Table 1 Yield Y 0 (mGy · mAs–1) of the reference device and α (mGy · mAs–1 · kVp–1) and β (mGy · mAs–1) for the five anode-filter combinations considered. The average value ± 1 standard deviation of five measurements of Y 0 for each anode-filter combination are reported. These measurements were performed at 28 kVp

From: Average absorbed breast dose in mammography: a new possible dose index matching the requirements of the European Directive 2013/59/EURATOM

  Mo-Mo Mo-Rh Rh-Rh W-Rh W-Ag
Y 0 0.0938 ± 0.0008 0.0739 ± 0.0008 0.0707 ± 0.0003 0.031 ± 0.001 0.0391 ± 0.0009
α 0.00997 ± 0.00005 0.00892 ± 0.00008 0.0078 ± 0.0002 0.00271 ± 0.00004 0.00371 ± 0.00009
β –0.186 ± 0.001 –0.180 ± 0.002 –0.149 ± 0.006 –0.047 ± 0.001 –0.066 ± 0.003