Table 2 Noise magnitude obtained for the low energy levels on virtual monoenergetic images (VMIs) and three reconstruction levels of both versions of deep learning spectral reconstruction (DLSR)

From: Comparison of image quality of two versions of deep-learning image reconstruction algorithm on a rapid kV-switching CT: a phantom study

DLSR version

Energy level

Reconstruction level

Mild

Standard

Strong

DLSR V1

40 keV

100.2 ± 0.40

86.6 ± 0.35

65.6 ± 0.31

50 keV

52.4 ± 0.32

45.3 ± 0.22

34.3 ± 0.28

60 keV

24.3 ± 0.09

21.0 ± 0.13

15.5 ± 0.19

70 keV

12.5 ± 0.10

10.6 ± 0.08

7.7 ± 0.12

DLSR V2

40 keV

68.5 ± 0.20

55.1 ± 0.27

44.6 ± 0.41

50 keV

34.3 ± 0.10

28.1 ± 0.14

22.7 ± 0.21

60 keV

16.6 ± 0.04

13.6 ± 0.07

10.9 ± 0.10

70 keV

13.1 ± 0.04

10.6 ± 0.05

8.1 ± 0.08

  1. Data are given as mean ± standard deviation