Skip to main content

Table 3 ADC of the proposed fat-containing diffusion phantom

From: Understanding ADC variation by fat content effect using a dual-function MRI phantom

GBD

g/50 mL

Fat content (%)

0

10

20

30

40

50

0

1.653 ± 0.002

1.419 ± 0.020

1.253 ± 0.014

1.124 ± 0.043

1.031 ± 0.006

0.819 ± 0.011

0.1

1.375 ± 0.007

1.205 ± 0.012

0.980 ± 0.002

0.797 ± 0.016

0.608 ± 0.024

0.547 ± 0.027

0.25

1.261 ± 0.005

1.114 ± 0.010

0.855 ± 0.034

0.691 ± 0.020

0.555 ± 0.003

0.458 ± 0.017

0.5

0.909 ± 0.021

0.648 ± 0.045

0.502 ± 0.038

0.393 ± 0.016

0.315 ± 0.014

0.271 ± 0.035

1.0

0.399 ± 0.006

0.312 ± 0.041

0.272 ± 0.023

0.222 ± 0.016

0.202 ± 0.015

0.145 ± 0.021

  1. ADC Apparent diffusion coefficient (×10−3 mm2/s), FC Fat content (%), GBD Glass bead density (g/50 mL). The ADC value is presented as mean ± standard deviation