Table 1 System parameters of the prototype silicon-based photon-counting computed tomography (Si-PCCT) and clinical energy-integrating detector CT (EIDCT) systems

From: Stent appearance in a novel silicon-based photon-counting CT prototype: ex vivo phantom study in head-to-head comparison with conventional energy-integrating CT

 

Si-PCCT

EIDCT

Detector material

Silicon semiconductor

Gemstone scintillator

CTDIvol (32 cm)

8.9 mGy

9 mGy

Scan mode

Helical

Helical

Pitch

0.99

0.98

Bowtie filter

Medium

Head (medium body)

Tube voltage

120 kVp

120 kVp

Tube current

255 mA

245 mA

Tube rotation

0.5 s

0.5 s

Focal spot size

0.6 mm

1.2 mm

z-coverage

40 mm

40 mm

Matrix

1024 × 1024

512 × 512

Field-of-view (FOV)

150 × 150 mm (150-mm FOV) 50 × 50 mm (50-mm FOV)

150 × 150 mm (150-mm FOV) 50 × 50 mm (50-mm FOV)

Pixel size

0.146 mm (150-mm FOV) 0.049 mm (50-mm FOV)

0.293 mm (150-mm FOV) 0.098 mm (50-mmFOV)

Slice thickness

0.42 mm

0.63 mm

Kernel

Bone + ASIR0

Bone + ASIR0

  1. ASIR0 Adaptive statistical iterative reconstruction with 0% blending, CTDIvol Volumetric computed tomography dose index, EIDCT Energy-integrating computed tomography, PCCT Photon-counting computed tomography, FOV Field-of-view