Table 1 Details of the contrast medium specifics

From: Impact of iodine concentration and scan parameters on image quality, contrast enhancement and radiation dose in thoracic CT

Phantom size

Scan parameter tube voltage:

Constant CTDI

Dose modulation

70 kVp

80 kVp

100 kVp

120 kVp

140 kVp

70 kVp

80 kVp

100 kV

120 kV

140 kV

Large

Rotation time (s)

1

1

0.5

0.5

0.5

0.5

0.5

0.5

0.5

0.5

mAs (range 80–500)

500

395

405

250

170

499

499

464

372

282

CTDIvol (mGy)

13.17

16.6

16.68

16.6

16.43

6.59

10.11

14.79

16.6

17.59

Medium

Rotation time (s)

1

0.5

0.5

0.5

0.5

0.5

0.5

0.5

0.5

0.5

mAs (range 80–500)

380

475

240

150

105

499

499

322

244

178

CTDIvol (mGy)

10

9.98

9.89

9.96

10.15

6.09

8.14

8.98

10

11.6

Small

Rotation time (s)

0.5

0.5

0.5

0.5

0.5

0.5

0.5

0.5

0.5

0.5

mAs (range 80–500)

500

330

170

105

70

484

630

219

106

123

CTDIvol (mGy)

6.59

6.93

7.00

6.97

6.77

4.67

5.18

5.73

7.05

8.75

  1. CTDIvol Volume computed tomography dose index. Scan technique: constant CTDIvol and dose modulation (SmartmA) using conventional, iterative reconstruction. Additional parameters: scan mode, helical; detector collimation, 80 × 0.625; pitch, 0.5; scan field of view, 50 cm; display field of view, 36 cm; reconstructed slice thickness, 2.5 mm; reconstruction kernel, standard; iterative reconstruction, adaptive statistical iterative reconstruction, ASIR-V 50%; noise index14.5 (scans with dose modulation)