Table 1 Main physical characteristics of the three mammography systems compared

From: Phantom-based analysis of variations in automatic exposure control across three mammography systems: implications for radiation dose and image quality in mammography, DBT, and CEM

System

Senographe Pristina

Selenia Dimensions

Selenia 3Dimensions

Manufacturer

GEHC

Hologic

Hologic

Detector type

CsI FPD

a-Se FPD

a-Se FPD

Pixel pitch (μm)

100

70

(2 × 2 binning in DBT)

70

Modalities

DM

DBT

CEM

DM

DBT

DM

DBT

CEM

Anode(s)

Mo and Rh

W

W

Filter(s)

Mo, Ag, Cua

Rh, Ag, Alb, Cua

Rh, Ag, Alb, Cua

DBT protocol

 Tube motion

Step-and-shoot

Continuous

Continuous

 Projections

9

15

15

 Angle

25° (± 12.5°)

15° (± 7.5°)

15° (± 7.5°)

 Geometry

Partial isocentric

Full isocentric

Full isocentric

 Scatter

Grid

No Grid

No Grid

N° AEC modes

DM: 3 (AOP/STD, AOP/STD+ , AOP/DOSE)

DBT: 2 (AOP/STD, AOP/STD+)

CEM: 1 (AOP/STD)

1 (AutoFilter)

1 (AutoFilter)

  1. AOP/STD, AOP/STD+ , AOP/DOSE, AutoFilter are the names given by the two manufacturers to their AEC modes
  2. FPD Flat panel detector, CsI Cesium iodine, a-Se Amorphous selenium, DM Digital mammography, DBT Digital breast tomosynthesis, CEM Contrast-enhanced mammography, AEC Automatic exposure control
  3. aCu filter is used by both systems (GE Pristina and Hologic 3Dimensions) for the only high-energy image in CEM mode
  4. bAl filter is used by Hologic systems in DBT mode