Table 4 Comparison between 2ABD computed from Eq. (1) and the average glandular dose (AGD) calculated by the Dance approach for 100% glandularity

From: Average absorbed breast dose (2ABD): an easy radiation dose index for digital breast tomosynthesis

Mammographic devicePhantom thickness (cm)Tube load (mAs)Tube voltage (kVp)2ABD (mGy)AGD 100% glandularity (mGy)Relative difference (%)
Hologic-A3.037.5280.94 ± 0.191.00 ± 0.20-6.0
Hologic-A3.540291.05 ± 0.201.07 ± 0.21-1.9
Hologic-A4.050291.21 ± 0.231.21 ± 0.240.0
Hologic-A4.550301.26 ± 0.231.27 ± 0.25-0.8
Hologic-A5.055311.45 ± 0.261.46 ± 0.29-0.7
Hologic-A5.560321.66 ± 0.301.67 ± 0.33-0.6
Hologic-A6.065331.87 ± 0.331.89 ± 0.38-1.1
Hologic-A6.570342.12 ± 0.372.14 ± 0.43-0.9
Hologic-A7.080352.51 ± 0.442.57 ± 0.51-2.3
Hologic-B3.530291.07 ± 0.201.14 ± 0.23-6.1
Hologic-B4.535301.20 ± 0.221.22 ± 0.24-1.6
Hologic-B5.542.5321.61 ± 0.291.64 ± 0.33-1.8
Hologic-B6.550342.06 ± 0.372.09 ± 0.42-1.4
Fuji-C3.528291.19 ± 0.241.21 ± 0.24-1.7
Fuji-C4.532321.63 ± 0.301.56 ± 0.314.5
Fuji-C5.542332.06 ± 0.372.04 ± 0.411.0
Fuji-C6.550352.66 ± 0.472.65 ± 0.530.4
Fuji-C4.5*45322.30 ± 0.432.15 ± 0.437.0
Fuji-C6.5*63353.35 ± 0.593.27 ± 0.652.4
Fuji-D3.528301.30 ± 0.261.38 ± 0.28-5.8
Fuji-D4.532321.59 ± 0.301.60 ± 0.32-0.6
Fuji-D5.550332.40 ± 0.432.43 ± 0.49-1.2
Fuji-D6.556352.91 ± 0.502.87 ± 0.571.4
Fuji-D3.5*32301.49 ± 0.281.55 ± 0.31-3.9
Fuji-D4.5*45322.24 ± 0.432.20 ± 0.441.8
Fuji-D5.5*63343.35 ± 0.613.22 ± 0.644.0
Fuji-D6.5*71353.69 ± 0.653.56 ± 0.713.7
  1. Data for 2ABD and AGD 100% glandularity are presented as mean ± standard deviation. The comparison was carried out by considering all mammographic devices. The exposure parameters were chosen as close as possible to the corresponding automatic exposure control settings for a given phantom thickness. For the AGD calculation by the Dance method, the measured air kerma was employed. For the Amulet Innovality devices both the standard acquisition mode and high-resolution acquisition mode (*) were employed