Table 4 Incident air kerma rate at the interventional reference point and average image quality of all three readers of fluoroscopic images acquired with detector dose-driven exposure control (DEC) and Ta-specific contrast-to-noise ratio-driven exposure control (CEC), while maintaining the incident air kerma rate

From: Tantalum-specific contrast-to-noise ratio or conventional detector dose-driven exposure control in angiography: radiation dose and image quality evaluation in a porcine model

Mean WET (cm)

Dose level

(DEC)

Incident air kerma rate at the interventional reference point

Average image quality

DEC

CEC

p value

DEC

CEC

AUC

p value

26

LD

6.6 ± 1.4

6.2 ± 1.3

0.003

2.6 ± 0.7

3.4 ± 0.7

0.82 [0.68–0.93]

0.018

ND

15.8 ± 3.9

15.4 ± 3.8

0.216

2.9 ± 0.4

4.2 ± 0.5

0.97 [0.89–1.00]

0.014

HD

28.8 ± 7.6

28.5 ± 7.5

0.147

3.3 ± 0.6

4.8 ± 0.4

0.98 [0.91–1.00]

0.017

31

LD

24.5 ± 5.1

23.5 ± 6.3

0.165

2.4 ± 0.8

3.3 ± 0.6

0.84 [0.64–1.00]

0.020

ND

53.0 ± 4.4

50.8 ± 3.4

0.137

2.8 ± 0.5

4.4 ± 0.6

0.97 [0.89–1.00]

0.022

HD

86.1 ± 4.6

81.0 ± 7.7

0.044

3.2 ± 0.6

4.6 ± 0.6

0.93 [0.79–1.00]

0.014

36

LD

60.7 ± 4.5

58.1 ± 7.1

0.182

2.1 ± 0.7

2.9 ± 0.8

0.85 [0.67–1.00]

0.019

ND

86.4 ± 5.7

83.7 ± 6.7

0.163

2.4 ± 0.6

3.3 ± 0.7

0.85 [0.68–0.93]

0.011

HD

129.4 ± 8.4

124.2 ± 14.1

0.212

2.8 ± 0.5

3.9 ± 0.7

0.93 [0.79–1.00]

0.014

  1. Values are given as mean ± standard deviation. AUC Area under curve, CEC Contrast-to-noise ratio-driven exposure control, DEC Detector dose driven exposure control, LD Low dose, ND Normal dose, HD High dose, WET Water equivalent thickness