Skip to main content

Table 1 3-T and 7-T protocols and radiologic assessment of coup/contrecoup injuries and microlesions

From: 7-T MRI for brain virtual autopsy: a proof of concept in comparison to 3-T MRI and CT

Sequence

3 T

7 T

3 T

7 T

3 T

7 T

T1-w. TFE

T1-w. MP2RAGE

T2-w. TSE SPAIR

T2-w. SPACE

VenBOLD

SWI

Repetition time (ms)

9.4

4,500

1,500

4,000

18

21

Echo time (ms)

4.6

2

230

118

26

14

Voxel size*

1.00 × 1.00 × 2.00

0.65 × 0.65 × 0.63

0.80 × 0.80 × 1.60

0.71 × 0.71 × 0.67

0.80 × 0.80 × 1.60

0.12 × 0.12 × 1.50

Slice orientation

Sagittal

Sagittal

Transversal

Sagittal

Transversal

Transversal

Scan time (min:s)

08:50

10:45

06:00

05:39

06:21

10:40

Coup injury

1−

2+

1+

3++

1+

3++

Contercoup injury

1−

2+

1+

3++

1+

3++

Microinfarct

1−

2−

1+

2++

1-

3-

Microhaemorrhage

1−

2−

1+

2++

1+

3+

  1. MP2RAGE Magnetisation prepared 2 rapid gradient echoes, SPACE Sampling perfection with application-optimised contrasts by using different flip angle evolutions, SPAIR Spectral attenuated inversion recovery, SWI Susceptibility-weighted imaging, TFE Turbo field echo, TSE Turbo spin echo, VenBOLD Venous blood oxygen level dependent
  2. *Frequency direction × phase direction × slice thickness