Skip to main content

Table 2 Experimentally estimated T1 and T2 values for synovial fluid, blood, LoB, LoD, and blood detection threshold at 1.5, 3, and 7 T

From: Detecting low blood concentrations in joints using T1 and T2 mapping at 1.5, 3, and 7 T: an in vitro study

 

Field strength

Relaxation time synovial fluid (ms)

Relaxation time blood (ms)

LoB (ms)

LoD (ms)

Blood detection threshold

T1

1.5 T

2641 ± 60

1258 ± 33

2541

2526

≥ 10%

3 T

2719 ± 72

1310 ± 33

2600

2560

≥ 25%

7 T

2355 ± 145

1272 ± 160

2117

2040

≥ 50%

T2

1.5 T

845 ± 56

93 ± 11

753

298

≥ 50%

3 T

592 ± 13

28 ± 20

579

546

≥ 5%

7 T

281 ± 5

Not measurable

272

218

= 25%

  1. Results are reported as mean values, and the standard deviations as data were normally distributed (Kolmogorov-Smirnov test, p > 0.05)
  2. Blood blood concentration 100%, Blood detection threshold blood concentration corresponding to the LoD, LoB limit of blank, LoD limit of detection, synovial fluid blood concentration 0%