Feasibility of aortic aneurysm sac embolization using a novel shape memory polymer embolic device

Abstract

Background

We investigated the feasibility of aneurysm sac embolization using a novel self-expanding porous shape memory polymer (SMP) device during endovascular aortic abdominal or thoracic aneurysm repair (EVAR).

Methods

Retrospective analysis of consecutive patients treated at 2 centers in Germany. Patients were treated from January 2019 to July 2021 with follow-up at 7 days and 3, 6, and 12 months. Aneurysm sacs were implanted with SMP devices immediately following endograft placement during the same procedure. Primary endpoint was technically successful SMP-device deployment into the aneurysm sac outside the endograft. Secondary endpoints were changes in aneurysm volume and associated complications (e.g., endoleaks).

Results

We included 18 patients (16 males), aged 72 ± 9 years, achieving 100% technical success. Mean preprocedure aortic aneurysm sac volume was 195 ± 117 mL with a perfused aneurysm volume of 97 ± 60 mL. A mean of 24 ± 12 SMP devices per patient were used (range 5–45, corresponding to 6.25–56.25 mL expanded embolic material volume). All evaluable patients exhibited sac regression except 2 patients yet to reach 3-month follow-up. At mean 11 ± 7 months (range 3–24), change in aneurysm volume from baseline was -30 ± 21 mL (p < 0.001). In 8 patients, aneurysm regression was observed despite type 2 endoleaks in 6 and type 1A endoleaks in 2, none of them requiring further intervention to date. No morbidity or mortality related to this treatment occurred.

Conclusions

SMP devices for aortic aneurysm sac embolization during endovascular repair appear feasible and safe in this small case series. Prospective studies are needed.

Key points

• Shape memory polymer is a novel, self-expanding, porous, and radiolucent embolic device material.

• Aortic aneurysm sacs were treated with polymer devices immediately following endograft placement.

• Aortic aneurysm sac regression was observed in all patients with over 3-month follow-up.

• Aortic aneurysm sac regression was observed even in the presence of endoleaks.

Background

Abdominal aortic aneurysm (AAA) sac failure to regress after endovascular aneurysm repair (EVAR) has been linked to increased long-term mortality [1, 2]. It is widely acknowledged that endoleaks contribute to sac expansion or failure to regress [3]. However, an absence of endoleak is not necessarily predictive of sac stability or regression [1, 2], and AAA sac management during and after EVAR remains a subject of substantial investigation.

Endoleak management, particularly for type 2, falls into two types, active and reactive, where the latter is currently more common practice. Reactive endoleak management options include a variety of techniques to reestablish endograft seals and/or aneurysm sac and/or branch vessel embolization techniques. Post-EVAR reactive sac/vessel embolization methods include transarterial and translumbar (direct injection) approaches. Embolic devices commonly used to embolize AAA sacs to treat endoleaks include peripheral vascular embolization coils, cyanoacrylate or fibrin glue, gelfoam [4,5,6], and Onyx [7,8,9,10,11,12]. Active AAA sac management options, i.e., at the same time as EVAR, include the use of the Nellix system in the endovascular aneurysm sealing procedure [13,14,15,16], intraprocedural branch vessel embolization [17,18,19,20,21,22], and intraprocedural sac embolization with coils, thrombin, fibrin glue, and/or gelfoam [23,24,25,26,27,28,29,30,31,32,33].

Ultra-low-density polyurethane shape memory polymer is a novel embolic material now available as vascular plugs [34,35,36]. Shape memory polymer devices (IMPEDE-FX Embolization Plug, Shape Memory Medical, Santa Clara, CA, USA) are supplied in a stable crimped form for catheter delivery, and the devices self-expand to a specific volume upon exposure to blood, as they “remember” the manufactured shape. Porcine model studies characterized an acute inflammatory response to the polymer and the formation of a thrombus throughout the expanded porous structure. Progressive healing over 90 days resulted in noninflammatory replacement of shape memory polymer with cellular, collagenous connective tissue and stable vessel occlusion without microscopic indication of a sustained chronic-active inflammatory response [37]. Separately, a porcine sidewall aneurysm model experiment showed distinctly greater regression in aneurysm sacs treated with shape memory polymer than when treated with embolic coils 180-day post-implantation [38].

The purpose of this study was to determine the feasibility of using shape memory polymer vascular plugs for the embolization of aortic aneurysm sacs, to report on the safety of implanting the devices, and to report on preliminary efficacy in terms of changes in sac size and the rate of complications commonly associated with EVAR.

Methods

Study design

Retrospective analysis of consecutive patients treated with aortic aneurysm sac embolization at the same time as EVAR at 2 vascular centers in Germany, at an academic tertiary-level university and a teaching hospital. Patients were treated between January 2019 and July 2021. All patients gave written informed consent for the intervention, and local ethics committee approval was obtained for retrospective analysis. Anonymized data were independently reviewed by 2 experienced radiologists (A.M., F.F.).

Patients were candidates for elective EVAR treatment with perfused aneurysmal lumens large enough after EVAR to manipulate the delivery sheath and accommodate the delivery of embolization devices (per the device description and delivery procedure outlined below) and the absence of relevant calcifications of iliac access vessels to enable a parallel wire approach to the aneurysmal lumen using large transfemoral access sheaths. Furthermore, the large aneurysm sacs had a non-thrombosed aneurysm lumen/a low thrombus burden. Based on the experience of the interventionalists, these were considered to potentially benefit from sac embolization.

The primary endpoint was technical success, defined as the deployment of shape memory polymer devices in an aortic aneurysm sac outside of an aortic endograft in the same procedure. Secondary endpoints were changes in aneurysm sac volume, the rate of endoleaks, and procedure-related morbidity and mortality. Procedure time and fluoroscopy time were recorded, and inflammation markers (body temperature, blood sedimentation rate, leucocyte count) were also evaluated at 7 days before the procedure and in line with the follow-up imaging schedule. An additional timer was used to record the additional time used for sac embolization in each case.

Imaging data

Preprocedural contrast-enhanced computed tomography angiography (CTA) image analysis and volumetry were performed in all patients on a dedicated workstation (Siemens Healthineers Syngo.Via, Erlangen, Germany). Follow-up imaging included combined contrast-enhanced ultrasound (CEUS) and multiphase computed tomography (CT) (non-contrast CT, arterial and venous phase, or perfusion CTA) at 7 days and 3, 6, and 12 months after the procedure, based on institutional standard of care. Critical nephropathy was not present in the treated patients. Annual follow-up through 5 years postprocedure is ongoing.

All reported aneurysm measurements were based on CTA imaging analysis. Three-dimensional aneurysm volume was measured between positions of non-aneurysmal aorta diameter proximal and distal to the aneurysm. Perfused aneurysm volume was the aneurysm volume exclusive of preexisting thrombus. The residual flow lumen volume was calculated by subtraction of the endograft volume (estimated from endograft dimensions in each product’s instructions for use) from the perfused aneurysm volume (Fig. 1).

Fig. 1
figure 1

Schematic of the volumes associated with preparation for sac embolization. a Three-dimensional aneurysm volume was measured between positions of non-aneurysmal aorta diameter proximal and distal to the aneurysm (dotted lines, with applicable adjustments for individual anatomy). Perfused aneurysm volume was the aneurysm volume exclusive of preexisting thrombus. b The residual flow lumen volume was calculated by subtraction of the endograft volume (estimated from endograft dimensions in each product’s instructions for use) from the perfused aneurysm volume

Devices

Aortic endografts were selected based on case-specific needs. The shape memory polymer devices were 12-mm diameter IMPEDE-FX Embolization Plugs, as shown in Fig. 2. The shape memory polymer is crimped for catheter delivery and self-expands to a porous structure upon contact with blood. Each 12-mm diameter IMPEDE-FX device has a volume of ~ 1.25 mL when fully expanded. The shape memory polymer is radiolucent, and the devices are visualized under fluoroscopy via a small proximal radiopaque marker (Fig. 2). These devices were used both for sac embolization and the occlusion of aortic side branch ostia. The majority of cases were performed using individual IMPEDE-FX Embolization Plugs; three of the cases were performed with the IMPEDE-FX RapidFill device after it became available, in which 5 IMPEDE-FX Embolization Plugs are loaded in an introducer and implanted at the same time.

Fig. 2
figure 2

a Illustration of the implanted shape memory polymer device in its crimped (above) and (below) expanded form. b The appearance of shape memory polymer devices under follow-up ultrasound (yellow asterisks indicate location of some polymer). A video of shape memory polymer device deployment illustrating the appearance under fluoroscopy is shown in the supplementary material

Sac embolization procedure

Procedures were performed by experienced interventional radiologists only. An additional timer was used to record the additional time used for sac embolization in each case. A transfemoral parallel wire technique was used to maintain access to the aneurysm sac after deployment of aortic endografts. For this purpose, an upsized introducer sheath was used (GORE DrySeal Flex, Flagstaff, Arizona, USA; or Medtronic Sentrant, Minneapolis, MN, USA), with a 2-Fr larger inner diameter than was indicated for the EVAR procedure. This allowed for simultaneous insertion of an 0.89-mm/0.035″ aneurysm access wire (Terumo Glidewire Advantage, Tokyo, Japan) and endograft delivery catheter via a single transfemoral percutaneous access sheath (Fig. 3). After withdrawal of the endograft delivery catheter, the parallel wire (in the aneurysm sac only) was used to advance a flexible, angulated, 6-Fr sheath with a smooth transition (Terumo Destination Guiding Sheath, Tokyo, Japan) adjacent to the deployed endografts into the aneurysm sac. The instructions for use recommend a minimum 0.070″ inner diameter delivery catheter for the 12-mm IMPEDE-FX Embolization Plugs, and the chosen 6-Fr sheath met the criterion.

Fig. 3
figure 3

Large transfemoral access sheath for aortic endograft and a parallel wire with 6-Fr sheath for shape memory polymer device deployment into the aneurysm sac behind the endograft. Inset: illustration of the crimped form of the shape memory polymer device, which is supplied in its introducer

Devices were delivered with an intention to fill the sac with embolic material as much as possible. Devices were distributed throughout the sac by initial arc-shaped insertion of the flexible sheath across the aneurysm from one to the other side and stepwise retraction during the deployment process. In case of device clustering, a standard 0.89 mm/0.035″ wire was used to distribute the devices throughout the non-thrombosed aneurysm sac. The supplementary material shows a video of representative device deployment. In cases with additional targeted occlusion of aortic side branch ostia (inferior mesenteric, internal iliac, or lumbar arteries), a small number of devices were delivered right at the ostium of the side branch to encourage thrombosis, where the ostia locations were identified in preprocedural CTA and catheter angiography prior to aortic endograft deployment. This is distinctly different from branch vessel embolization remote from the sac via vessel cannulation. The decision criterion for ostia embolization was the presence of a prominent aortic side branch diameter > 3 mm, which is known to be prone to type 2 endoleak. In the single case including placement of the devices at the ostium of the internal iliac artery, devices were placed behind the endograft limb as the delivery catheter was withdrawn. Following deployment of the shape memory polymer devices, the endograft placement was completed with ballooning of the limbs, completion angiography, and vessel closure.

Periprocedural standard medication included 5000 international units of heparin and a prophylactic single-shot antibiosis. Therapeutic heparinization (body-weight-adapted subcutaneous injection of low-molecular-weight heparin) was continued for 48 h, and acetylsalicylic acid 100 mg/day is prescribed for life.

Data analysis

Percent change in follow-up aneurysm volumes was based on preprocedure baseline data. The mean percent change was calculated as the mean of individual percent changes in volume from baseline to follow-up. Continuous data are reported as the mean ± standard deviation and range (minimum − maximum). Categorical data are reported as number (percent of total). Continuous sac volume data were compared using Student’s t-test after confirming data were normally distributed (Kolmogorov–Smirnov test). A value of p < 0.05 was considered statistically significant.

Results

Eighteen patients (2 females, age 72 ± 9 years; range 61–88 years) were treated with aortic aneurysm sac embolization at the same time as EVAR. Patient demographic and baseline characteristics are shown in Table 1. The majority of aneurysms were infrarenal AAAs (n = 16), plus an aortoiliac aneurysm extending to the external iliac artery, and a thoracic aneurysm of the proximal descending thoracic aorta. Fourteen patients underwent sac embolization alone, and the ostia of aortic side branches were embolized in addition to sac embolization in an additional 4 patients. The ostia of a total of 8 side branches were targeted for embolization in the 4 patients; 4 inferior mesenteric arteries, 3 lumbar arteries in 2 patients, and 1 internal iliac artery. The aneurysms were large, and the residual flow volume was considered large enough and with morphology able to accommodate safe delivery of the embolization devices and potentially benefit from sac embolization. A variety of aortic endografts were used, and graft selection was based on interventionalist preference for individual cases (Table 1).

Table 1 Baseline and procedural data

Technical success was achieved in all cases, with a mean of 24 ± 12 devices implanted per patient (Table 1). This included the devices positioned at the ostia of side branches to encourage thrombosis, where a mean of 3 ± 2 (range 1–6) devices were used in this manner. The mean ratio of implanted shape memory polymer volume to the calculated residual flow volume was 108% ± 119% (range 10–433%), based on the fully expanded device volume of 1.25 mL per device. The overall sac filling rate of the cohort was 50.3% (a total calculated residual flow volume of 1,063 mL and a total of 428 devices, equivalent to 535 mL of expanded porous material).

Although the extent of sac filling was not a criterion for technical success, it was intended to fill the sac with embolic material as much as possible. However, in some cases, it was not possible to implant the intended number of devices (to fill the sac as much as possible) due to difficulties in positioning the delivery catheter at different locations in the aneurysm sac. In some patients, the number of devices implanted was larger than necessary to fill the sac based on the fully expanded device volume, which was likely due to the fast filling of the aneurysm sac before the shape memory polymer could fully expand. A representative example of fluoroscopic imaging immediately after sac embolization illustrating the distribution of the shape memory polymer devices (based on their radiopaque proximal markers) throughout an infrarenal AAA sac is shown in Fig. 4.

Fig. 4
figure 4

Completion fluoroscopic image post aortic endograft (Lombard Altura) deployment and implantation of 31 shape memory polymer devices, with a total expanded embolic material volume of 38.75 mL, based on 1.25 mL per device. The small proximal radiopaque markers indicate the positioning of the devices throughout the aneurysm sac behind the endograft (black arrow indicates a single device/marker)

Aneurysm sac embolization increased procedure and fluoroscopy times by a mean of 21 ± 11 (range 10–45) min and 6 ± 4 (range, 2–15) min, respectively. The maximum additional procedure time of 45 min was while treating one of the first three patients, due to learning curve, and difficulty in advancing the aneurysm access sheath along the already deployed bifurcated aortic endograft due to atherosclerotic-related high friction.

There was no additional administration of contrast media during aneurysm sac embolization as no further angiographic imaging was necessary.

All aneurysm sacs in the 16 patients with at least 3-month imaging data decreased in volume over a mean of 11 ± 7 months (Table 2). The mean change in aneurysm volume of -30 ± 21 mL was a significant change in volume from baseline (p < 0.001). In the patient group with 3- to 6-month follow-up (n = 8), the mean change in aneurysm volume of -27 ± 20 mL was a significant change in volume from baseline (p = 0.018). In the patient group with 12- to 24-month follow-up (n = 8), the mean change in aneurysm volume of -31 ± 20 mL was a significant change in volume from baseline (p = 0.019). The patient group with 12- to 24-month follow-up exhibited sac regression at the 6-month mark and continued to regress through longer-term follow-up. These preliminary data indicate that aneurysm sac shrinkage occurs in all cases during the first 6 months following aneurysm sac embolization with this novel shape memory polymer embolic device. The baseline and follow-up CTAs of a patient with a reduction in aneurysm volume of 48 mL at 12 months are shown in Fig. 5.

Table 2 Follow-up aneurysm sac volume data based on computed tomography angiography
Fig. 5
figure 5

a Preprocedure computed tomography angiography (CTA) of a 54-mm diameter, 104-mL infrarenal AAA sac. b Follow-up CTA at 12 months, illustrating a reduction in aneurysm volume of 48 mL. The sac was treated with 31 shape memory polymer devices, totaling 38.75 mL of embolic material, which represented 117% of the calculated residual flow lumen volume outside of the endograft (Lombard Altura) based on preprocedural CTA analysis. Yellow arrows indicate the markers of the shape memory polymer devices

Two-type IA endoleaks were identified by CEUS and CTA in 2 patients at 7-day follow-up but were completely resolved at 3 months without intervention. Eight type 3 endoleaks were observed in 8 patients at 7-day follow-up CEUS/CTA. In the cohort with 3–6 months follow-up, 2 endoleaks had completely resolved. The other 2 endoleaks persisted with reduced intensity and three-dimensional size. In the cohort with ≥ 12-month follow-up, 1 endoleak had completely resolved. The other 3 endoleaks were still present with reduced intensity and three-dimensional size. Three of the 4 patients with focused ostial embolization had no endoleak, and 1 patient exhibited a lumbar artery endoleak that had not been targeted for ostium embolization. Importantly, all patients that had endoleaks also had sac volume reduction, including the patients with persistent type 2 endoleaks.

There was no dislodgement of the devices from the sac during or after the procedure. There was also no off-target embolization from migration of the devices. No access site complications were observed. There was no morbidity or mortality related to aneurysm sac embolization throughout the follow-up to date. There was also no aneurysm-related morbidity or mortality in the same time frame. There were no clinical or laboratory test-based signs of inflammation; body temperature, white blood cell count, and blood sedimentation rate were within normal limits.

Discussion

This retrospective study evaluated the performance of a newly available polymer embolization device for aortic aneurysm sac embolization at the same time as EVAR. There are numerous reports of aortic aneurysm sac embolization at the same time as EVAR with widely available embolic agents, including randomized controlled trials of the use of coils and fibrin glue versus no sac embolization [27, 32]. However, the use of radiopaque devices for aneurysm sac embolization hinders the ability to identify endoleaks during follow-up. Furthermore, the use of liquid or flowing embolics may result in off-target embolization and resulting complications (e.g., spinal cord ischemia).

The properties of the shape memory polymer-based devices evaluated in this study are attractive for aortic sac embolization. The expanded, porous shape memory polymer material has been shown to support aneurysm sac regression [38]; the material has also been shown to support the growth of cellular, collagenous connective tissue throughout its structure without indication of a chronic inflammatory response [37]; and the majority of the device is radiolucent. Furthermore, the shape memory polymer devices self-expand to a prescribed, and therefore, predictable volume.

The instructions for use of the device also indicate the polymer bioabsorbs over time. Although each shape memory polymer device has an expanded volume of ~ 1.25 mL, it is important to recognize that the expanded porous structure of the device does not displace that volume of blood. In addition, the 1.25-mL volume comprises very little volume of polymer material, rather that the expanded, porous polymer structure serves as a scaffold supporting initial thrombus formation throughout its structure, and then followed by the formation of collagenous connective tissue intra-device.

In our experience, the deployment of shape memory polymer devices into aortic aneurysm sacs with a transfemoral parallel wire technique was relatively straightforward after a short learning curve. The devices are pushable, and the time required for device delivery did not extend the procedure or fluoroscopy time more than is reasonable, and the time is likely to decrease with increasing experience. Three of the cases were performed with the IMPEDE-FX RapidFill device after it became available, in which 5 IMPEDE-FX Embolization Plugs are loaded in an introducer and implanted at the same time. We decided to place devices at the ostia of branch vessels to encourage thrombosis if a vessel diameter was > 3 mm and patent, which is a widely acknowledged risk factor for type 2 endoleak development. The mean ratio of implanted shape memory polymer volume to the calculated residual flow volume was 108% per patient.

We acknowledge that the aortic endograft volume, and therefore the residual flow volume, may have changed on aortic endograft deployment, and therefore, this ratio is an approximation. Our intention was to fill the sac as much as possible. The shape memory polymer devices used in this study take > 5 min to expand to their predefined volume, and the expanded device has a low radial force. Consequently, we did not experience any adverse effects with filling the sacs more than 100%, based on the fully expanded volume of the device and the calculated residual flow volume. Furthermore, we did not observe any stent graft distortion.

The results of this study in terms of reduction in aneurysm sac size were encouraging, based on comparison with those reported by Piazza et al. [27] and Fabre et al. [32]. We reported our sac size data based on volume rather than diameter, since we believe volume is a more comprehensive means of measuring aneurysm sac expansion or regression. It is noteworthy that patients in our study were exhibiting a large amount of sac regression as early as 3–6 months postprocedure. In randomized controlled trials comparing with and without sac embolization, Piazza et al. and Fabre et al. reported a difference in volume-based sac regression between the two groups at 6 months and 12 months, respectively [27, 32]. Although preliminary data is in a retrospective study, our patients appear to be exhibiting larger changes in volume in the majority of patients and are continuing to regress over time. We believe this is a consequence of filling the sacs with embolic material, which is in turn enabled by the unique properties of the shape memory polymer.

The type 1 endoleaks observed in our patient set were clearly originating from the proximal sealing zone, were very small, did not warrant intervention, and resolved within 3 months. The type 2 endoleaks observed in our patients were also not significant enough to indicate intervention to date. Although all patients with endoleaks also exhibited a reduction in sac size, standard-of-care vigilance is ongoing. Piazza et al. demonstrated that outcomes may be dependent on the volume of embolic material in the sac [26, 27]. The small size of our study did not enable us to draw any conclusions on this topic; however, an increase in the amount of shape memory polymer placed in the sac may result in a decrease in the rate of endoleaks. It is noteworthy that the endoleaks were clearly visible in the presence of the shape memory polymer devices, i.e., there was no artifact interference. The shape memory polymer is radiolucent but is echogenic, and Fig. 2 shows the appearance of the shape memory polymer under ultrasound.

No safety concerns arose during the cases or follow-up to date. The potential for post implantation syndrome was a concern; however, no clinical or laboratory signs of inflammation occurred. There was no migration of the devices, and the treatment of an aortoiliac aneurysm without issues suggests the shape memory polymer devices may be used for such cases if there is sufficient sealing of the iliac aneurysm with the endograft limbs to prevent caudal migration. In a total of 201 patients, Piazza et al. and Fabre et al. reported very low compilation rates and no differences in complication rates between arms in randomized controlled trials comparing the outcomes of EVAR with and without sac embolization [27, 32].

The shape memory polymer device is priced in line with other similar peripheral vessel embolization devices in Germany when used for vessel embolization. Cost analyses of sac embolization should include consideration of all aspects of postoperative care, including reinterventions, and this retrospective feasibility study is not positioned to perform such analyses. However, when comparing with and without sac embolization with coils in a randomized controlled trial, Piazza et al. noted the additional cost of sac embolization more than the offset cost of higher number of reinterventions in the arm without sac embolization [27]. Fabre et al. also noted the cost of sac embolization should balance against the cost of follow-up and secondary interventions associated with aneurysm enlargement [32].

Our initial experience with shape memory polymer embolization devices serves as a foundation for further investigation into the use of these shape memory polymer devices for active aortic aneurysm sac management as sac regression was observed in all patients after three months, even in the presence of endoleaks. Optimization of patient selection will be essential to determine which patients may benefit from active sac embolization to promote sac regression and prevent type 2 endoleaks or minimize the consequences of type 2 endoleaks. A clear requirement is sufficient non-thrombosed aneurysmal lumen for the placement of the embolic devices. Independent risk factors for type 2 endoleak are the presence of a patent inferior mesenteric artery diameter ≥ 3 mm, patent lumbar, sacral or renal accessory arteries ≥ 2 mm, and aortoiliac aneurysm extension [39]. Based on these criteria, ~ 50% of patients referred for EVAR would be potential candidates for active aortic aneurysm sac embolization [21]. Moreover, indication for active sac management with embolization should be considered in terms of patient age, comorbidities, and aneurysm volume [39].

This retrospective feasibility study has obvious limitations, including lack of a control arm/comparison with standard EVAR. The retrospective nature of the study means inclusion/exclusion criteria were not predefined. The imaging data were analyzed retrospectively without core lab adjudication, and the study is small with limited follow-up of some patients to date. This study represents initial experience and with a variety of aortic endografts and different amounts of embolic material.

In conclusion, the use of shape memory polymer devices for the active treatment of aortic aneurysm sacs appears feasible based on this small case series. Technical success of the delivery of shape memory polymer devices into the residual flow lumen behind aortic endografts was achieved in all cases. A wide range aortic endografts were used. No safety concerns arose to date. The volume-based sac regression observed over the medium term supports further investigation into the potential of using this embolic material in active aortic aneurysm sac management. Prospective studies are needed to evaluate the effectiveness of active aneurysm sac management with these novel devices in specific patient populations.

Availability of data and materials

All datasets on which the conclusions of the manuscript rely are presented in the main paper.

Abbreviations

AAA:

Abdominal aortic aneurysm

CEUS:

Contrast-enhanced ultrasound

CTA:

Computed tomography angiography

EVAR:

Endovascular aneurysm repair

References

  1. Deery SE, Ergul EA, Schermerhorn ML et al (2018) Aneurysm sac expansion is independently associated with late mortality in patients treated with endovascular aneurysm repair. J Vasc Surg 67:157–164. https://doi.org/10.1016/j.jvs.2017.06.075

    Article  PubMed  Google Scholar 

  2. O’Donnell TFX, Deery SE, Boitano LT et al (2019) Aneurysm sac failure to regress after endovascular aneurysm repair is associated with lower long-term survival. J Vasc Surg 69:414–422. https://doi.org/10.1016/j.jvs.2018.04.050

    Article  PubMed  Google Scholar 

  3. White GH, Yu W, May J (1996) Endoleak–a proposed new terminology to describe incomplete aneurysm exclusion by an endoluminal graft. J Endovasc Surg 3:124–125. https://doi.org/10.1177/152660289600300119

    Article  CAS  PubMed  Google Scholar 

  4. Stavropoulos SW, Park J, Fairman R, Carpenter J (2009) Type 2 endoleak embolization comparison: Translumbar embolization versus modified transarterial embolization. J Vasc Interv Radiol 20:1299–1302. https://doi.org/10.1016/j.jvir.2009.07.003

    Article  PubMed  Google Scholar 

  5. Sarac TP, Gibbons C, Vargas L et al (2012) Long-term follow-up of type II endoleak embolization reveals the need for close surveillance. J Vasc Surg 55:33–40. https://doi.org/10.1016/j.jvs.2011.07.092

    Article  PubMed  Google Scholar 

  6. Yang RY, Tan KT, Beecroft JR, Rajan DK, Jaskolka JD (2017) Direct sac puncture versus transarterial embolization of type II endoleaks: an evaluation and comparison of outcomes. Vascular 25:227–233. https://doi.org/10.1177/1708538116663992

    Article  PubMed  Google Scholar 

  7. Ameli-Renani S, Pavlidis V, Morgan RA (2017) Early and midterm outcomes after transcatheter embolization of type I endoleaks in 25 patients. J Vasc Surg 65:346–355. https://doi.org/10.1016/j.jvs.2016.06.101

    Article  PubMed  Google Scholar 

  8. Abularrage CJ, Patel VI, Conrad MF, Schneider EB, Cambria RP, Kwolek CJ (2012) Improved results using Onyx glue for the treatment of persistent type 2 endoleak after endovascular aneurysm repair. J Vasc Surg 56:630–636. https://doi.org/10.1016/j.jvs.2012.02.038

    Article  PubMed  Google Scholar 

  9. Massis K, Carson WG 3rd, Rozas A, Patel V, Zwiebel B (2012) Treatment of type II endoleaks with ethylene-vinyl-alcohol copolymer (Onyx). Vasc Endovascular Surg 46:251–257. https://doi.org/10.1177/1538574412442401

    Article  PubMed  Google Scholar 

  10. Jamieson RW, Bachoo P, Tambyraja AL (2016) Evidence for ethylene-vinyl-alcohol-copolymer liquid embolic agent as a monotherapy in treatment of endoleaks. Eur J Vasc Endovasc Surg 51:810–814. https://doi.org/10.1016/j.ejvs.2016.02.015

    Article  CAS  PubMed  Google Scholar 

  11. Wojtaszek M, Wnuk E, Maciag R et al (2016) Improving the results of transarterial embolization of type 2 endoleaks with the embolic polymer Onyx. Wideochir Inne Tech Maloinwazyjne 11:259–267. https://doi.org/10.5114/wiitm.2016.64747

    Article  PubMed  PubMed Central  Google Scholar 

  12. Marcelin C, Le Bras Y, Petitpierre F et al (2017) Embolization for persistent type IA endoleaks after chimney endovascular aneurysm repair with Onyx((R)). Diagn Interv Imaging 98:849–855. https://doi.org/10.1016/j.diii.2017.04.005

    Article  CAS  PubMed  Google Scholar 

  13. Carpenter JP, Cuff R, Buckley C et al (2017) One-year pivotal trial outcomes of the Nellix system for endovascular aneurysm sealing. J Vasc Surg 65:330-336 e334. https://doi.org/10.1016/j.jvs.2016.09.024

    Article  PubMed  Google Scholar 

  14. Carpenter JP, Cuff R, Buckley C et al (2016) Results of the Nellix system investigational device exemption pivotal trial for endovascular aneurysm sealing. J Vasc Surg 63:23-31 e21. https://doi.org/10.1016/j.jvs.2015.07.096

    Article  PubMed  Google Scholar 

  15. Carpenter JP, Lane JS 3rd, Trani J et al (2018) Refinement of anatomic indications for the Nellix system for endovascular aneurysm sealing based on 2-year outcomes from the EVAS forward IDE trial. J Vasc Surg 68:720-730 e721. https://doi.org/10.1016/j.jvs.2018.01.031

    Article  PubMed  Google Scholar 

  16. O’Donnell TFX, Carpenter JP, Lane JS 3rd et al (2020) Endovascular aneurysm sealing is associated with higher medium-term survival than traditional EVAR. Ann Vasc Surg 63:145–154. https://doi.org/10.1016/j.avsg.2019.08.094

    Article  PubMed  Google Scholar 

  17. Axelrod DJ, Lookstein RA, Guller J et al (2004) Inferior mesenteric artery embolization before endovascular aneurysm repair: technique and initial results. J Vasc Interv Radiol 15:1263–1267. https://doi.org/10.1097/01.RVI.0000141342.42484.90

    Article  PubMed  Google Scholar 

  18. Sheehan MK, Hagino RT, Canby E et al (2006) Type 2 endoleaks after abdominal aortic aneurysm stent grafting with systematic mesenteric and lumbar coil embolization. Ann Vasc Surg 20:458–463. https://doi.org/10.1007/s10016-006-9103-2

    Article  PubMed  Google Scholar 

  19. Fukuda T, Matsuda H, Tanaka H, Sanda Y, Morita Y, Seike Y (2018) Selective inferior mesenteric artery embolization during endovascular abdominal aortic aneurysm repair to prevent type II endoleak. Kobe J Med Sci 63:E130–E135

    PubMed  PubMed Central  Google Scholar 

  20. Chew DK, Dong S, Schroeder AC, Hsu HW, Franko J (2019) The role of the inferior mesenteric artery in predicting secondary intervention for type II endoleak following endovascular aneurysm repair. J Vasc Surg 70:1463–1468. https://doi.org/10.1016/j.jvs.2019.01.090

    Article  PubMed  Google Scholar 

  21. Samura M, Morikage N, Otsuka R et al (2020) Endovascular aneurysm repair with inferior mesenteric artery embolization for preventing type II endoleak: a prospective randomized controlled trial. Ann Surg 271:238–244. https://doi.org/10.1097/SLA.0000000000003299

    Article  PubMed  Google Scholar 

  22. Vaillant M, Barral PA, Mancini J et al (2019) Preoperative inferior mesenteric artery embolization is a cost-effective technique that may reduce the rate of aneurysm sac diameter enlargement and reintervention following EVAR. Ann Vasc Surg 60:85–94. https://doi.org/10.1016/j.avsg.2019.03.012

    Article  PubMed  Google Scholar 

  23. Muthu C, Maani J, Plank LD, Holden A, Hill A (2007) Strategies to reduce the rate of type II endoleaks: routine intraoperative embolization of the inferior mesenteric artery and thrombin injection into the aneurysm sac. J Endovasc Ther 14:661–668. https://doi.org/10.1177/152660280701400509

    Article  PubMed  Google Scholar 

  24. Pilon F, Tosato F, Danieli D, Campanile F, Zaramella M, Milite D (2010) Intrasac fibrin glue injection after platinum coils placement: the efficacy of a simple intraoperative procedure in preventing type II endoleak after endovascular aneurysm repair. Interact Cardiovasc Thorac Surg 11:78–82. https://doi.org/10.1510/icvts.2009.231167

    Article  PubMed  Google Scholar 

  25. Ronsivalle S, Faresin F, Franz F, Rettore C, Zanchetta M, Olivieri A (2010) Aneurysm sac “thrombization” and stabilization in EVAR: a technique to reduce the risk of type II endoleak. J Endovasc Ther 17:517–524. https://doi.org/10.1583/09-3004.1

    Article  PubMed  Google Scholar 

  26. Piazza M, Frigatti P, Scrivere P et al (2013) Role of aneurysm sac embolization during endovascular aneurysm repair in the prevention of type II endoleak-related complications. J Vasc Surg 57:934–941. https://doi.org/10.1016/j.jvs.2012.10.078

    Article  PubMed  Google Scholar 

  27. Piazza M, Squizzato F, Zavatta M et al (2016) Outcomes of endovascular aneurysm repair with contemporary volume-dependent sac embolization in patients at risk for type II endoleak. J Vasc Surg 63:32–38. https://doi.org/10.1016/j.jvs.2015.08.049

    Article  PubMed  Google Scholar 

  28. Fabre D, Fadel E, Brenot P et al (2015) Type II endoleak prevention with coil embolization during endovascular aneurysm repair in high-risk patients. J Vasc Surg 62:1–7. https://doi.org/10.1016/j.jvs.2015.02.030

    Article  PubMed  Google Scholar 

  29. Mascoli C, Freyrie A, Gargiulo M et al (2016) Selective intra-procedural AAA sac embolization during EVAR reduces the rate of type II endoleak. Eur J Vasc Endovasc Surg 51:632–639. https://doi.org/10.1016/j.ejvs.2015.12.009

    Article  CAS  PubMed  Google Scholar 

  30. Natrella M, Rapellino A, Navarretta F et al (2017) Embo-EVAR: a technique to prevent type II endoleak? A single-center experience. Ann Vasc Surg 44:119–127. https://doi.org/10.1016/j.avsg.2017.01.028

    Article  PubMed  Google Scholar 

  31. Dosluoglu HH, Rivero M, Khan SZ, Cherr GS, Harris LM, Dryjski ML (2019) Pre-emptive nonselective perigraft aortic sac embolization with coils to prevent type II endoleak after endovascular aneurysm repair. J Vasc Surg 69:1736–1746. https://doi.org/10.1016/j.jvs.2018.10.054

    Article  PubMed  Google Scholar 

  32. Fabre D, Mougin J, Mitilian D et al (2021) Prospective, randomised two centre trial of endovascular repair of abdominal aortic aneurysm with or without sac embolisation. Eur J Vasc Endovasc Surg 61:201–209. https://doi.org/10.1016/j.ejvs.2020.11.028

    Article  PubMed  Google Scholar 

  33. Massmann A, Fries P, Kubale R, Seidel R (2018) Multiple treatment approaches following incomplete endovascular closure of abdominal aneurysm rupture-related aortocaval fistula. EuroIntervention 14:836–837. https://doi.org/10.4244/EIJ-D-18-00165

    Article  PubMed  Google Scholar 

  34. Morgan RA, Loftus I, Ratnam L et al (2021) Clinical experience with a shape memory polymer peripheral vascular embolisation plug: a case series. CVIR Endovasc 4:29. https://doi.org/10.1186/s42155-021-00214-w

    Article  PubMed  PubMed Central  Google Scholar 

  35. Castellano D, Boghi A, Di Maggio L, Rapellino A, Savio D (2021) Shape memory polymer ovarian vein embolisation in a patient with nickel allergy. CVIR Endovasc 4:25. https://doi.org/10.1186/s42155-021-00212-y

    Article  PubMed  PubMed Central  Google Scholar 

  36. Jansen AS, van Schaik PM, Martens JM, Reijnen M (2020) Embolization of the false lumen using IMPEDE-FX embolization plugs as part of treatment of an infrarenal post-dissection aneurysm: a case report. CVIR Endovasc 3:91. https://doi.org/10.1186/s42155-020-00183-6

    Article  PubMed  PubMed Central  Google Scholar 

  37. Jessen SL, Friedemann MC, Ginn-Hedman A et al (2020) Microscopic assessment of healing and effectiveness of a foam-based peripheral occlusion device. ACS Biomat Sci Eng 5:2588–2599. https://doi.org/10.1021/acsbiomaterials.9b00895

    Article  CAS  Google Scholar 

  38. Horn J, Hwang W, Jessen SL et al (2017) Comparison of shape memory polymer foam versus bare metal coil treatments in an in vivo porcine sidewall aneurysm model. J Biomed Mater Res B Appl Biomater 105:1892–1905. https://doi.org/10.1002/jbm.b.33725

    Article  CAS  PubMed  Google Scholar 

  39. Samura M, Morikage N, Mizoguchi T et al (2018) Identification of anatomical risk factors for type II endoleak to guide selective inferior mesenteric artery embolization. Ann Vasc Surg 48:166–173. https://doi.org/10.1016/j.avsg.2017.10.016

    Article  PubMed  Google Scholar 

Download references

Funding

No funding was received for this work.

Author information

Authors and Affiliations

Authors

Contributions

Conception and design, AM. Data collection, AM, PF, FF, RSK, and PB. Analysis and interpretation, AM, PF, RSK, AB, PB, and FF. Writing the manuscript, AM, PF, and AB. Critical revision, AM, PF, RSK, AB, PB, and FF. Approval of the manuscript, AM, PF, RSK, AB, PB, and FF. The authors read and approved the final manuscript.

Corresponding author

Correspondence to Alexander Massmann.

Ethics declarations

Ethics approval and consent to participate

The local ethics committee (Ärztekammer des Saarlandes) approval was obtained for retrospective analysis.

Consent for publication

All patients gave written informed consent for the intervention and scientific publication. Our manuscript does not contain any individual person’s data.

Competing interests

Prof. A. Massmann is a consultant to Shape Memory Medical. All other authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Video 1. Fluoroscopy IMPEDE deployment.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Massmann, A., Fries, P., Shayesteh-Kheslat, R. et al. Feasibility of aortic aneurysm sac embolization using a novel shape memory polymer embolic device. Eur Radiol Exp 7, 12 (2023). https://doi.org/10.1186/s41747-023-00328-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-023-00328-x

Keywords