From: Average absorbed breast dose (2ABD): an easy radiation dose index for digital breast tomosynthesis
Phantom thickness (cm) | Digital mammography (single view) (W/Rh and W/Ag*) | Digital breast tomosynthesis (W/Al) | ||||
---|---|---|---|---|---|---|
Tube load (mAs) | Tube voltage (kVp) | 2ABD (mGy) | Tube load (mAs) | Tube voltage (kVp) | 2ABD (mGy) | |
2.5 | 50 | 26 | 0.49 ± 0.11 | 39 | 27 | 0.95 ± 0.21 |
3.0 | 67 | 26 | 0.61 ± 0.15 | 38 | 28 | 0.95 ± 0.19 |
3.5 | 72 | 27 | 0.67 ± 0.15 | 40 | 29 | 1.05 ± 0.20 |
4.0 | 82 | 28 | 0.77 ± 0.16 | 49 | 29 | 1.18 ± 0.23 |
4.5 | 98 | 28 | 0.85 ± 0.18 | 50 | 30 | 1.26 ± 0.23 |
5.0 | 134 | 29 | 1.20 ± 0.24 | 55 | 31 | 1.45 ± 0.26 |
5.5 | 138 | 30 | 1.26 ± 0.23 | 60 | 32 | 1.66 ± 0.30 |
6.0 | 155 | 31 | 1.43 ± 0.26 | 65 | 33 | 1.86 ± 0.33 |
6.5 | 161 | 32 | 1.51 ± 0.28 | 71 | 34 | 2.13 ± 0.38 |
7.0* | 159 | 30 | 1.73 ± 0.36 | 78 | 35 | 2.43 ± 0.43 |
7.5* | 174 | 31 | 1.96 ± 0.40 | 82 | 36 | 2.70 ± 0.47 |