Skip to main content

Table 1 Comparison between measured and calculated (Eq. (2)) incident air kerma for the reference Hologic Selenia Dimensions (device A)

From: Average absorbed breast dose (2ABD): an easy radiation dose index for digital breast tomosynthesis

Phantom thickness (cm)

Tube load (mAs)

Tube voltage (kVp)

Measured air kerma (mGy)

Calculated air kerma (mGy)

Relative difference (%)

2.0

37.5

26

2.11 ± 0.08

2.06 ± 0.15

-2.6

2.5

40

27

2.61 ± 0.10

2.54 ± 0.19

-2.8

3.0

37.5

28

2.74 ± 0.11

2.71 ± 0.20

-1.3

3.5

40

29

3.33 ± 0.13

3.24 ± 0.24

-2.7

4.0

50

29

4.14 ± 0.17

4.12 ± 0.31

-0.4

4.5

50

30

4.66 ± 0.19

4.59 ± 0.34

-1.6

5.0

55

31

5.80 ± 0.23

5.60 ± 0.42

-3.6

5.5

60

32

7.08 ± 0.28

6.73 ± 0.50

-5.3

6.0

65

33

8.18 ± 0.33

7.99 ± 0.59

-2.4

6.5

70

34

9.78 ± 0.39

9.40 ± 0.70

-4.1

7.0

80

35

12.03 ± 0.48

11.69 ± 0.87

-2.9

7.5

80

36

13.02 ± 0.52

12.68 ± 0.94

-2.7

8.5

80

38

15.19 ± 0.61

14.78 ± 1.10

-2.8

8.5

90

42

21.52 ± 0.86

20.55 ± 1.53

-4.7

  1. Data for measured air kerma and calculated air kerma are presented as mean ± standard deviation. Note that the exposure parameters were chosen as close as possible to the corresponding automatic exposure control settings for a given phantom thickness