Table 3 Assessment of image noise and artefacts in the bone and soft tissue for cone-beam (CBCT) and multidetector CT (MDCT) by two radiologists (R1, R2) using a five-point Likert scale

From: 3D cone-beam CT with a twin robotic x-ray system in elbow imaging: comparison of image quality to high-resolution multidetector CT

 

Score

Regular-dose CBCT

Low-dose CBCT

Regular-dose MDCT

Low-dose MDCT

  

R1

R2

R1

R2

R1

R2

R1

R2

Noise in the bone

5

16 (100.0)

14 (87.5)

6 (37.5)

10 (62.5)

2 (12.5)

4

2 (12.5)

10 (62.5)

7 (43.8)

4 (25.0)

8 (50.0)

6 (37.5)

4 (25.0)

3

6 (37.5)

3 (18.8)

2 (12.5)

4 (25.0)

3 (18.8)

4 (25.0)

2

2 (12.5)

3 (18.8)

4 (25.0)

1

-

4 (25.0)

4 (25.0)

Median

5.0

5.0

4.0

4.0

5.0

4.0

3.0

2.5

Noise in the soft tissue

5

1 (6.3)

2 (12.5)

2 (12.5)

4

10 (62.5)

14 (87.5)

6 (37.5)

2 (12.5)

3

5 (31.3)

2 (12.5)

7 (43.8)

9 (56.3)

8 (50.0)

10 (62.5)

2 (12.5)

2 (12.5)

2

5 (31.3)

4 (25.0)

2 (12.5)

2 (12.5)

2 (12.5)

1

4 (25.0)

3 (18.8)

12 (75.0)

12 (75.0)

Median

4.0

4.0

2.0

3.0

3.5

3.0

1.0

1.0

Artefacts in the bone

5

16 (100.0)

16 (100.0)

16 (100.0)

15 (93.8)

13 (81.3)

12 (75.0)

13 (81.3)

12 (75.0)

4

1 (6.3)

2 (12.5)

4 (25.0)

2 (12.5)

2 (12.5)

3

1 (6.3)

2 (12.5)

2

1

Median

5.0

5.0

5.0

5.0

5.0

5.0

5.0

5.0

Artefacts in the soft tissue

5

13 (81.3)

12 (75.0)

13 (81.3)

12 (75.0)

4

15 (93.8)

12 (75.0)

15 (93.8)

7 (43.8)

2 (12.5)

3 (18.8)

2 (12.5)

2 (12.5)

3

1 (6.3)

2 (12.5)

1 (6.3)

5 (31.3)

1 (6.3)

1 (6.3)

1 (6.3)

2 (12.5)

2

2 (12.5)

4 (25.0)

1

Median

4.0

4.0

4.0

3.0

5.0

5.0

5.0

5.0

  1. Image quality ordinal score: 5 = minimal artefacts or noise; 4 = little artefacts or noise; 3 = moderate artefacts or noise; 2 = considerable artefacts or noise; 1 = strong artefacts or noise). Scale results are displayed as frequencies (percentages) and median values