From: Patient organ and effective dose estimation in CT: comparison of four software applications
 | Slice thickness (%) | Pitch (%) | Collimation (%) | Scanner model (%) |
---|---|---|---|---|
Organs inside the scan region | 0–1 | < 1 | < 1 | 3–13 |
Partially irradiated organs | 0–3 | 6–40 | 1–4 | 0–50 |
Organs outside the scan region | 1–3 | 7–30 | 1–10 | 5–40 |
Distributed organs | 0 | 2–6 | < 2 | 1–2 |