Fig. 3

From: Synthetic interpolated DSA for radiation exposure reduction via gamma variate contrast flow modeling: a retrospective cohort study

Fig. 3

a Heatmaps demonstrating the percent error of the relative x-ray intensities at each voxel over the entire study when comparing the original angiographic data to the synthetically reconstructed study. Red represents ~ 15% relative x-ray intensity error and green representing < 3% error. From left to right, the error maps of a single study for each level of subtraction are presented. b Our system allows users to inspect every voxel of the error map on the left and demonstrates the gamma variate function fits for each level of frame rate reduction on the right. Drop 1, 2, and 3 represent 50%, 66%, and 75% frame reduction respectively. R is the percent error