Table 4 SNR and CNR calculations

From: Photon-counting CT of elbow joint fractures: image quality in a simulated post-trauma setting with off-center positioning

 

Kernel

Position

Dose protocol

PCD-CT (mean ± SEM)

EID-CT (mean ± SEM)

p value

SNR

Br76/Ur77

OH

2 mGy

0.6 ± 0.11

0.3 ± 0.05

< 0.001

6 mGy

0.7 ± 0.11

0.3 ± 0.04

< 0.001

12 mGy

0.7 ± 0.11

0.6 ± 0.15

0.001

ABD

2 mGy

0.5 ± 0.16

0.3 ± 0.05

< 0.001

6 mGy

0.7 ± 0.16

0.3 ± 0.05

< 0.001

12 mGy

0.7 ± 0.17

0.4 ± 0.07

< 0.001

Br89/Ur89

OH

2 mGy

0.3 ± 0.07

0.3 ± 0.01

0.031

6 mGy

0.4 ± 0.09

0.3 ± 0.02

< 0.001

12 mGy

0.4 ± 0.12

0.2 ± 0.03

< 0.001

ABD

2 mGy

0.3 ± 0.58

0.4 ± 0.03

< 0.001

6 mGy

0.3 ± 0.08

0.3 ± 0.01

0.235

12 mGy

0.3 ± 0.11

0.3 ± 0.02

0.774

CNR

Br76/Ur77

OH

2 mGy

0.9 ± 0.16

0.4 ± 0.04

< 0.001

6 mGy

1.0 ± 0.14

0.6 ± 0.05

< 0.001

12 mGy

1.0 ± 0.15

1.0 ± 0.12

0.013

ABD

2 mGy

0.5 ± 0.32

0.4 ± 0.10

< 0.001

6 mGy

0.9 ± 0.20

0.4 ± 0.04

< 0.001

12 mGy

0.9 ± 0.21

0.5 ± 0.06

< 0.001

Br89/Ur89

OH

2 mGy

0.4 ± 0.11

0.4 ± 0.03

0.031

6 mGy

0.6 ± 0.13

0.3 ± 0.02

< 0.001

12 mGy

0.7 ± 0.18

0.3 ± 0.04

< 0.001

ABD

2 mGy

0.4 ± 0.71

0.4 ± 0.04

< 0.001

6 mGy

0.4 ± 0.11

0.1 ± 0.03

< 0.001

12 mGy

0.5 ± 0.13

0.4 ± 0.02

0.011

  1. Calculations of SNR and CNR after a ROI-based measurement of signal attenuation for different dose protocols, arm scanning positions and reconstruction kernels. The results are given as mean values with the standard error of mean (± SEM). The p value of the Mann-Whitney U test analysis is reported. ABD Abdominal scanning position, CNR Contrast-to-noise ratio, EID-CT Energy-integrating detector computed tomography, OH Overhead scanning position, PCD-CT Photon-counting detector computed tomography, ROI Region of interest, SNR Signal-to-noise ratio