Skip to main content

Table 1 Pulse sequence parameters employed in this study

From: Comparison of gadolinium-based contrast agents for MR cholangiography in saline, blood and bile: a phantom study

Sequence

Repetition time (ms)

Echo time (ms)

Flip angle (degrees)

T1-weighted spin-echo

600–700

10

70

T1-weighted turbo field-echo

“shortest”

“shortest”

8

3D multiecho fast field-echo

“shortest”

4.6

20

Multiecho gradient- and spin-echo

“shortest”

“shortest”

90

Short tau inversion-recovery

5,000

30

90

T1-weighted high-resolution isotropic volume excitation

“shortest”

“shortest”

10

2D MIXED

920 (spin-echo) 2,300 (inversion-recovery)

“shortest”

90

T2-weighted turbo spin-echo

3,000

80

90

Fluid-attenuated inversion-recovery

11,000

125

90

  1. 2D Two-dimensional, 3D Three-dimensional