Table 2 Comparison between k a,i (mGy) measured (k a,i meas ) and calculated by Equation 4 (k a,i calc ) for different mammography units. Y tb (mGy · mAs–1) is the yield of the mammographic device (calculated at 28 kVp, different for each anode-filter combination) and d (cm) is the breast thickness simulated by different solid-water phantoms. FID is the focus-image distance (cm) for the device considered

From: Average absorbed breast dose in mammography: a new possible dose index matching the requirements of the European Directive 2013/59/EURATOM

Device Anode-filter FID Y tb d kVp mAs k a,i meas k a,i calc 1- (ka,i calc/ka,i meas)
Fujifilm Amulet 1 Mo-Mo 64 0.094 4 28 20 2.14 2.13 –0.01
Fujifilm Amulet 2 Mo-Mo 64 0.090 6 28 20 2.23 2.16 –0.03
Fujifilm Amulet 1 Mo-Rh 64 0.075 2 28 20 1.62 1.61 –0.01
Fujifilm Amulet 2 Mo-Rh 64 0.079 4 28 20 1.80 1.91 –0.06
GE Senographe DS Rh-Rh 63.5 0.072 5 29 50 4.80 4.65 0.03
GE Senographe DS Rh-Rh 63.5 0.072 3 27 40 2.83 2.67 0.06
Fujifilm Amulet 1 W-Rh 64 0.031 4 28 20 0.71 0.79 0.10
Fujifilm Amulet 2 W-Rh 64 0.030 6 28 20 0.71 0.81 0.11
Giotto class W-Ag 67 0.0394 4 26 93 3.36 3.43 0.02
Giotto class W-Ag 67 0.0394 6 29 93 5.06 4.98 –0.02