weight/body composition tailored CM approach. For the current study, the dataset from only weight/body composition tailored CM approach was used and the patient dataset was prospectively collected between 2019 and 2021 at a University Hospital in Norway for the CM dose calculation. All CT chest examinations were performed on the same scanner and had identical patient positioning, scan delay and CM was Iohexol 350 mgI/mL (Omnipaque, GE Healthcare, Oslo, Norway). The injection duration was 18 s, and consequently, the flow rate was manually adjusted individually for the included patients. CM injection was followed by a 40-mL saline flush at the same flow rate. Exclusion criteria were hemodynamic instability, cardiac failure, pacemaker, renal insufficiency (estimated glomerular filtration rate < 30 mL/min/1.73 m2), contraindications to CM, and age < 18 years. As this dataset was only utilized to calculate potential CM doses for TBW, LBM, and BSA approaches, no evaluation of image quality was performed.

Eighty-nine patients were included and categorized as either normal, muscular or overweight based on pre-defined objective and subjective criteria (Fig. 1). Criteria for inclusion in the normal group was a BMI < 25. Participants 30-years-old and younger with a BMI < 25 and waist circumference smaller than the overweight category were included in the muscular group. For patients > 30 years, a subjective assessment was performed to determine if they belonged to the muscular or other two groups. Participants were included in the overweight category if their BMI was ≥ 25 with a waist circumference ≥ 88 cm for females and ≥ 102 cm for males. The assessment was performed by a restricted number of specialized CT technologists/radiographers with experience and training to perform subjective assessment for the study [19].

Fig. 1
figure 1

Schematic illustration of the selected criteria for determining body composition. BMI, Body mass index

Participants' age, TBW, height, BMI, waist circumferences, and BFP were recorded. The BFP was measured using a foot-to-foot bioelectrical impedance analyzer (BIA) (Tanita® mod. DC-430MA, Tanita Corp., Tokyo, Japan). The test took approximately 2−3 min, including entering the information into the software and generating the report.

Patient allometric parameters

Patient allometric parameters were calculated using the formulas in Table 1. LBM reflects the central blood volume, extracellular fluid space of well-perfused tissues, and the small volume of the extracellular fluid space of poorly perfused tissues, such as the skeletal system [20]. LBM calculations using the James and Boer formulas estimate CM dose by considering the patient’s height, weight, and gender, unlike traditional TBW protocols [11]. Calculating LBM using the BIA formula accounts for BFP rather than height and does not differentiate between gender [12]. BSA (m2) was calculated using the Mosteller formula, which estimates body surface using height and weight [1] and is also independent of gender. Used often in clinical practice to correct drug dosage based on body size and as an alternative to the LBM methods, it is easily memorized and can be calculated on a handheld calculator [1, 20].

Table 1 Equations for calculating allometric parameters using three different LBM and BSA contrast medium strategies

CM dose calculations

Calculation of CM doses based on alternative allometric parameters described in Table 1 were achieved using the CM dose administered in mL to patients from the original data set. The CM dose was initially determined based on findings in a survey study, where CM dosages in chest CT ranged from 0.6 to 2.0 mL/kg with a mean and median CM dose of 1.2 mL/kg and 1.3 mL/kg TBW, respectively [21]. The total amount of CM dose (8,560 mL) received by the 89 included patients in the earlier larger prospective study, but this time independent of their body composition, was divided by TBW (6,967 kg), LBM James (5,064 kg), LBM Boer (5,067 kg), LBM BIA (4,997 kg) and BSA (173 m2) based on equations shown in Table 1. This resulted in 1.2 mL/kg using TBW, 1.7 mL/kg using LBM for the three LBM methods, and 49.0 mL/m2 for BSA. These resultant calculations were further used to calculate the CM dose based on the three included CM dose calculation strategies in the present retrospective study.

Statistical analysis

Descriptive statistics for CM doses are expressed as means ± standard deviations with ranges. Normality was tested using graphical assessment together with the Shapiro–Wilk test. Using Pearson’s correlation coefficient r, correlations between calculated CM doses that patients would have received for each CM strategy and their TBW, height, BMI, waist circumference, and BFP, were calculated overall and separately for each body composition. Correlations were considered weak (r, -0.00 to 0.40 or 0.00 to -0.40), moderate (r, 0.40 to 0.80 or -0.40 to -0.80), or strong (r, 0.8 to 1.0 or -0.8 to -1.00); p values < 0.05 were considered statistically significant for paired t-tests. Excel Version 16.59 (Microsoft, Redmond, WA, USA) and STATA statistical software Version 16.0 (Stata Corp, College Station, TX, USA) were used for all calculations. A statistician was consulted, and analysis was performed accordingly.

Results

Population

The population of the study consisted of a total of 89 patients categorized as either normal (n = 31), muscular (n = 28), or overweight (n = 30). Descriptive and allometric parameters from the patient dataset in this study are shown in Table 2. Overall, average patient TBW, height, and BMI were 78 ± 18 kg, 175 ± 9 cm, and 25 ± 5 kg/m2, respectively. BMI and waist size varied between 14 to 40 kg/m2, and 64 to 131 cm, respectively.

Table 2 Descriptive statistics of demographic factors by body habitus presented as mean ± standard deviation with ranges

Calculated CM dose

Table 3 presents calculated CM doses according to normal, muscular, and overweight patient categories. Differences between mean CM dose (mL/kg) for each body composition using the included CM dose strategies in the present study are shown in Fig. 2. These differences were statistically significant (p < 0.050) for the three body compositions using TBW versus LBM and BSA strategies. Furthermore, differences in the normal group using LBM BIA versus LBM James (1.26 ± 0.10 mL/kg versus 1.33 ± 0.08 mL/kg; p = 0.005) and LBM Boer (1.26 ± 0.10 mL/kg versus 1.34 ± 0.12 mL/kg; p = 0.004) were statistically significant. In the muscular group a statistically significant (p = 0.003) difference was observed when comparing LBM BIA and BSA. Calculated CM dose was higher in muscular patients when using LBM BIA when compared to BSA as well as LBM James and LBM Boer (Table 3). Calculated CM doses (mL/kg) for all three-body compositions were, as expected, identical using the TBW strategy while a significant difference was observed with all other strategies included in this study (Table 3).

Table 3 Calculated contrast medium doses for each CM strategy in mL/kg and mL for the defined body compositions
Fig. 2
figure 2

Box plot showing the calculated CM dose (mL/kg) for each CM strategy by body composition. Plot shows the mean ± standard deviation, the minimum and maximum values. The box represents 25th and 75th percentile values, and the whiskers indicate 1.5 times the interquartile range. Outlier values are plotted with dots. * = p < 0.05 between LBM BIA and other LBM strategies, including James versus BIA (p = 0.005) and Boer versus BIA (p = 0.004) for normal size patient group and BIA versus BSA (p = 0.003) for muscular patient group. BIA, Bioelectric impedance analysis; BSA, Body surface area; CM, Contrast medium; LBM, Lean body mass

Effect of patient demographics on calculated CM dose

Correlational analysis using Pearson’s correlation coefficient was performed to investigate the impact of different patient characteristics on CM dose for each body composition, according to TBW, LBM (James, Boer, and BIA) and BSA, as shown in Table 4. There were significant differences (p < 0.05 for all) in correlation between calculated CM doses and several demographic factors as shown in Table 4. The table also shows a slightly stronger negative correlation for BFP using BIA than the other LBM calculations. These differences had a statistically significant impact (p < 0.05) for normal and overweight groups using LBM BIA (p = 0.03 and p = 0.00, respectively) and only for the overweight group using LBM James and Boer (p = 0.01 and p = 0.03 respectively).

Table 4 Pearson’s correlation coefficients (r) and p values of demographic factors and calculated contrast medium dose

Discussion

This study aimed to evaluate which method of CM dose calculation (TBW, three LBM CM dose calculation methods, and BSA) would be the most appropriate approach to be used in a larger future prospective study evaluating contrast enhanced thoracic CT examinations. The evaluation was performed based on CM dose calculation according to several methods and analyzed the impact of demographic factors and body habitus in an existing dataset, including three different body compositions.

Calculated CM doses for each body habitus group produced similar results using the James, Boer, and BSA methods (Tables 3 and 4). Among LBM strategies, the calculated CM dose was lower in the normal group (1.26 mL/kg) and overweight group (1.08 mL/kg) and higher in muscular group (1.37 mL/kg) using BIA compared to James and Boer as shown in Table 3, demonstrating BIA had the largest variation in CM dose according to body composition. When compared to the normal group and using BIA, the muscular group received higher CM dose by 11% (0.11 mL/kg) (p = 0.00) and lower than 14% (0.18 mL/kg) for overweight participants (p = 0.00). Clinically, using the BIA method, a 90-kg patient would receive 113 mL, 123 mL, and 97 mL of CM having a normal, muscular, or overweight body habitus, respectively, showing a higher variability between the three body habitus groups when compared to James (120, 120, and 100 mL) and Boer (121, 120, and 100 mL) methods as well as TBW (90 mL for all three groups). This greater variation in estimated CM dose using BIA is due to the larger variation in BFP% between the body composition categories.

The use of personalized CM doses has been reported with varied results concerning both CM dose and CM enhancement. While some studies have demonstrated minimal variation between CM dose calculations [5, 22], others have demonstrated higher variations [8, 20]. Zanardo et al. [22] found comparable image quality using TBW and LBM. A limitation of this study, however, was that it did not include patients with a varied body habitus. When a more diverse population is included, variation in CM enhancement has been reported [7, 8]. Differences in CM enhancement have also been reported using different LBM strategies [6, 11]. Rengo et al. [6] found that measurements of LBM using BIA were optimal for tailoring CM dose for hepatic CT exams. As the BIA method better reflects the variations in patient body habitus, it accounts for an increase in CM needed for highly vascular parenchymal organs within muscular patients, and a decrease in CM required to diffuse adipose tissue in overweight patients [1, 6, 8]. Clinically, due to this higher vascularization [10], muscles are perfused with proportionally more blood and consequently more contrast medium than adipose tissue. Therefore, this strategy may better account for this difference, resulting in a more homogenous enhancement regardless of body composition. It is also known that LBM strategies correct for high BFP by correlating with extracellular fluid volumes [4, 6, 12]. This is confirmed in our results that show a higher BFP for normal and overweight groups, as reported in Table 2. Another benefit of using BIA is that it accounts for only weight and BFP [12] compared to James and Boer that include height, weight and gender [11].

It is important to note that no contrast enhancement was analyzed and included in our study, as the CM calculations and analysis were performed retrospectively. Contrast enhancement will be included in the future larger prospective clinical study.

The correlational analysis of patient demographic factors produced some expected results. The ratio between weight and CM dose for the TBW approach was 1.0, which was predictable given that weight was the determinant factor in the strategy. Similarly, the strong correlation between patient weight and height with CM dose when the James, Boer, and BSA methods were used was predictable, as these factors were directly used in the calculations presented in Table 1. While the BIA formula does not utilize patient height, there remained a strong correlation between height and CM dose for this strategy, independent of body habitus. This finding was intuitively expected due to the proportional relationship between weight and height (i.e., taller people generally weigh more than shorter people) [8]. Of the LBM methods, the BIA method had the strongest statistically significant inverse relationship between BFP and CM dose, as expected from the equation of LBM BIA. Therefore, patients with a higher BFP would receive lower CM dose. This finding supports using BIA, as it does not overestimate CM doses for overweight patients with low metabolic mass. The BIA method also had weaker correlations than other included strategies between both BMI and waist size to CM dose. This is ideal given that BMI is a poor estimate of total body fat content in muscular people with high BMIs due to their muscle density [8]. In this sense, the James formula is shown to be less accurate in obese patients with BMI ≥ 37 and ≥ 43 kg/m2 for female and male, respectively, as estimates reach a level where the estimated LBM decreases with increasing BMI [11, 23]. However, the current study included only one patient with BMI 39.5 kg/m2 and is therefore limited in ability to investigate the impact of using the James or Boer methods. Moreover, formulas cannot accurately substitute for quantitative imaging or measurements [24]. This is especially important when analysis focuses on measurements of muscle mass or changes in body composition. Chamchod et al. [24] concluded that CT based LBM assessments were more reliable than formulas in cancer patients at risk for illness and therapy associated changes in body composition. With regard to CM enhancement and increased vascularization in muscle tissue, the use of BIA could be of benefit to discriminate high muscle mass subjects from obese ones, especially in patients with comparable body size measures like BMI.

Our study has some limitations. Our calculations were based on predefined CM doses from a series of patients who received the standard CM dose in clinical practice. Also, this study used only theoretical calculations which have not yet been supported by image analysis to confirm our results. However, we expect that the prospective ongoing study confirms our hypothesis. Thirdly, the use of BIA analyzer could carry some known inaccuracy and reproducibility issues; however, the scientifically validated analyzer was used with a standardized procedure and the technique itself has been validated in previous studies [25,26,27]. Lastly, a subjective analysis of only 16 muscular patients older than 30 years old might result in less accurate assignment of body composition category, however, the technologists involved in the patient recruitment were trained carefully, and thus any negative impact on patient categorization would be limited. Moreover, BFP results suggest the categorization is adequate.

In conclusion, our estimations of CM dose based on various CM dose calculations showed that there were significant differences between TBW and all the included more compound CM dosing strategies. The BIA method demonstrated a closer correlation to various patient demographics included in the current study and determined greater differences, especially in muscular and overweight patients. To fully investigate the clinical impact of different dose calculation methods on various body compositions, further studies need to be carried out comparing variations in contrast dosage depending on body habitus, assessing both contrast enhancement and image quality.

Availability of data and materials

The dataset is available from the corresponding author on reasonable request.

Abbreviations

BFP:

Body fat percentage

BIA:

Bioelectric impedance analysis

BMI:

Body mass index

BSA:

Body surface area

CM:

Contrast medium

CT:

Computed tomography

TBW:

Total body weight

References

  1. Walgraeve MS, Pyfferoen L, Van De Moortele K, Zanca F, Bielen D, Casselman JW (2019) Implementation of patient-tailored contrast volumes based on body surface area and heart rate harmonizes contrast enhancement and reduces contrast load in small patients in portal venous phase abdominal CT. Eur J Radiol 121:108630. https://doi.org/10.1016/j.ejrad.2019.07.031

    Article  PubMed  Google Scholar 

  2. Sardanelli F (2017) Trends in radiology and experimental research. Eur Radiol Exp 1:1. https://doi.org/10.1186/s41747-017-0006-5

    Article  PubMed  PubMed Central  Google Scholar 

  3. Bae KT (2010) Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology 256:32–61. https://doi.org/10.1148/radiol.10090908

    Article  PubMed  Google Scholar 

  4. Svensson A, Nouhad J, Cederlund K et al (2012) Hepatic contrast medium enhancement at computed tomography and its correlation with various body size measures. Acta Radiol 53:601–606. https://doi.org/10.1258/ar.2012.120268

    Article  PubMed  Google Scholar 

  5. Eijsvoogel NG, Hendriks BMF, Nelemans P et al (2020) Personalization of CM injection protocols in coronary Computed Tomographic Angiography (People CT Trial). Contrast Media Mol Imaging 2020:5407936. https://doi.org/10.1155/2020/5407936

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Rengo M, Bellini D, Businaro R et al (2017) MDCT of the liver in obese patients: evaluation of a different method to optimize iodine dose. Abdom Radiol (NY) 42:2420–2427. https://doi.org/10.1007/s00261-017-1156-x

    Article  PubMed  Google Scholar 

  7. Zanardo M, Doniselli FM, Esseridou A et al (2018) Abdominal CT: a radiologist-driven adjustment of the dose of iodinated contrast agent approaches a calculation per lean body weight. Eur Radiol Exp 2:41. https://doi.org/10.1186/s41747-018-0074-1

    Article  PubMed  PubMed Central  Google Scholar 

  8. Bae KT, Seeck BA, Hildebolt CF et al (2008) Contrast enhancement in cardiovascular MDCT: effect of body weight, height, body surface area, body mass index, and obesity. AJR Am J Roentgenol 190:777–784. https://doi.org/10.2214/ajr.07.2765

    Article  PubMed  Google Scholar 

  9. Janek E, Edl D, Taflinski A et al (2021) Contrast media dose adjustment to allometric parameters of body mass in multiphasic CT of the liver: a comparison of different metrics. Eur J Radiol 141:109778. https://doi.org/10.1016/j.ejrad.2021.109778

    Article  PubMed  Google Scholar 

  10. Hochhegger B, Rottenfusser R, Marchiori E (2017) When is the use of contrast media in chest CT indicated? J Bras Pneumol 43(5):400. https://doi.org/10.1590/s1806-37562017000000179

    Article  PubMed  PubMed Central  Google Scholar 

  11. Caruso D, De Santis D, Rivosecchi F et al (2018) Lean body weight-tailored iodinated contrast injection in obese patient: Boer versus James formula. Biomed Res Int 2018:8521893. https://doi.org/10.1155/2018/8521893

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Lemos T, Gallagher D (2017) Current body composition measurement techniques. Curr Opin Endocrinol Diabetes Obes 24:310–314. https://doi.org/10.1097/med.0000000000000360

    Article  PubMed  PubMed Central  Google Scholar 

  13. Faisal W, Tang HM, Tiley S, Kukard C (2016) Not all body surface area formulas are the same, but does it matter? J Glob Oncol 2:436–437. https://doi.org/10.1200/jgo.2016.005876

    Article  PubMed  PubMed Central  Google Scholar 

  14. U.S. Food and Drug Administration. FDA Drug Shortages. https://www.accessdata.fda.gov/scripts/drugshortages/default.cfm. Accessed 20 Feb 2023.

  15. Cavallo JJ, Pahade JK (2022) Practice management strategies for imaging facilities facing an acute iodinated contrast media shortage. AJR Am J Roentgenol 219:666–670. https://doi.org/10.2214/ajr.22.27969

    Article  PubMed  Google Scholar 

  16. Eibschutz LS, Gholamrezanezhad A (2022) How low can we go? Strategies and recommendations to combat the iodinated contrast shortage. Emerg Radiol 29:925–928. https://doi.org/10.1007/s10140-022-02077-7

    Article  PubMed  PubMed Central  Google Scholar 

  17. Barreto IL, Grajo JR, Brock DE, Magnelli L, Patel P, Hochhegger B (2022) A health system’s response to the ongoing global shortage of iodinated contrast media. AJR Am J Roentgenol 219:974–979. https://doi.org/10.2214/ajr.22.28060

    Article  PubMed  Google Scholar 

  18. Yamashita Y, Komohara Y, Takahashi M et al (2000) Abdominal helical CT: evaluation of optimal doses of intravenous contrast material–a prospective randomized study. Radiology 216:718–723. https://doi.org/10.1148/radiology.216.3.r00se26718

    Article  CAS  PubMed  Google Scholar 

  19. Dastan K, Henning MK, England A, Aalokken TM, Johansen S (2021) An investigation into the variability of radiographers assessing body composition prior to CT contrast media administration. Radiography (Lond) 27:168–172. https://doi.org/10.1016/j.radi.2020.07.012

    Article  CAS  PubMed  Google Scholar 

  20. Kidoh M, Nakaura T, Oda S et al (2013) Contrast enhancement during hepatic computed tomography: effect of total body weight, height, body mass index, blood volume, lean body weight, and body surface area. J Comput Assist Tomogr 37:159–164. https://doi.org/10.1097/RCT.0b013e31827dbc08

    Article  PubMed  Google Scholar 

  21. Henning MK, Aaløkken TM, Johansen S (2021) Contrast medium protocols in routine chest CT: a survey study. Acta Radiol 284185121997111 https://doi.org/10.1177/0284185121997111

  22. Zanardo M, Doniselli FM, Esseridou A et al (2020) Lean body weight versus total body weight to calculate the iodinated contrast media volume in abdominal CT: a randomised controlled trial. Insights Imaging 11:132. https://doi.org/10.1186/s13244-020-00920-4

    Article  PubMed  PubMed Central  Google Scholar 

  23. Nyman U (2016) James lean body weight formula is not appropriate for determining CT contrast media dose in patients with high body mass index. Radiology 278:956–957. https://doi.org/10.1148/radiol.2016152031

    Article  PubMed  Google Scholar 

  24. Chamchod S, Fuller CD, Mohamed AS et al (2016) Quantitative body mass characterization before and after head and neck cancer radiotherapy: a challenge of height-weight formulae using computed tomography measurement. Oral Oncol 61:62–69. https://doi.org/10.1016/j.oraloncology.2016.08.012

    Article  PubMed  PubMed Central  Google Scholar 

  25. Segal KR, Van Loan M, Fitzgerald PI, Hodgdon JA, Van Itallie TB (1988) Lean body mass estimation by bioelectrical impedance analysis: a four-site cross-validation study. Am J Clin Nutr 47:7–14. https://doi.org/10.1093/ajcn/47.1.7

    Article  CAS  PubMed  Google Scholar 

  26. Ward LC (2019) Bioelectrical impedance analysis for body composition assessment: reflections on accuracy, clinical utility, and standardisation. Eur J Clin Nutr 73:194–199. https://doi.org/10.1038/s41430-018-0335-3

    Article  PubMed  Google Scholar 

  27. Marra M, Sammarco R, De Lorenzo A et al (2019) Assessment of body composition in health and disease using Bioelectrical Impedance Analysis (BIA) and Dual Energy x-ray Absorptiometry (DXA): a critical overview. Contrast Media Mol Imaging 2019:3548284. https://doi.org/10.1155/2019/3548284

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors thank Kendra Underhill and Sydney Trenholm for their contribution and cooperation.

Funding

None. The authors received no financial support.

Author information

Authors and Affiliations

Authors

Contributions

MKH carried out data collection. JA participated in review and revision of the manuscript. CG participated in review of the manuscript. SJ led the research team, drafted the manuscript in collaboration with MKH, contributed to data analysis, carried out the study design, contributed to manuscript finalization, and contributed to the revision. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Safora Johansen.

Ethics declarations

Ethics approval and consent to participate

This study utilized a patient dataset collected during the period of 2019–2021 at Oslo University Hospital. The study was approved by the Norwegian Regional Medical Research Ethics Committee (No. 2016/674) and Data Protection Officer at the institution. Informed consent was waived from patients. Trial registration: ClinicalTrials.gov, Identifier no. NCT05645796, registered on the 9th December 2022.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Henning, M.K., Gunn, C., Arenas-Jiménez, J. et al. Strategies for calculating contrast media dose for chest CT. Eur Radiol Exp 7, 29 (2023). https://doi.org/10.1186/s41747-023-00345-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-023-00345-w

Keywords