Skip to main content

Table 2 Objective image noise: pure versus hybrid iterative reconstruction

From: Development and implementation of an ultralow-dose CT protocol for the assessment of cerebrospinal shunts in adult hydrocephalus

Anatomical cavity

Mean IN-pure IR

Mean IN-hybrid IR

Noise reduction

p-value

Head

7.81 ± 3.06

52.94 ± 14.46

85.3%

p < 0.0001

Neck

9.52 ± 9.98

57.13 ± 44.65

83.3%

p < 0.0001

Thorax

8.69 ± 4.95

84.24 ± 38.52

89.7%

p < 0.0001

Abdomen

12.5 ± 7.18

98.59 ± 43.29

87.3%

p < 0.0001

Pelvis

9.23 ± 5.55

87.15 ± 48.1

89.4%

p < 0.0001

  1. IN Image noise, IR Iterative reconstruction
  2. Data represent mean ± standard deviation. Difference in means was assessed using a paired sample t-test. A p-value < 0.05 was considered statistically significant