Table 2 Imaging protocols and pulse sequence parameters for 0.55-T, 1.5-T and 3-T scanner systems

From: Visual and quantitative assessment of hip implant-related metal artifacts at low field MRI: a phantom study comparing a 0.55-T system with 1.5-T and 3-T systems

Field strength

Plane

Sequence

TR (ms)

TE (ms)

TI (ms)

BW (Hz/pixel)

FOV (mm)

Matrix

ST (mm)

TA (min)

0.55 T

Coronal

T2 TSE TIRM

4,440

35

100

299

220 × 220

134 × 192

3.5

02:48

Coronal

T2 TSE

3,220

77

-

434

220 × 220

224 × 320

3.5

02:21

Axial

T2 TSE TIRM

4,260

25

100

302

220 × 220

192 × 240

5

04:58

Axial

T1 TSE

517

9.4

-

401

220 × 220

213 × 304

5

03:40

Sagittal

T1 TSE

517

9.4

-

401

220 × 220

213 × 304

5

03:40

1.5 T

Coronal

T2 TSE TIRM

4,000

34

150

455

220 × 220

192 × 256

3.5

01:48

Coronal

T2 TSE

3,400

73

-

430

220 × 220

358 × 448

3.5

01:25

Axial

T2 TSE TIRM

5,210

54

150

400

180 × 180

224 × 320

5

03:33

Axial

T1 TSE

661

9.5

-

305

200 × 200

218 × 256

5

05:44

Sagittal

T1 TSE

596

8.5

-

460

220 × 220

256 × 320

5

02:32

3 T

Coronal

T2 TSE TIRM

4,800

42

210

505

240 × 240

240 × 320

3.5

01:36

Coronal

T2 TSE

4,500

73

-

505

220 × 220

224 × 320

3.5

01:30

Axial

T2 TSE TIRM

4,200

39

200

500

180 × 180

205 × 256

5

02:06

Axial

T1 TSE

687

17

-

520

200 × 200

320 × 320

5

02:53

Sagittal

T1 TSE

600

9.3

-

520

240 × 240

384 × 384

5

02:32

  1. BW Bandwidth, FOV Field of view, ST Slice thickness, T1 T1-weighted, T2 T2-weighted, TA Acquisition time, TE Echo time, TI Inversion time, TIRM Turbo inversion-recovery magnitude, TR Repetition time, TSE Turbo spin-echo