Table 5 Comparison between 2ABD in digital mammography and digital breast tomosynthesis for different phantom thicknesses on the reference Hologic Selenia Dimensions (device A)

From: Average absorbed breast dose (2ABD): an easy radiation dose index for digital breast tomosynthesis

Phantom thickness (cm)Digital mammography (single view) (W/Rh and W/Ag*)Digital breast tomosynthesis (W/Al)
Tube load (mAs)Tube voltage (kVp)2ABD (mGy)Tube load (mAs)Tube voltage (kVp)2ABD (mGy)
2.550260.49 ± 0.1139270.95 ± 0.21
3.067260.61 ± 0.1538280.95 ± 0.19
3.572270.67 ± 0.1540291.05 ± 0.20
4.082280.77 ± 0.1649291.18 ± 0.23
4.598280.85 ± 0.1850301.26 ± 0.23
5.0134291.20 ± 0.2455311.45 ± 0.26
5.5138301.26 ± 0.2360321.66 ± 0.30
6.0155311.43 ± 0.2665331.86 ± 0.33
6.5161321.51 ± 0.2871342.13 ± 0.38
7.0*159301.73 ± 0.3678352.43 ± 0.43
7.5*174311.96 ± 0.4082362.70 ± 0.47
  1. Data for 2ABD are presented as mean ± standard deviation. In DM modality, the W-Rh anode-filter combination was selected by the AEC system for a range of 2.5–6.5 cm of phantom thickness, while for a thickness of 7 cm and 7.5 cm the W-Ag (*) was automatically set. On the other hand, only the W-Al option was available for DBT