Table 1 Sequence parameters of the 7-T shoulder MRI protocol.

From: 7-T clinical MRI of the shoulder in patients with suspected lesions of the rotator cuff

Weighting/type

Orientation

Type of sequence

Fat suppression

In-plane resolution (mm2)

Slice thickness (mm)

Number of slices

Distance factor (%)

Acquisition time (min)

TR (ms)

TE (ms)

FOV (mm2)

Standard sequences

 PD

Axial

Turbo spin-echo

Fat-sat

0.43 × 0.43

2.5

30

30

4:05

4,350

30

180 × 180

 PD*

Coronal

Turbo spin-echo

Fat-sat

0.43 × 0.43

2.5

23

30

4:05

4,350

30

180 × 180

 T1

Sagittal

Gradient-echo

None

0.45 × 0.35

2.0

40

20

5:24

172

4.1

180 × 180

Additional sequences

 PD

Sagittal

Turbo spin-echo

None

0.40 × 0.40

2.5

30

20

3:38

4,500

35

200 × 200

 PD

Coronal

Turbo spin-echo

None

0.40 × 0.40

2.5

30

20

3:15

4,500

35

200 × 200

 STIR

Coronal

Turbo spin-echo

Inversion time 300 ms

0.70 × 0.70

3.0

20

20

2:03

4,500

31

180 × 180

 MEDIC

Coronal

Multi-echo spoiled gradient-echo

None

0.35 × 0.35

1.5

39

50

5:21

1,000

15

180 × 180

 DESS 3D

Coronal

Steady-state free precession

Water excitation

0.70 × 0.70

0.7

128

0

4:56

8

2.4

190 × 190

 T1

Coronal

Gradient-echo

None

0.45 × 0.35

2.0

40

20

5:24

172

4.1

180 × 180

 T1

Coronal

Turbo spin-echo

None

0.40 × 0.40

2.5

9

20

1:43

900

11

200 × 200

  1. 3D Three-dimensional, DESS Dual-echo steady state, Fat-sat Fat saturated, FOV Field of view, MEDIC Multi-echo data image combination, PD Proton density, STIR Short-tau inversion recovery, TE Echo time, TR Repetition time. Except for T1-weighted TSE/GRE, all sequences used parallel imaging technique 2
  2. *According to guidelines, a double echo (PD- and T2-weighted) sequence should be used. However, as T2-imaging with adequate signal-to-noise ratio remains challenging at ultra-high fields, we decided to forego the second echo