Table 1 Scanning and reconstruction parameters for multislice computed tomography using energy-integrating detectors (EID-CT) and multislice computed tomography using photon-counting detectors (PCD-CT)

From: Photon-counting detector CT and energy-integrating detector CT for trabecular bone microstructure analysis of cubic specimens from human radius

Scanner

Tube voltage (kVp)

Quality reference tube load (mAs)

Effective tube load (mAs)

Pitch

CTDIvol (mGy)

Field of view (mm)

Matrix size

Intra-slice voxel size (μm)

Slice increment (μm)

Slice thickness (μm)

Voxel size (μm)

Kernel

PCD-CT 51

120

450

~170

0.6

29

51

1,024

50

50

200

50

Qr80u

PCD-CT 30

120

450

~170

0.6

29

30

2,048

15

50

200

50

Qr80u

PCD-CT 18

120

450

~170

0.6

29

18

2,048

8.8

50

200

50

Qr80u

EID-CT 1000

120

1,000

~360

0.6

49

51

512

100

100

400

100

Ur81-3

EID-CT 600

120

600

~210

0.6

29

51

512

100

100

400

100

Ur77-3

  1. The computed tomography dose index (CTDIvol) was supplied by the CT scanner software. The numbers after the type of scanner correspond to the field of view in mm (PCD-CT) and quality reference tube load in mAs (EID-CT)