Patient | Tumour volume | Injected Ho | SPECT recovery | CT recovery | Volume of voxels with HU > 100 |
---|
| cm3 | MBq | mg | MBq | % | mg | % | cm3 | % |
---|
1 | 44.6 | 366.7 | 36.5 | 309.2 | 84.3 | 32.4 | 88.8 | 2.7 | 6.1 |
2 left | 5.6 | 53.9 | 13.7 | 9.6 | 17.8 | 2.3 | 16.5 | 0.5 | 8.0 |
2 right | 6.1 | 63.8 | 16.2 | 16.9 | 26.5 | 3.8 | 23.5 | 0.6 | 9.3 |
3 | 3.9 | 17.6 | 3.1 | 8.5 | 48.3 | 1.5 | 48.6 | 0.9 | 23.1 |
- Three patients with recurrent head and neck cancer were treated with radioactive 166Ho-microspheres and underwent SPECT/CT imaging. Patient 2 was treated on both sides of the neck. Base on the injected activity, the corresponding amount of mg Ho was calculated. Injected activity compared to SPECT recovery was 17.8–84.3%. CT recovery was in line with SPECT imaging ranging from 16.5 to 88.8%. Tumour volume was 3.9–44.6 cm3 based on the > 100 HU threshold; from 6.1 to 23.1% of tumour volume was filled with 166Ho-microspheres