Skip to main content

Table 4 Histologic comparison of vessel injury and healing

From: Prospective study to assess the tissue response to HPC-coated p48 flow diverter stents compared to uncoated devices in the rabbit carotid artery model

Implant group

Injury score

Thrombus score

Intimal/medial fibrin score

Inflammation score

Giant cells score

RBCs score

Intimal coverage

Adventitial inflammation score

p48 MW HPC (n = 8)

0.46 ± 0.43

0.04 ± 0.12

0.00 ± 0.00

2.12 ± 0.75

2.04 ± 0.84

0.00 ± 0.00

Complete

0.13 ± 0.25

Uncoated P48 MW (n = 8)

0.54 ± 0.35

0.00 ± 0.00

0.00 ± 0.00

1.96 ± 0.79

1.96 ± 0.79

0.00 ± 0.00

Complete

0.08 ± 0.24

p value

0.737

0.317

1.000

0.707

0.789

1.000

NA

0.589

  1. Data are presented as mean ± standard deviation. RBC Red blood cell, NA Not available