Table 2 Parameters of diffusion tensor imaging and T2-weighted sequences

From: Diffusion tensor imaging in anisotropic tissues: application of reduced gradient vector schemes in peripheral nerves

 

Diffusion tensor imaging

T2-weighted sequences

Scanner

Prismafit

TimTrio

Prismafit

TimTrio

Repetition time (ms)

4,000

4,000

6,969

8,150

Echo time (ms)

87

92.8

54

54

Field of view

160 × 160 mm2

160 × 160 mm2

140 × 140 mm2

160 × 160 mm2

Matrix

128 × 128

128 × 128

512 × 358

512 × 333

Slice thickness (mm)

4.0

4.0

3.5

3.5

Number of slices

18

18

35

41

Slice gap (mm)

0.4

1.2

0.35

0.35

Number of averages

3

3

3

2

Parallel imaging

GRAPPA

GRAPPA

GRAPPA

GRAPPA

Acceleration factor

2

3

2

2

Acquisition time (min:s)

4:34

4:32

04:22

5:29

Fat saturation

SPAIR

SPAIR

Spectral

Spectral

b value

0, 1,000 s/mm2

0, 1,000 s/mm2

 − 

 − 

Diffusion gradients

20

20

 − 

 − 

  1. GRAPPA Generalized autocalibrating partial parallel acquisition, SPAIR Spectral adiabatic inversion recovery