Skip to main content

Table 2 Diagnostic accuracy, sensitivity, specificity, and inter-rater reliability (Cohen κ) for signs of implant degeneration

From: Photon-counting CT of degenerative changes and rupture of silicone breast implants: a pilot study

 

Accuracy (95% CI)

Sensitivity (95% CI)

Specificity (95% CI)

Inter-rater reliability

Folds

0.81 (0.69, 0.90)

0.75 (0.62, 0.88)

 > 0.99 (0.70, > 0.99)

0.93

Keyhole sign

0.90 (0.79, 0.96)

0.82 (0.70, 0.95)

 > 0.99 (0.86, > 0.99)

0.59

Fluid outside

0.78 (0.65, 0.87)

0.64 (0.39, 0.89)

0.82 (0.70, 0.93)

0.58

Fluid inside

 > 0.99 (0.94, > 0.99)

 > 0.99 (0.63, > 0.99)

 > 0.99 (0.93, > 0.99)

 > 0.99

Silicone outside

0.95 (0.86, 0.99)

0.50 (0.01, > 0.99)

0.98 (0.95, > 0.99)

0.65

Linguini sign

 > 0.99 (0.94, > 0.99)

 > 0.99 (0.16, > 0.99)

 > 0.99 (0.94, > 0.99)

 > 0.99

  1. CI Confidence interval